Comparison of accuracy between augmented reality/mixed reality techniques and conventional techniques for epidural anesthesia using a practice phantom model kit
BMC Anesthesiology volume 23, Article number: 171 (2023)
Abstract
Background
This study used an epidural anesthesia practice kit (model) to evaluate the accuracy of epidural anesthesia using standard techniques (blind) and augmented/mixed reality technology and whether visualization using augmented/mixed reality technology would facilitate epidural anesthesia.
Methods
This study was conducted at the Yamagata University Hospital (Yamagata, Japan) between February and June 2022. Thirty medical students with no experience in epidural anesthesia were randomly divided into augmented reality (-), augmented reality (+), and semi-augmented reality groups, with 10 students in each group. Epidural anesthesia was performed using the paramedian approach with an epidural anesthesia practice kit. The augmented reality (-) group performed epidural anesthesia without HoloLens2Ⓡ and the augmented reality (+) group with HoloLens2Ⓡ. The semi-augmented reality group performed epidural anesthesia without HoloLens2Ⓡ after 30 s of image construction of the spine using HoloLens2Ⓡ. The epidural space puncture point distance between the ideal insertion needle and participant’s insertion needle was compared.
Results
Four medical students in the augmented reality (-), zero in the augmented reality (+), and one in the semi-augmented reality groups failed to insert the needle into the epidural space. The epidural space puncture point distance for the augmented reality (-), augmented reality (+), and semi-augmented reality groups were 8.7 (5.7–14.3) mm, 3.5 (1.8–8.0) mm (P = 0.017), and 4.9 (3.2–5.9) mm (P = 0.027), respectively; a significant difference was observed between the two groups.
Conclusions
Augmented/mixed reality technology has the potential to contribute significantly to the improvement of epidural anesthesia techniques.
Background
Epidural anesthesia, the gold standard for postoperative analgesia, is used with general anesthesia in thoracic, abdominal, pelvic, and lower extremity surgery. The analgesic effect of epidural anesthesia is superior to that of paravertebral nerve block, and it may be effective in reducing postoperative cognitive dysfunction and stress response [1,2,3]. However, epidural anesthesia is difficult. Clinicians must locate the epidural space by feeling the epidural needle tip with their finger and “walking” it, which is performed blindly and requires experience, tactile acuity, and anatomical knowledge.
The incidence of insertion failure and epidural anesthesia difficulty is approximately 7% and as high as 26% among anesthesia residents [4, 5]. These occurrences are attributed to obesity, atypical spine anatomy, or supraspinous ligament ossification due to advanced age [6,7,8]. Difficulty inserting an epidural catheter is increasingly painful for the patient. Multiple epidural punctures also cause local pain. The risk of nerve damage increases and the procedure may be interrupted if the patients are unable to maintain posture during the puncture [5, 9]. Additionally, epidural anesthesia is associated with complications such as intravascular cannula insertion (0.67%), unintentional dural puncture (0.61%), and epidural hematoma [10,11,12]. Blind “walking” procedures performed by inexperienced clinicians have a high potential for complications. However, there are few educational tools for learning this technique in epidural anesthesia, and training is difficult. Methods for safe and accurate epidural anesthesia have been explored extensively, including devices for epidural needles and ultrasound echo-based epidural and fiber optic-based techniques; however, few devices support epidural anesthesia [13,14,15,16,17,18,19,20].
Recent developments in augmented reality/mixed reality (AR/MR) technology have facilitated the visualization of the invisible and have been utilized in various fields. In medicine, they are used as educational tools for ventricular drainage and extracorporeal membrane oxygenation cannulation [21,22,23,24]. Specifically, head-mounted AR/MR devices, such as Hololens2Ⓡ, can project images constructed by computed tomography (CT) of the patient’s anatomy onto their body, which would aid in clinical settings and medical training [25].
The blind technique is the gold standard for epidural anesthesia. Since the practice process is often performed blind, it is important to understand the anatomy of the patient’s spine. We hypothesized that the use of AR/MR technology to visualize internal structures using spine images constructed from the patient’s own CT images and projected onto the patient’s back would help facilitate a smooth and accurate epidural anesthesia procedure. Since the current epidural anesthesia practice method is only a blinded method using an epidural anesthesia practice kit, the blinded technique (standard technique) was used as a control group, and this study aimed to examine the usefulness of using a tool (AR/MR technology) for visualization during the procedure. In addition, the purpose of this study was to examine the usefulness of using AR/MR technology as a tool for constructing images to improve the accuracy of the procedure. This study examined the possibility of using AR/MR for epidural needle placement in epidural anesthesia in accordance with textbooks and whether AR/MR technology can be used as a training tool for medical students (MSs) in epidural anesthesia techniques.
If the feasibility of epidural anesthesia using AR/MR is proven, difficulties during epidural catheter insertion can be avoided, and patient pain and anesthesia induction time can be reduced. Improved epidural anesthesia techniques may resolve issues with epidural anesthesia in obese patients and patients with atypical spinal anatomy or supraspinous ligament ossification.
Methods
This study was approved by the Ethics Committee of Yamagata University and conducted in accordance with the Declaration of Helsinki (The approval numbers are 2021 − 346 and UMIN000046701). Written informed consent was obtained from all participants.
Model
We used a commercially available phantom epidural anesthesia practice kit for the lumbar spine (Lumbar Puncture Simulator II; KYOTO Kagaku, Kyoto, Japan) equipped with silicone skin and air-filled epidural space, and an 18G Tuohy needle (PericanⓇ, B. Braun, Melsungen, Germany) (Fig. 1a). A CT scanner (Aquilion-One Vision, Canon Medical Systems Corporation, Otawara, Japan) created 0.5-mm cross-sectional images at 0.25-mm intervals for the construction of images of the epidural anesthesia practice kit. The images were processed for metal removal with metal artifact reduction algorithms to improve image quality and prevent obstructions in the area of interest. An ideal insertion model was captured when a needle was inserted into the epidural puncture pad inside the epidural anesthesia practice kit (Supplementary Fig. 1). The insertion method was determined to be in accordance with the textbook, where the needle entry point was 1 cm lateral and caudal to the space between the cephalic and caudal spinous processes, and the needle entry angle was 15° to the vertical; the needle was then “walked” cephalad and placed in the epidural space using the loss of resistance method (Fig. 1b) [26]. The epidural needle was cut with nippers (Fig. 1c). The hologram data sets of the ideal insertion model and epidural puncture pad were imported into the HoloLens2Ⓡ device (Microsoft Corp., Redmond, WA, USA) and stored as an application for use in epidural anesthesia. The AR marker was attached to the head side of the epidural anesthesia practice kit. HoloLens2Ⓡ was set to recognize the AR marker, and a hologram of the ideal insertion model was projected at the same location as the pad inside the epidural anesthesia practice kit (Fig. 1d). Vuforia SDK (PTC Inc., Boston, MA, USA) was used for the AR markers and Unity 2019.4.9 (Unity Technologies, San Francisco, CA, USA) for the holograms. A developmental toolkit for MR applications (MRTK 2.4.0, Microsoft Corp.) was also used. The operator was able to observe a hologram of the ideal insertion model projected on the back of the epidural anesthesia practice kit while using HoloLens2Ⓡ (Supplementary Fig. 2).
Procedure
This study included 30 MSs with no previous experience performing epidural anesthesia undergoing bedside training at the Yamagata University Hospital between February and June 2022. In this study, MSs over the age of 30 years were excluded from the study because of the possibility that they might have had previous experience in other medical work and to avoid allowing this experience to influence the results of the study. MSs who were unable to perform the procedure according to the video lectures and procedures described below were also excluded. MSs were randomly divided into AR(-), AR(+), and semi-augmented reality (SemiAR) groups, with 10 MSs in each group (Fig. 2). In this study, the sample size was set at 10 persons per group based on previous similar studies [24]. All MSs watched a 10-min video lecture on epidural anesthesia using the paramedian approach; the AR(+) and SemiAR groups were given an additional 1-min lecture on AR-specific visualization. Figure 3 shows the grouping of Test 1 (first procedure) and Test 2 (second procedure).
Test 1
For the first epidural puncture, epidural anesthesia was performed using the paramedian approach with the epidural anesthesia practice kit according to the following method for each group. The epidural anesthesia procedure was written on paper and placed in a visible position for the operator during the procedure. The procedure manual was designed according to Miller’s Anesthesiology [26]. The procedure time was measured from the start of the epidural needle puncture until the end of the procedure.
-
The AR(-) group performed epidural anesthesia according to the procedure manual and was designated AR(-)_test1.
-
In the AR(+) group, participants used HoloLens2Ⓡ after confirming its activation and observed a hologram of the ideal insertion model for 30 s. Epidural anesthesia was performed according to the procedure manual while using HoloLens2Ⓡ. This group was designated AR(+)_test1.
-
In the SemiAR group, after confirming its activation, HoloLens2Ⓡ was worn, and a hologram of the ideal insertion model was observed for 30 s. HoloLens2Ⓡ was then removed, and epidural anesthesia was performed according to the procedure manual without using HoloLens2Ⓡ. This group was designated as SemiAR_test1.
Epidural anesthesia was performed using the loss of resistance method [26]. The procedure ended when the MS declared that they had finished. The needle was then cut with nippers and retained in place; the epidural puncture pad was removed from the epidural practice kit so that it was not visible to the operator. The removed epidural puncture pad was analyzed by CT to identify the surface puncture point of the spine, epidural needle angle, and epidural space puncture point. Using Blender 2.83.9 software (Blender Foundation, Amsterdam, Netherlands), the distance to the spinal puncture point between the needle of the model that the MS punctured and ideal insertion model (surface puncture point distance [SPPD]), angle difference of the needle (puncture angle [PA]), and distance to the epidural space puncture point (epidural space puncture point distance [ESPPD]) were measured. The PA was measured using the inner product of vectors (Supplementary Fig. 3). Measurements were taken by an individual who was not involved in the study and blinded to the model groups.
Test 2
For the second epidural puncture, all groups (AR(-), AR(+), and SemiAR) performed epidural anesthesia via the paramedian approach without HoloLens2Ⓡ using the epidural anesthesia practice kit. Written instructions were provided and placed in a position visible to the operator during the procedure.
-
The AR(-) group performed the epidural anesthesia again after being instructed that the puncture should be more precise than the previous one. This group was designated AR(-)_test2.
-
The AR(+) group retained the image of the hologram of the ideal insertion model at Test 1, and epidural anesthesia was performed without HoloLens2Ⓡ. This group was designated AR(+)_test2.
-
The SemiAR group retained the image of the hologram of the ideal insertion model at Test 1, and epidural anesthesia was performed without HoloLens2Ⓡ. This group was designated SemiAR_test2.
Data acquisition and measurements were performed as in Test 1.
User experience questionnaire
The user experience questionnaire (UEQ) was administered to all MSs (Supplementary Table 1) [27]. MSs in the AR(-) group had no experience with HoloLens2Ⓡ; therefore, they used HoloLens2Ⓡ and observed the holograms after Test 2 was completed before completing the questionnaire. The UEQ comprised 26 questions that scored attractiveness, perspicuity, novelty, stimulation, efficiency, and dependability. The questions were scored from 1 to 7, with negative answers receiving a -3 score and positive answers a + 3 score.
Evaluation
Figure 1E shows the surface puncture point, PA, and epidural space puncture point in the ideal insertion model. Test 1 measured the SPPD, PA, and ESPPD for the needle of the model with the MS puncture and needle of the ideal insertion model, with the AR(-) group as the control and the AR(+) and SemiAR groups as comparisons. Test 2 examined the effect of learning using AR/MR by comparing the ESPPD among the groups in Tests 1 and 2.
Dural punctures were excluded from evaluation in the epidural practice kit as anatomically improbable situations can arise (Supplementary Fig. 4). To evaluate models in which the needle could not be implanted in the epidural space, the needle was extended using Blender 2.83.9 software, and the virtual entry point into the epidural space was evaluated as the epidural space puncture point. The primary outcome was the ESPPD of Test 1 in each group, and the secondary outcomes were the SPPD of Test 1, PA of Test 1, ESPPD of Test 2, PA of Test 2, SPPD of Test 2, time required for the procedure, and UEQ in each group.
Statistical analyses
EZR, version 1.41, was used for statistical analyses. For continuous variables, histogram visualization and the Shapiro–Wilk test were used to determine the pattern of variable distribution. The F-test was used to confirm equal variances. Continuous variables with normal distributions and equal variances were expressed as means ± standard deviations and analyzed with the Student’s t-test. Continuous variables that were not normally distributed or equally distributed were expressed as medians (interquartile range, 25–75%) and analyzed with the Mann–Whitney U test. As we believe that administering epidural anesthesia with visualization is superior to administering it blindly, a one-tailed test was used to examine the SPPD, PA, and ESPPD of the model with MS insertion and model with the needle inserted according to the textbook. One-tailed tests were also performed for the comparison of the epidural anesthesia procedure time and UEQ. Tests 1 and 2 were compared using the one-tailed Wilcoxon signed-rank sum test. P-values < 0.05 were considered statistically significant.
Results
The mean age of the MSs was 24.0 (23.0–25.0) years (19 male; 11 female) (Supplementary Table 2); none had experience with mixed reality head-mounted smart glasses (e.g., HoloLens2Ⓡ). There were no missing values with respect to the data used to assess the participant’s needle insertion position.
Test 1
Four MSs in AR(-)_test1, zero in AR(+)_test1, and one in SemiAR_test1 were unable to puncture the epidural space. Table 1 shows the SPPD, PA, and ESPPD for AR(-)_test1 and AR(+)_test1. Table 2 shows the SPPD, PA, and ESPPD for AR(-)_test1 and SemiAR_test1. The SPPD in AR(-)_test1 and AR(+)_test1 was 11.2 ± 4.8 and 9.1 ± 5.1 mm, respectively (P = 0.178); the PA in AR(-)_test1 and AR(+)_test1 was 15.5 ± 3.6 and 8.5 ± 4.4 degree, respectively (P < 0.01); and the ESPPD in AR(-)_test1 and AR(+)_test1 was 8.7 (5.7–14.3) and 3.5 (1.8–8.0) mm (P = 0.017), respectively. This indicates a significant difference in the PA and ESPPD (Table 1, Supplementary Fig. 5).
In AR(-)_test1 and SemiAR_test1, the SPPD was 11.2 ± 4.8 and 7.5 ± 4.2 mm, respectively (P = 0.04); the PA was 15.5 ± 3.6 and 8.0 ± 4.4 degrees, respectively (P < 0.01); and the ESPPD in AR(-)_test1 and SemiAR_test1 was 8.7 (5.7–14.3) and 4.9 (3.2–5.9) mm (P = 0.027), respectively. Significant differences were observed for SPPD, PA, and ESPPD (Table 2, Supplementary Fig. 5).
Figure 4 A–B shows the distribution of all surface puncture points and PAs in Test 1, respectively. AR(-)_test1 showed dispersed PAs, while AR(+)_test1 and SemiAR_test1 showed uniform PAs. Figure 4 C shows the distribution of epidural space puncture points: AR(-)_test1 was distributed, while AR(+)_test1 and SemiAR_test1 were concentrated.
The execution time of Test 1 was 117.5 (112.0–143.5) s, 148.0 (139.0–164.8) s, and 122.0 (95.5–136.3) s for the AR(-)_test1, AR(+)_test1, and SemiAR_test1, respectively. (Tables 1 and 2).
Test 2
Test 2 evaluated whether MSs could insert the epidural needle in the appropriate position without AR/MR, as compared with Test 1. One MS in AR(-)_test2, zero in AR(+)_test2, and zero in SemiAR_test2 could not puncture the epidural space. When comparing Tests 1 and 2, there was a tendency for the ESPPD to approach the ideal insertion model in the AR(-) and SemiAR groups. However, the difference was not statistically significant (Supplementary Table 3). The ESPPD did not differ significantly between AR(-)_test2 and AR(+)_test2 (5.9 [4.0–10.6 vs. 5.0 [2.5–5.9] mm [P = 0.192]) (Table 3). However, there was a significant difference in the ESPPD between AR(-)_test2 and SemiAR_test2, 5.9 (4.0–10.6) vs. 3.4 (2.5–5.5) mm (P = 0.045) (Table 4).
The execution time for Test 2 was 100.0 (79.5–128.5) s, 115.0 (97.3–143.0) s, and 93.0 (79.8-159.5) s for AR(-)_test2, AR(+)_test2, and SemiAR_test2, respectively (Tables 3 and 4). The total time for Tests 1 and 2 was 226.5 (201.0-251.5) s for the AR(-) group, 288.5 (239.5-382.3) s for the AR(+) group, and 232.5 (211.0-293.0) s for the SemiAR group (Tables 5 and 6).
Results of the user experience questionnaire
The AR(+) group had significantly more positive answers to questions in the perspicuity, novelty, stimulation, dependability, and efficiency categories of the UEQ than the AR(-) group. The SemiAR group also provided significantly more positive responses in all items of the UEQ than the AR(-) group (Supplementary Fig. 6).
Discussion
In Test 1, AR(+)_test1 MSs were able to place the epidural needle close to the ideal locations for the PA and ESPPD. In SemiAR_test1, the epidural needle was placed close to the ideal needle in all cases (SPPD, PA, and ESPPD).
These results indicate that epidural anesthesia with AR may be more appropriate for epidural needle placement than epidural anesthesia with textbook or video learning alone. The large variance in the SPPD for the AR(+)_test1 MSs may be attributed to none of the participants having experience using mixed reality head-mounted smart glasses; thus, the MSs may have misidentified the surface puncture point because they had not mastered AR sensation. Alternatively, the registration accuracy, and inherent overlay accuracy and stability could have also had some impact on the variance in the SPPD [28].
Test 2 resulted in the SemiAR_test2 group placing the epidural needle closer to the ideal needle model at the epidural space puncture point compared to AR(-)_test2. The SemiAR group performed image training using a HoloLens2Ⓡ and performed epidural anesthesia without AR in both Tests 1 and 2. This tended to improve epidural anesthesia technique.
However, a comparison of AR(+)_test1 and AR(+)_test2 showed no trend toward successful epidural anesthesia. This could be attributed to the fact that the AR(+) group used HoloLens2Ⓡ in Test 1 and did not use it in Test 2, resulting in a significantly different environment between the first and second sessions.
The UEQ results indicate that many students had a positive perception of AR. We believe that AR is a useful learning tool in terms of attracting students’ interest. In terms of time, the results of Tests 1 and 2 showed that the use of AR/MR technology did not contribute to a reduction in procedure time.
A previous study has examined the possibility of epidural anesthesia using AR [29]. However, ours is the first to compare the accuracy of epidural anesthesia performed by MSs with no epidural anesthesia experience using AR with that of conventional epidural anesthesia. This study suggests that AR/MR has the potential to aid in the understanding of anatomical structures and improve epidural anesthesia techniques [30].
There is no practice in clinical practice, it is always an actual procedure, which in the case of this study is the Test 1 situation. In Test 1, the SemiAR group had the best procedure accuracy, suggesting that epidural anesthesia using AR/MR technology may be effective for image training before the actual surgery. For residents inexperienced in epidural anesthesia, AR/MR technology appears to be a sufficient tool to learn the optimal insertion angle technique when “walking” the needle tip. Since most patients undergoing thoracic and abdominal surgery undergo CT, we believe that AR/MR technology can be applied clinically and that previewing epidural approaches with actual patient CT images before performing epidural anesthesia promotes safety. Furthermore, epidural anesthesia AR/MR is expected to be used by experienced anesthesiologists as a technical learning tool to perform epidural anesthesia more accurately and safely.
Some HoloLens2Ⓡ medical applications have been approved by the US Food and Drug Administration. In the future, epidural anesthesia may be performed with HoloLens2Ⓡ, thereby improving the understanding of the anatomy and safety of epidural anesthesia. This will facilitate epidural anesthesia in patients with obesity and spinal stenosis, for whom locating the epidural space is difficult, and avoid unintentional dural and vascular puncture by the epidural needle.
Limitations
There are eight limitations of this study. First, the commercially available phantoms are practice kits for epidural anesthesia and have a larger gap than the actual vertebral arch, which can cause the epidural needle to accidentally penetrate the epidural space puncture point. Future research should be conducted using epidural anesthesia practice kits with smaller gaps or epidural anesthesia AR/MR in actual clinical practice. Second, blinding was impossible for MSs. Third, the sample size was determined with reference to previous similar studies. Increasing the sample size in future studies may provide more reliable results. Fourth, the AR markers were used to visualize the virtual spine, and the epidural anesthesia practice kit was held in place to prevent misalignment; however, the pressure applied by the participant during the puncture may have caused some movement, resulting in misalignment of the virtual spine. Fifth, we did not consider the possibility that the needle could successfully approach the epidural space even when the surface puncture point was located far from the ideal surface puncture point in this study. In clinical practice, the needle can approach the epidural space by changing the angle of the needle, even when the insertion point is located far from the ideal point. This study determined how close the epidural needle can be inserted to the model entry point described in Miller’s Anesthesiology (textbook), which is currently considered ideal. Therefore, there may be differences in clinical situations. We believe it is important to continue our research to further improve the practicality of applying AR/MR technology while validating it in clinical situations. Sixth, it may be necessary to consider the time factor in the data acquisition process. In the AR(-) group, there was a period of time between the video lecture and start of epidural anesthesia. In the AR(+) and SemiAR groups, there was a period of time between the video lecture and start of hologram observation, and a period of time between the end of hologram observation and start of epidural anesthesia. Thus, there were different gaps in time periods in each group. In addition, although we took care to avoid time gaps in timing, such as the replacement of the epidural puncture pad in all groups, there could have been a difference of a few seconds in each group. Seventh is an examination of the time factor when AR/MR technology is applied in actual clinical practice. Although AR/MR data acquisition, processing, and registration are likely to be required in actual clinical settings, these times were not measured in this study. We have confirmed that it takes approximately 1 min from the time the HoloLens2Ⓡ is turned on to the time the application is launched, although it is necessary to examine the exact time required in actual clinical settings. Eighth, we have not examined the comparison between the administration of epidural anesthesia using AR/MR technology and other technologies. In future studies, it would be appropriate to consider comparisons with other techniques such as Fiducial Registered CT imaging and ultrasound-guided access in future studies.
Conclusions
These findings suggest that AR/MR technology is a useful tool to construct an image of epidural anesthesia and can significantly contribute to improved epidural anesthesia skills, potentially stabilizing patient conditions, and solving medically important epidural anesthesia problems.
Data Availability
The data underlying this article will be shared on reasonable request to the corresponding author.
Abbreviations
- AR/MR:
-
Augmented reality/mixed reality
- CT:
-
Computed tomography
- ESPPD:
-
Epidural space puncture point distance
- MSs:
-
Medical students
- PA:
-
Puncture angle
- SemiAR:
-
Semi-augmented reality
- SPPD:
-
Surface puncture point distance
- UEQ:
-
User experience questionnaire
References
Tamura T, Mori S, Mori A, Ando M, Yokota S, Shibata Y, et al. A randomized controlled trial comparing paravertebral block via the surgical field with thoracic epidural block using ropivacaine for post-thoracotomy pain relief. J Anesth. 2017;31:263–70.
Schreiber KL, Chelly JE, Lang RS, Abuelkasem E, Geller DA, Marsh JW, et al. Epidural versus paravertebral nerve block for postoperative analgesia in patients undergoing open liver resection: a randomized clinical trial. Reg Anesth Pain Med. 2016;41:460–8.
Meng Z, Gao C, Li X, Shen J, Hong T, He X, et al. Effects of combined epidural anesthesia and general anesthesia on cognitive function and stress responses of elderly patients undergoing liver cancer surgery. J Oncol. 2021;2021:8273722.
Kula AO, Riess ML, Ellinas EH. Increasing body mass index predicts increasing difficulty, failure rate, and time to discovery of failure of epidural anesthesia in laboring patients. J Clin Anesth. 2017;37:154–8.
Faitot V, Ourchane R, Dahmani S, Magheru M, Nebout S, Gomas F, et al. An observational study of factors leading to difficulty in resident anaesthesiologists identifying the epidural space in obstetric patients. Int J Obstet Anesth. 2011;20:124–7.
Prakash S, Mullick P, Kumar SS, Diwan S, Singh R. Factors predicting difficult spinal block: a single centre study. J Anaesthesiol Clin Pharmacol. 2021;37:395–401.
Chien I, Lu IC, Wang FY, Soo LY, Yu KL, Tang CS. Spinal process landmark as a predicting factor for difficult epidural block: a prospective study in taiwanese patients. Kaohsiung J Med Sci. 2003;19:563–8.
Murata H, Sakai T, Goto S, Sumikawa K. Three-dimensional computed tomography for difficult thoracic epidural needle placement. Anesth Analg. 2008;106:654–8.
Kang X, Zhu Y, Lin K, Xie L, Wen H, Geng W, et al. The incidence of and risk factors for localized pain at the epidural insertion site after epidural anesthesia: a prospective survey of more than 5000 cases in nonobstetric surgery. Risk Manag Healthc Policy. 2021;14:2171–80.
Tanaka K, Watanabe R, Harada T, Dan K. Extensive application of epidural anesthesia and analgesia in a university hospital: incidence of complications related to technique. Reg Anesth. 1993;18:34–8.
Moen V, Dahlgren N, Irestedt L. Severe neurological complications after central neuraxial blockades in Sweden 1990–1999. Anesthesiology. 2004;101:950–9.
Lerner SM, Gutterman P, Jenkins F. Epidural hematoma and paraplegia after numerous lumbar punctures. Anesthesiology. 1973;39:550–1.
Sawada A, Kii N, Yoshikawa Y, Yamakage M. Epidrum(Ⓡ): a new device to identify the epidural space with an epidural Tuohy needle. J Anesth. 2012;26:292–5.
Chin A, Crooke B, Heywood L, Brijball R, Pelecanos AM, Abeypala W. A randomised controlled trial comparing needle movements during combined spinal-epidural anaesthesia with and without ultrasound assistance. Anaesthesia. 2018;73:466–73.
Pan PH, Sintay BJ. Eyes to the needle: to assist identification of the epidural space. Anesthesiology. 2010;112:1073–5.
Jain K, Puri A, Taneja R, Jaiswal V, Jain A. Preprocedural ultrasound as an adjunct to blind conventional technique for epidural neuraxial blockade in patients undergoing hip or knee joint replacement surgery: a randomised controlled trial. Indian J Anaesth. 2019;63:924–31.
Beigi P, Malenfant P, Rasoulian A, Rohling R, Dube A, Gunka V. Three-dimensional ultrasound-guided real-time midline epidural needle placement with epiguide: a prospective feasibility study. Ultrasound Med Biol. 2017;43:375–9.
Rasoulian A, Lohser J, Najafi M, Rafii-Tari H, Tran D, Kamani AA, et al. Utility of prepuncture ultrasound for localization of the thoracic epidural space. Can J Anaesth. 2011;58:815–23.
Grau T, Bartusseck E, Conradi R, Martin E, Motsch J. Ultrasound imaging improves learning curves in obstetric epidural anesthesia: a preliminary study. Can J Anaesth. 2003;50:1047–50.
Tran D, Kamani AA, Al-Attas E, Lessoway VA, Massey S, Rohling RN. Single-operator real-time ultrasound-guidance to aim and insert a lumbar epidural needle. Can J Anaesth. 2010;57:313–21.
Hecht R, Li M, de Ruiter QMB, Pritchard WF, Li X, Krishnasamy V, et al. Smartphone augmented reality CT-based platform for needle insertion guidance: a phantom study. Cardiovasc Intervent Radiol. 2020;43:756–64.
Dennler C, Jaberg L, Spirig J, Agten C, Götschi T, Fürnstahl P, et al. Augmented reality-based navigation increases precision of pedicle screw insertion. J Orthop Surg Res. 2020;15:174.
Schneider M, Kunz C, Pal’a A, Wirtz CR, Mathis-Ullrich F, Hlaváč M. Augmented reality-assisted ventriculostomy. Neurosurg Focus. 2021;50:E16.
Wolf J, Wolfer V, Halbe M, Maisano F, Lohmeyer Q, Meboldt M. Comparing the effectiveness of augmented reality-based and conventional instructions during single ECMO cannulation training. Int J Comput Assist Radiol Surg. 2021;16:1171–80.
Blum T, Heining SM, Kutter O, Navab N. Advanced training methods using an augmented reality ultrasound simulator 8th IEEE International Symposium on Mixed and Augmented Reality, 2009. Orlando, Florida: ISMAR, October 19–22, 2009; 2009. p. 177-8.
Brown D. Subarachnoid anesthesia, epidural anesthesia, sacral anesthesia. In: Miller RD, editor. Miller’s anesthesia. 6th ed. Tokyo: Medical Science International; 2007. pp. 1287–310.
Laugwitz B, Held T, Schrepp M. Construction and evaluation of a user experience questionnaire. In: Lect Notes Comput Sci. Holzinger A, editor. HCI and Usability for Education and Work. USAB 2008. Berlin, Heidelberg: Springer;5298:63–76.
Condino S, Carbone M, Piazza R, Ferrari M, Ferrari V. Perceptual limits of optical see-through Visors for Augmented reality Guidance of Manual Tasks. IEEE Trans Bio Med Eng. 2020;67:411–9.
Lim S, Ha J, Yoon S, Tae Sohn Y, Seo J, Chul Koh J, et al. Augmented reality assisted surgical navigation system for epidural needle intervention. Annu Int Conf IEEE Eng Med Biol Soc. 2021;2021:4705–8.
Kassutto SM, Baston C, Clancy C. Virtual, augmented, and alternate reality in medical education: socially distanced but fully immersed. ATS Sch. 2021;2:651–64.
Acknowledgements
We thank the doctors and radiologists in the Department of Anesthesiology and Diagnostic Radiology, Yamagata University Hospital, and Mr. Toshimitsu Sato and Mr. Kazunori Sato for their cooperation in this study.
Funding
The authors have no sources of funding to declare for this manuscript.
Author information
Authors and Affiliations
Contributions
Study design: TH, YO. Data analysis: TH, KK. Revision of this manuscript: YO, HS, MN, MK, and KK. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
This study was approved by the Ethics Committee of Yamagata University and conducted in accordance with the Declaration of Helsinki (The approval numbers are 2021 − 346 and UMIN000046701). Written informed consent was obtained from all participants.
Consent for publication
Not applicable.
Competing interests
The authors declare that they have no competing interests.
Additional information
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Electronic supplementary material
Below is the link to the electronic supplementary material.
12871_2023_2133_MOESM4_ESM.doc
Supplementary Figure 1: CT scanner image of the ideal insertion model: CT scanner images of the ideal insertion model from the side, top, and front views
12871_2023_2133_MOESM5_ESM.doc
Supplementary Figure 2: Posture of epidural anesthesia operator (without and with HoloLens2Ⓡ) : The epidural anesthesia operator wearing HoloLens2Ⓡ can observe the hologram of the ideal insertion model projected on the back of the epidural anesthesia practice kit
12871_2023_2133_MOESM6_ESM.doc
Supplementary Figure 3: Formula for the inner product of vectors: The epidural needle puncture angle of the ideal insertion model and the participants’ epidural needle puncture angle were calculated using the inner product of vectors
12871_2023_2133_MOESM7_ESM.doc
Supplementary Figure 4: Evaluation of dural puncture and internal structure of epidural anesthesia practice kit: Dural punctures were excluded from the evaluation of the epidural anesthesia practice kit because of the possibility of anatomically improbable puncture situations, as indicated by the arrows
12871_2023_2133_MOESM8_ESM.doc
Supplementary Figure 5: Comparison of the SPPD, PA, and ESPPD in AR(+)_test1 and SemiAR_test1, with AR(-)_test1 as the control group: Regarding the SPPD, SemiAR_test1 had a surface puncture point closer to the ideal needle model (A2). Regarding the PA, both AR(+)_test1 and SemiAR_test1 had a puncture angle close to the ideal needle model (B1, B2). Regarding the ESPPD, both the AR(+)_test1 and SemiAR_test1 groups had epidural space puncture points closer to the ideal needle model (C1, C2). Note: * indicates significant between-group difference (P<0.05)
12871_2023_2133_MOESM9_ESM.doc
Supplementary Figure 6: UEQ results for the AR(-) group as the control group: The AR(+) group had significantly more positive answers to questions in the perspicuity, novelty, stimulation, dependability, and efficiency categories of the UEQ than the AR(-) group. The SemiAR group also obtained significantly more positive responses in all items of the UEQ than the AR(-) group. Note: * indicates a significant between-group difference (P<0.05)
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
About this article
Cite this article
Hayasaka, T., Kawano, K., Onodera, Y. et al. Comparison of accuracy between augmented reality/mixed reality techniques and conventional techniques for epidural anesthesia using a practice phantom model kit. BMC Anesthesiol 23, 171 (2023). https://doi.org/10.1186/s12871-023-02133-w
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s12871-023-02133-w