Comparing suction rates of novel DuCanto catheter against Yankauer and standard suction catheter using liquids of different viscosity—a technical simulation

Abstract

Purpose

Aspiration is a feared complication that may occur during airway management, and can significantly contribute to morbidity and mortality. Availability of a suctioning device with a suction catheter capable of clearing the airway is mandatory for airway management. However, suction performance may be significantly different amongst different suction catheters. The aim of this study was to compare suction rates of a standard 14 Ch suction catheter (SC), a Yankauer catheter (Y) and a DuCanto catheter (DC) using 4 fluids with different viscosity.

Methods

In this simulation trial, 4 preparations with standardized viscosity were prepared using a Xanthane-based medical fluid thickener. Lowest viscosity was achieved using tap water without thickener, syrup-like viscosity was achieved by adding 10 g per liter tap water, honey-like viscosity was achieved by adding 20 g per liter, and a pudding-like viscosity was achieved by adding 30 g of thickening powder per liter tap water. Each preparation was suctioned for 15 s with the three different suctioning devices. Measurements were repeated four times. The amount of removed preparation by suctioning was measured using a tared scale.

Results

Suction rates for water were 580 ± 34 mg for SC, 888 ± 5 mg for Y and 1087 ± 15 for DC; for syrup-like viscosity it was 383 ± 34(SC) vs. 661 ± 64(Y) vs. 935 ± 42(DC); for honey-like viscosity it was 191 ± 21(SC) vs. 426 ± 34(Y) vs. 590 ± 68(DC); and for pudding-like viscosity 74 ± 13(SC) vs. 164 ± 6(Y) vs. 211 ± 8(DC).

Conclusion

Suctioning liquids of different viscosity, the new DuCanto catheter was more effective than the Yankauer catheter that was more effective than a standard suctioning catheter. The relative superiority of the DuCanto was highest in fluids with high viscosity.

Peer Review reports

Introduction

Failed airway management remains a major contributor towards morbidity and mortality in modern anesthesia, intensive care and emergency medicine [1,2,3]. The knowledge of the importance of a secure airway for general anesthesia is almost as old as modern anesthesia itself. Following the first public demonstration of general anesthesia with ether by William Thomas Morton in 1846, James Young Simpson reported the first death due to pulmonary aspiration during chloroform induced anesthesia in 1848 [4]. Today, anesthesiologists, critical care and emergency doctors take multiple precautions in order to prevent pulmonary aspiration when inducing general anesthesia, but regurgitation of gastric content can occur despite all precautions, as a review of 166,491 anesthesia records in a recent study showed: in 20 cases, they found pulmonary aspiration, and in an additional 20 cases there were signs of regurgitation without pulmonary aspiration [5]. Pulmonary aspiration can have varying clinical consequences – ranging from clinically none, to aspiration pneumonia due to bacteria transported into the lung alongside with the aspirate, [6] right up to severe acute respiratory distress syndrome (ARDS) with distinct hypoxemia – named “Mendelson Syndrome" after Curtis L. Mendelson, who described a case series of 66 healthy obstetrical patients with chemical pneumonitis after ether anesthesia [7]. Whenever a patient is at risk for aspiration, a modified intubation technique called “rapid sequence induction” (RSI) is accepted as the technique of choice for securing the airway [8]. When performing RSI, induction agents are injected in rapid sequence and without bag-mask ventilation [9]. Usually, a large-bore suction catheter will be available to suction any aspirate that might occur during the procedure. These suction catheters may have difficulties as suction rates can decline with increasing viscosity of the aspirate and may be blocked by particles within the aspirate [10]. However, there are different possibilities of what kind of fluid must be suctioned to clear the airway – from watery solutions such as beverages, to stomach acid, airway secretions, blood, and coagulated blood, and different kinds of food. Accordingly, different levels of viscosity must be expected and handled.

To our knowledge. The DuCanto catheter represent the largest-bore commercially available suction catheter for adult patients. Therefore, in this study, we evaluated the DuCanto catheter against our current clinical standard Yankauer suction catheter, and the largest clinically available standard suction catheter, using liquids of four standardized viscosities in a prospective technical simulation study. Primary endpoint was suction rate (in mg/min) of the different catheters with different standardized viscosities. Our formal hypothesis was that there are no differences between groups.

Materials and methods

Ethics approval

This study was designed as a technical simulation and was conducted in our laboratory at the University Hospital of Cologne. As neither patients nor patient data were investigated, approval from the local Ethics Committee was not required for this type of study.

Experimental setup

We compared (a) a standard 14 Charrière (Ch) suction catheter (ProFlo with funnel, outer diameter 4.667 mm [Ch 14], length 60 cm; ConvaTec Ltd., Deeside, United Kingdom), (b) a standard large-bore rigid Yankauer suction catheter (Yankauer Midi, outer diameter 6.0 mm [Ch 18], length 27 cm; EXTRUDAN Surgery ApS, Birkerød, Denmark), both of which were in use in clinical daily routine, and (c) the novel SSCOR DuCanto Catheter (DuCanto Catheter with outer diameter 9.3 mm [Ch 28], length 24.9 cm; SSCOR, Inc., Sun Valley, CA, United States of America). Figure 1 shows the suction catheters used. These were connected to a standard suction line used for clinical routine in our hospital (suction line 1.8 m, Ch 25; Ref. 1324.1800.80ML, Extrudan Surgery ApS, Denmark). Negative pressure of the suction unit was set to the maximum, which was -0.8 bar (approx.—600 mmHg).

Fig. 1
figure 1

ConvaTec 14Ch suction catheter (left), Extrudan Yankauer 18Ch suction catheter (middle) and SSCOR DuCanto 28Ch suction catheter (right)

We measured the suction rates of each catheter using four different fluids with standardized levels of viscosity. To achieve these, a Xanthan-based thickening medium for use in patients with dysphagia (seneoPro VISCOclear instant, biozoon GmbH, Bremerhaven, Germany) was added to tap water to achieve standardized viscosities, as instructed by the manufacturer [11]. Lowest viscosity (A) was achieved using tap water without thickener, syrup-like viscosity (B) was achieved by adding 10 g per liter tap water, honey-like viscosity (C) was achieved by adding 20 g per liter, and a pudding-like viscosity (D) was achieved by adding 30 g of thickening powder per liter tap water. These mixtures were stirred well using a whisk to ensure a homogenously distribution, and thickening was allowed for 5 min prior to starting to ensure that a proper effect had taken place according to the manufacturer’s specifications.

Experimental procedure

Measurements were repeated four times for each suction catheter. Suctioning was accomplished by keeping the tip of every catheter beneath the surface, but markedly above the bottom of the container to avoid measuring errors by inlet of air, or with the bottom of the container that might reduce the actual suction rate. Suctioning of the respective fluid with the respective catheter was executed for exactly 15 s each, measured using a laboratory timer (Timer 38.2010, TFA Dostmann GmbH & Co. KG, Wertheim-Reicholzheim, Germany). After each attempt, the suctioning bag was removed, the scale was tared using the tare function to the weight of an empty suctioning bag, and the full suctioning bag was weighted by means of a calibrated, gauged infant scale (seca 376, seca GmbH & Co. KG, Hamburg, Germany). Suction rates were calculated accordingly. The primary outcome of this study was the weight measured by full grams of mobilized test fluid in a 15 s timeframe.

Statistical analysis

Data was analyzed using Sigmaplot 14 (Systat, San Jose, CA). Data was tested for normal distribution using Shapiro–Wilk test, and equal variances were evaluated with Brown-Forsythe test. If given, a one-way analysis of variance (ANOVA) was used to determine statistically significant differences. Finally, an all pairwise multiple comparison was made using Holm-Sidak method. Figures were created using Prism 9 (GraphPad, San Diego, CA, USA).

Results

The suction rates achieved by the suction catheters and the different viscosities are shown in Table 1, and a visualization of the suction performance can be found in Fig. 2.

Table 1 Mean values and standard deviations (SD) of the different suction catheters and the different viscosities
Fig. 2
figure 2

Visualization of mean values of suction performance per 15 s using the three different suction catheters at fluids of four different viscosities

For water, after testing for normality with Shapiro–Wilk test and equal variances with Brown-Forsythe test, one-way analysis of variance (ANOVA) revealed a statistically significant difference (p < 0.001). An all pairwise multiple comparison (Holm-Sidak method) revealed significant differences between each suction catheter (Table 2).

Table 2 Comparison of the performance of the different suction catheters with water, syrup, honey, and pudding

For syrup, after passed Shapiro–Wilk and Brown-Forsythe testing for normality and equal variances, one-way ANOVA revealed a significant difference. All pairwise comparisons (Holm-Sidak test) showed significant differences between each suction catheter (Table 2).

For honey, after testing for normality (Shapiro–Wilk) and equal variance (Brown-Forsythe), one-way ANOVA showed significant difference. All pairwise comparisons (Holm-Sidak test) showed significant differences between each suction catheter (Table 2).

For pudding, normality was again verified with Shapiro–Wilk and equal variance with Brown-Forsythe tests, and one-way ANOVA showed a significant difference. All pairwise comparisons (Holm-Sidak test) showed significant differences between all suction catheters (Table 2).

Discussion

In this study, we evaluated the suction rates of a standard suction catheter, a large-bore Yankauer suction catheter, and the new large-bore DuCanto suction catheter, when using fluids with four different standardized degrees of viscosity in a technical simulation. The results of this study showed that large-bore catheters were able to achieve significantly higher suction rates in all degrees of viscosity.

Failed airway management due to gastric regurgitation or hemorrhage and subsequent impeded visibility is a dreaded situation in anesthesia, critical care and emergency medicine [12]. Every second of poor visibility and obstructed airway control may have incremental effects in patient health and lead to hypoxemia or death [13]. The importance of a fully functioning, ready to use high performance suction device is further recommended to be incorporated into every workplace that provides elective and emergency airway management [14].

There are several studies and case reports that compared different sizes of the Yankauer with different sizes of other suction devices (including endotracheal tubes connected to a meconium aspirator) [15,16,17,18]. In these studies, a lager diameter of the suction device improved the suction performance, as it is intuitively expected, and be explainable by the law of Hagen and Poiseulle. Interestingly, in this study the difference between the standard catheter and the DuCanto is less than twice the suctioning rate for water, whereas for the pudding like consistence it is approximately three times higher, which can be explained by more turbulent flows in catheters with smaller diameters and increasing viscosity.

It is noteworthy, that a suctioning rate of rounded 74 mg per 15 s with a standard catheter, even an amount of regurgitated stomach contents of only 250 mL will take nearly one minute to be completely removed; this is a very long time in an emergency situation with regurgitation, aspiration and most probably in consequence deteriorating arterial oxygenation [19]. The largest-bore of the catheters we compared would need only 17.5 s for the same amount with the same viscosity.

Further, the longer it takes to remove the regurgitated stomach contents, the higher the likelihood is that massive aspiration occurs, which may result in a severe or even deadly Mendelson syndrome [20]. We deliberately did not use preparations with corpuscular element since these cannot be standardized. One may imagine that every meal has a different consistence, depending on eating behavior, dental status, contact time with digestive enzymes and gastric acidity. However, depending on the latter, it is most likely that normal stomach contents include corpuscular elements, which may further have the potential to deteriorate suction rate or even obstruct suction catheters with lower diameters. So, our values may be considered as maximum rates when no corpuscular elements are present, and can be expected much lower in real-life situations.

The Yankauer catheter is one of the most common suction devices and has become the standard tool used by emergency physicians during endotracheal intubation [15]. In this regard, the new DuCanto Catheter with its 50% better performance in suctioning rate in high viscosity preparations seems to be a very promising alternative.

We have chosen to use four different levels of viscosity to simulate fluids that can typically occur in clinical situation (water as a reference, syrup for stomach acid, honey for airway secretes, and pudding for freshly coagulated blood). Of course, our mixture of xanthan-based thickener and water only gives a surrogate of how these could be, but on the other side, in real life, coagulated blood or stomach acid are not standardized either. This is why we chose to use this product to ensure reproducibility of our experiments. As for the suction unit, we have chosen -600 mbar to represent the maximum suction rate that is available at all work places of a huge university hospital, to simulate the situation that would occur in the clinical setting.

We face several limitations in this study. First, a bench model study is always a limitation by itself, e.g. it can never simulate the stressful situation of a real emergency situation. However, on the other side it is nearly if not totally impossible to study aspirations in prospective controlled clinical trials. Observational studies need large numbers of observed patients and can never be standardized. Thus, we are confident our model is a reproduceable and valid method to assess function of suction devices, such as the new DuCanto Catheter. Secondly, the study staff could not be blinded to the catheter they used. Third, as mentioned above we did not use corpuscular stomach contents, which would be different in humans. Last, we did not test these catheters in an intubation scenario that may be able to demonstrate usefulness of the DuCanto for this standard anesthesia procedures [21,22,23].

In conclusion, suctioning substances of different viscosity from the upper airway the new DuCanto catheter was more effective than the Yankauer catheter, which was more effective compared to a standard suctioning catheter. The relative superiority of the DuCanto was highest in fluids of high viscosity.

Availability of data and materials

The original datasets analysed during the current study are available from the corresponding author on reasonable request.

References

  1. Apfelbaum JL, Hagberg CA, Connis RT, Abdelmalak BB, Agarkar M, Dutton RP, et al. 2022 American Society of Anesthesiologists Practice Guidelines for Management of the Difficult Airway. Anesthesiology. 2022;136:31–81.

    Article  Google Scholar 

  2. Higgs A, McGrath BA, Goddard C, Rangasami J, Suntharalingam G, Gale R, Cook TM, Difficult Airway Society, Intensive Care Society, Faculty of Intensive Care Medicine, Royal College of Anaesthetists. Guidelines for the management of tracheal intubation in critically ill adults. Br J Anaesth. 2018;120:323–52.

    Article  CAS  Google Scholar 

  3. Karamchandani K, Wheelwright J, Yang AL, Westphal ND, Khanna AK, Myatra SN. Emergency Airway Management Outside the Operating Room: Current Evidence and Management Strategies. Anesth Analg. 2021;133:648–62.

    Article  Google Scholar 

  4. Simpson JY. Remarks on the alleged case of death from the action of chloroform. Lancet. 1848;5:175–6.

    Article  Google Scholar 

  5. Sun J, Wei G, Hu L, Liu C, Ding G. Perioperative pulmonary aspiration and regurgitation without aspiration in adults: a retrospective observational study of 166,491 anesthesia records. Ann Palliat Med. 2021;10:4037–46.

    Article  Google Scholar 

  6. Mandell LA, Niederman MS. Aspiration Pneumonia. N Engl J Med. 2019;380:651–63.

    Article  CAS  Google Scholar 

  7. Mendelson CL. The aspiration of stomach contents into the lungs during obstetric anesthesia. Am J Obstet Gynecol. 1946;52:191–205.

    Article  CAS  Google Scholar 

  8. Lossius HM, Roislien J, Lockey DJ. Patient safety in prehospital emergency tracheal intubation: a comprehensive meta-analysis of the intubation success rates of EMS providers. Crit Care. 2012;16:R24.

    Article  Google Scholar 

  9. Wetsch WA, Hinkelbein J. Current national recommendations on rapid sequence induction in Europe. How standardised is the “standard of care”? Eur J Anaesthesiol. 2014;31:443–4.

    Article  Google Scholar 

  10. Pilbery R, Teare MD, Millins M. Soiled airway tracheal intubation and the effectiveness of decontamination by paramedics: a randomised controlled manikin study protocol. Br Paramed J. 2018;3(3):16–22.

    Article  Google Scholar 

  11. ViscoClear Xanthan-based thickener for safe thickening of food & beverages; https://biozoon.de/shop/en/product/seneopro-viscoclear-instant/ (Accessed: 2022–07–14)

  12. Vandenberg JT, Vinson DR. The inadequacies of contemporary oropharyngeal suction. Am J Emerg Med. 1999;17:611–3.

    Article  CAS  Google Scholar 

  13. Sakles JC, Chiu S, Mosier J, Walker C, Stolz U. The importance of first pass success when performing orotracheal intubation in the emergency department. Acad Emerg Med. 2013;20:71–8.

    Article  Google Scholar 

  14. BDA, DGAI. Minimalanforderungen an den anaesthesiologischen Arbeitsplatz. Anaesth Intensivmed 2013;54:1–4

  15. Andreae M, Cox R, Shy B, Wong N, Strayer R. 319 Yankauer outperformed by alternative suction devices in evacuation of simulated emesis. Ann Emerg Med. 2016;68:123.

    Article  Google Scholar 

  16. Kei J, Mebust DP. Comparing the effectiveness of a novel suction set-up using an adult endotracheal tube connected to a meconium aspirator vs. a traditional yankauer suction instrument. J Emerg Me. 2016;9:006.

    Google Scholar 

  17. Vandenberg JT, Rudman NT, Burke TF, Ramos DE. Large-diameter suction tubing significantly improves evacuation time of simulated vomitus. Am J Emerg Med. 1998;16:242–4.

    Article  CAS  Google Scholar 

  18. Weingart SD, Bhagwan SD. A novel set-up to allow suctioning during direct endotracheal and fiberoptic intubation. J Clin Anesth. 2011;23:518–9.

    Article  Google Scholar 

  19. Lipe DN, Lindstrom R, Tauferner D, Mitchell C, Moffett P. Evaluation of Karl Storz CMAC tip device versus traditional airway suction in a cadaver model. West J Emerg Med. 2014;15:548–53.

    Article  Google Scholar 

  20. Wetsch WA, Spöhr FA, Hinkelbein J, Padosch SA. Emergency extracorporeal membrane oxygenation to treat massive aspiration during anaesthesia induction. A case report. Acta Anaesthesiol Scand. 2012;56(6):797–800.

    Article  CAS  Google Scholar 

  21. Pehböck D, Wenzel V, Voelckel W, Jonsson K, Herff H, Mittlböck M, Nagele P. Effects of Preoxygenation on Desaturation Time during Hemorrhagic Shock in Pigs. Anesthesiology. 2010;113:593–9.

    Article  Google Scholar 

  22. Mitterlechner T, Maisch S, Wetsch WA, Herff H, Paal P, Stadlbauer KH, et al. A suction laryngoscope facilitates intubation for physicians with occasional emergency medical service experience–a manikin study with severe simulated airway haemorrhage. Resuscitation. 2009;80:693–5.

    Article  Google Scholar 

  23. Mitterlechner T, Nerbl N, Herff H, Paal P, Moritz M, et al. Effects of a suction laryngoscope in a model with simulated severe airway hemorrhage. Anesth Analg. 2008;106:1505–8.

    Article  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

Open Access funding enabled and organized by Projekt DEAL. This study was solely supported by institutional resources. We acknowledge support for the Article Processing Charge from the DFG (German Research Foundation, 491454339).

Author information

Authors and Affiliations

Authors

Contributions

SF and WAW concepted the study, SF and WAW conducted the experiments. HH and WAW analysed the data. WAW, HE, DCS and HH drafted the manuscript. DS, BWB and DCS helped with data analysis, interpretation and revised the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Wolfgang A. Wetsch.

Ethics declarations

Ethics approval and consent to participate

Not applicable, not required for this type of study by national legislation.

Consent for publication

Not applicable.

Competing interests

There are no competing interests in regard of this study.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Finke, SR., Schroeder, D.C., Ecker, H. et al. Comparing suction rates of novel DuCanto catheter against Yankauer and standard suction catheter using liquids of different viscosity—a technical simulation. BMC Anesthesiol 22, 285 (2022). https://doi.org/10.1186/s12871-022-01830-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12871-022-01830-2

Keywords