A comparison of first-attempt cannulation success of peripheral venous catheter systems with and without wings and injection ports in surgical patients—a randomized trial

Abstract

Background

A peripheral venous catheter (PVC) is the most widely used device for obtaining vascular access, allowing the administration of fluids and medication. Up to 25% of adult patients, and 50% of pediatric patients experience a first-attempt cannulation failure. In addition to patient and clinician characteristics, device features might affect the handling and success rates. The objective of the study was to compare the first-attempt cannulation success rate between PVCs with wings and a port access (Vasofix® Safety, B. Braun, abbreviated hereon in as VS) with those without (Introcan® Safety, B. Braun, abbreviated hereon in as IS) in an anesthesiological cohort.

Methods

An open label, multi-center, randomized trial was performed. First-attempt cannulation success rates were examined, along with relevant patient, clinician, and device characteristics with univariate and multivariate analyses. Information on handling and adherence to use instructions was gathered, and available catheters were assessed for damage.

Results

Two thousand three hundred four patients were included in the intention to treat analysis. First-attempt success rate was significantly higher with winged and ported catheters (VS) than with the non-winged, non-ported design (IS) (87.5% with VS vs. 78.2% with IS; PChi < .001). Operators rated the handling of VS as superior (rating of “good” or “very good: 86.1% VS vs. 20.8% IS, PChi < .001). Reinsertion of the needle into the catheter after partial withdrawal—prior or during the catheterization attempt—was associated with an increased risk of cannulation failure (7.909, CI 5.989–10.443, P < .001 and 23.023, CI 10.372–51.105, P < .001, respectively) and a twofold risk of catheter damage (OR 1.999, CI 1.347–2.967, P = .001).

Conclusions

First-attempt cannulation success of peripheral, ported, winged catheters was higher compared to non-ported, non-winged devices. The handling of the winged and ported design was better rated by the clinicians. Needle reinsertions are related to an increase in rates of catheter damage and cannulation failure.

Trial registration

ClinicalTrials.gov, Identifier: NCT02213965, Date: 12/08/2014.

Peer Review reports

Introduction

A peripheral venous catheter (PVC) is the most widely used device for obtaining continuous vascular access, allowing the administration of fluids and medication to a variety of patients. It is estimated that 30–80% of all hospitalized patients receive a PVC during their stay [1,2,3,4]. For procedural and safety reasons, vascular access in the surgical settings is virtually indispensable. Along with attachment of monitoring devices, the insertion of a PVC usually occurs immediately upon arrival in the induction or operating room, allowing for the administration of fluids, induction of anesthesia and management of potential complications.

Since establishing venous access occurs while the patient is awake, and usually precedes administration of any anesthetic agents, PVC placement can be an uncomfortable and painful procedure. Repeated and unsuccessful attempts to place a PVC can be stressful for both patient and clinician, causing localized pain and swelling, creating a portal of entry for micro-organisms, and forcing clinicians to seek alternative sites, where cannulation may be more difficult or dangerous [5,6,7]. Therefore, a successful PVC placement in the first attempt is always desirable. Reports indicate that up to a quarter of adult patients, and half of pediatric patients experience a first-attempt cannulation failure [8,9,10].

Cannulation success rates can be influenced by the venous conditions and body-mass index (BMI) of the patient, as well as the level of experience of the healthcare provider [11], but also the type of PVC used may impact success rates. There are several types of PVCs available in the European market, and features such as size, bevel type, and the presence of notches, wings and ports might affect the handling of these devices [12]. Ported and winged catheters are widespread in German hospitals, but there is no evidence that they are superior to non-ported catheters, especially in relation to first-attempt cannulation success. The objective of this study is to compare the rate of first-attempt placement success between two PVC designs, namely a ported, winged catheter (Vasofix® Safety, B. Braun Melsungen AG, Germany, abbreviated hereon in as VS) and a non-ported, non-winged catheter (Introcan® Safety, B. Braun Melsungen AG, Germany, abbreviated hereon in as IS) (see Fig. 1).

Fig. 1
figure 1

Introcan® Safety (IS) and Vasofix® Safety (VS), B. Braun Melsungen AG, Germany (images retrieved from https://www.bbraun.de/de/products/b0/vasofix-safety.html and https://www.bbraun.de/de/products/b/introcan-safety.html, adapted and printed with permission from B. Braun Melsungen)

Methods

This open-label, multicenter study randomized patients to receive one of two distinct types of PVCs prior to elective surgery. This sponsor-initiated study compared two CE-marked, Class II medical devices in accordance with their intended purpose and use instructions [13, 14]. Data was collected between December 2014 and March 2016 by the anesthesiology departments of two German tertiary care university hospitals, namely the Charité – Universitätsmedizin Berlin with its three separate sites, Campus Charité Mitte (CCM), Campus Virchow Klinikum (CVK), and Campus Benjamin Franklin (CBF), and the Universitätsklinikum Tübingen. In accordance with inclusion and exclusion criteria (see Supplemental Fig. 1), adult and pediatric patients undergoing elective surgery were recruited by the anesthesiology staff during the routine pre-surgical assessment, and written informed consent was obtained from all participants or their proxies. Peripheral venous cannulation was required for surgery, so that patients were not subjected to any additional procedure or distress in the context of this investigation. All catheters were placed immediately prior to surgery, at the induction or operating room. The study was approved by the ethic committees of both participating universities (Berlin: EA2/104/14; Tübingen: 494/2014BO2), as well as by their respective data protection officers. The study was conducted in accordance with the Declaration of Helsinki and its amendments and was registered at ClinicalTrials.gov (NCT 02213965, 12/08/2014).

After providing informed consent, patients were randomly selected to receive a ported, winged fluorethylene propylene catheter (VS) or a non-ported, non-winged polyurethane catheter (IS). Study personnel prepared and conducted randomization via sealed envelopes, stratified by study site in a 1:1 ratio. General patient data was then collected, including age, gender, BMI, American Society of Anesthesiologists Physical Status (ASA-PS) score, as well as venous status (subjective assessment: good, moderate, or poor). Whenever possible, study personnel accompanied patients to the operating theater to ensure that venous cannulation took place with the designated catheter and that documentation was completed appropriately. The staff member placing the PVC (operator) completed a paper-based case report form (CRF), which included questions regarding the number and sites of cannulation attempts, subjective reasons for failure, adherence to use instructions (assessed by documenting needle reinsertion into the catheter, i.e. the practice of moving the needle stylet in the cannula prior to puncture or during the catheterization attempt; the manufacturer’s instructions for use (IFU) explicitly warn against this practice, and the variable was included as a marker for misuse), catheter handling information (e.g. 6-point Likert scale, blood spillage, use of ports), as well as their professional function and level of experience with venous catheterization. In case of a failed cannulation attempt, the catheter was collected and sent to the Bergische Universität Wuppertal, Faculty of Mechanical and Safety Engineering for a macro- and microscopic examination for signs of damage. Observation continued until patient discharge from the operating or recovery room, whereas any adverse events (e.g. catheter occlusion, displacement, or extravasation) were noted and the respective catheters collected. Used catheters were also collected following decannulation of ambulatory patients. No treatment changes occurred, all standard operating procedures remained in place throughout the investigation, and the attending physician had ultimate authority in all treatment decisions.

The primary outcome was successful cannulation at first attempt. Secondary outcomes included overall number of cannulation attempts, subjective reasons for cannulation failure, operators’ evaluation of catheter handling, frequency of needle reinsertion into the catheter, rate of PVC damage, and adverse events.

The sample size calculation, which was aimed for a power of 80%, an α of 0.05, and a 5% delta for a 75% first-attempt cannulation success for the control group (VS), yielded 1134 patients per trial arm (n = 2268). Additionally, in order to retain the prognostic balance provided by the randomization process [15], all data were analyzed according to the intention-to-treat (ITT) principle. Metric data was not normally distributed, so that data was reported as median and interquartile ranges and differences among groups were evaluated using the Wilcoxon-Mann–Whitney U test. Ordinal and nominal data were reported using frequencies and percentages, and groups were compared using Chi-Square Test, or Fisher’s Exact Test when small cell exceptions were present. Missing data was treated as such, with no replacement by estimates. Univariate and multiple logistic regression analyses was conducted to determine the most relevant factors for cannulation success at the first attempt, adjusting for possible confounders. The stepwise regression model was based on the available literature, which was supplemented by input from the stakeholders, who were familiar with the complaints and expectations from users in the field. The included variables depict patient (age, gender, BMI, cannulation site and venous condition) [16], operator (function and experience) [17], and product (catheter type and size) [17, 18] characteristics that might affect the cannulation success rates. Additionally, as part of an exploratory analysis, rate of needle reinsertion into catheters (any, before, or during the catheterization attempt) was added among operator variables as a misuse indicator, along with a detailed risk assessment for cannulation failure and catheter damage. Statistical analysis was performed with SPSS (Version 27.0, Armonk, NY: IBM Corp., USA).

Results

Among the four study sites, 2304 patients were included in the ITT analysis, as shown in Supplemental Fig. 1 (Consort Diagram). Of these patients, 1133 received an IS catheter and 1171 received a VS catheter. Overall, there were no statistically relevant differences among patients in terms of age, gender, BMI, ASA score, venous conditions, or cannulation site. However, significant differences were observed in terms of catheter size (see Table 1). Patient characteristics among sites are provided in Supplemental Tables 1 and 2.

Table 1 All data shown as frequencies and percentages. BMI body-mass index, ASA American Society of Anesthesiologists. Missing data is treated as such
Table 2 All data shown as frequencies and percentages. ITT intention to treat, BMI body-mass index. Missing data is treated as such

Overall, the first-attempt cannulation success was achieved in 82.9% of cases and was higher with VS (87.5%) than with IS (78.2%) (P < 0.001). This significance was confirmed in a per protocol group analysis (87.5% vs. 80.5% success rate; n = 2267, P < 0.001), as well as in a sub-analysis including only patients ≥ 18 years of age (88.2% vs. 78.0% success rate; n = 2202, P < 0.001). Relevant factors influencing first-attempt success rates are shown in Table 2. While the success rates with VS were significantly higher in patients with at least 45 years of age, differences ceased to be significant for those aged between 18 and 44, and observed success rates did not differ significantly for patients below the age of 18. There were no differences in success rates among patients with a BMI < 18.5 kg/m2, although VS was superior in all other BMI categories. VS was also superior in good and moderate vein conditions, while the success rate was not significantly different in patients with poor venous status. The majority of venous catheters were placed on the back of the hand (78.7%), where success rates with VS were significantly higher than at other cannulation sites. Here, a success rate of 89.8% was achieved with VS, compared to 78.9% with IS. No statistical differences were detected between success rates for cannulas placed in the forearm or antecubital fossa. Compared to IS, first-attempt success rates were significantly higher with VS for nurses and physicians alike (P < 0.001), although this difference could not be observed among other groups (i.e. medical students, interns). There were differences among centers regarding operator function. While physicians performed the majority of cannulations in CBF and UCT (60.7% and 54.6%, respectively), nurses performed the majority of cannulations in CCM and CVK (83.7% and 61.4%, respectively). 72.9% of operators had over 3 years of work experience, placing an average of 17 (± 9) catheters per week. Here, nurses placed significantly higher number of catheters per week in comparison to physicians (P < 0.001), and only 7% reported having less than 3 years of experience, compared to 50% of physicians. The multivariate analysis (n = 1157) identified venous conditions, BMI, operator function and experience, adherence to IFU, catheter size and catheter type as the most significant determinants of cannulation success at the first attempt (see Table 3). In a regression model without misuse markers (n = 1956, Supplemental Table 3), operator function ceased to be significant.

Table 3 Multivariate analysis; VS Vasofix® Safety, IS Introcan® Safety; adult participants only; All data analyzed as categorical variables

The most commonly reported reasons for cannulation failure were poor vein status (48.8% of respondents, n = 313), poor handling (27.8% or respondents, n = 127; 90.8% IS vs. 9.2% VS, P < 0.001), and blunt cannulas (21.4% of respondents, n = 67; 76.1% IS vs. 23.9% VS, P = 0.73). Handling of VS was rated by operators as “good” or “very good” in 86.1% of cases, while only 20.8% rated IS as such (P < 0.001) (see Table 4). Accordingly, VS operators said that the wings and ports influenced catheter placement in 25.7% of cases. The port of the VS was used in 65.0% of cases for flushing, and/or drug administration.

Table 4 Handling characteristics. Subjective assessment by corresponding operators

Data on misuse parameters was limited with a 58.9% response rate. Partial needle withdrawal and reinsertion into the cannula prior to the catheterization attempt (to ensure mobility of the needle stylet) was reported in 318 of 1355 documented cases (23.5%), while the same maneuver during the catheterization attempt (i.e. partial needle withdrawal to test whether a flashback in the cannula occurs, followed by needle reinsertion and a further catheterization attempt) was reported in 61 of 1348 cases (4.5%). These maneuvers were associated with an 8- and 23-fold risk of cannulation failure, respectively (see Table 5; a detailed analysis by catheter is available in Supplemental Table 4). Of the 461 catheters returned for laboratory evaluation, 32.5% displayed signs of damage, and needle reinsertion prior or during the catheterization attempt were associated with a twofold increase in the risk of catheter damage (see Table 5; a detailed analysis by catheter is available in Supplemental Table 5).

Table 5 Catheter misuse in relation to risk of cannulation failure and catheter damage. Catheter misuse refers to needle movement, the practice of moving the needle in the cannula prior to puncture or during the catheterization attempt (the manufacturer’s use instructions explicitly warn against such maneuvers). Catheter damage refers to abnormalities in catheter structure, macroscopic or microscopic, such as cuts, tears, compressions or loss of material

Following an unsuccessful first attempt, two further attempts were usually required to secure venous access (75.9%; median 3, IQR [3, 3]), regardless of the type of cannula used. Overall, adverse events took place in 77 (3.3%) cases, whereas 65 of these events involved the IS cannula (P < 0.001). No serious adverse events were observed and only one needle stick injury took place during the trial (with a VS PVC), which took place prior to patient (or blood) contact.

Discussion

In general, success rates were significantly higher with the ported, winged catheter VS than with the non-ported, non-winged IS. With 83% in the ITT and 84% in the per protocol (PP) populations, the success rates observed were higher than the 73–74% success rates reported in emergency departments [9, 11], but failed to reach the desired 90% threshold [19]. The results are comparable to recent studies including surgical patients, which report a 79–85% first-attempt success rate [12].

Key factors affecting first-attempt success rates of peripheral cannulation were described by Carr et al. as patient, operator, and product variables [20]. Patient factors included age, BMI, and venous status, while operator characteristics included function and experience. Product variables included catheter size, but this study could demonstrate that handling and differences in product design can also affect success rates. For the surgical cohort, the most significant factors in the multifactorial analysis were venous status, catheter type, higher BMI, and operator experience. Additionally, adherence to IFU (i.e. no needle reinsertion) was shown to be a significant factor, although the analysis encompassed only a subpopulation of the cohort. The operator-reported reasons for cannulation failure were most often poor venous status, poor handling of the PVC (primarily IS), and blunt needles.

Overall, the VS design was shown to be more favorable. This PVC design, which has wings and an injection port, appears to facilitate catheter placement and fixation, as indicated by the consistently higher successful cannulation rates, improved backflash visibility, as well as lower rates of spillage and soiled dressings. The majority of operators (86.1%) rated the VS PVC as “good” or “very good”, while IS received such ratings from only 20% of operators. The injection port also appears to be well accepted, being utilized during the surgical procedure for flushing and/or drug administration in approximately 2/3 of cases, despite the availability of 3-way stopcocks. It must be noted, however, that such ports have been associated with a greater risk of infection, likely due to the inability to properly decontaminate the port area [21].

Interestingly, the superiority of the VS PVC was diminished with younger age and lower BMI. The slimmer and smaller IS design seems to be more suitable for children and patients with BMI < 18.5 kg/m2, although these differences remained statistically insignificant, likely due to the smaller size of these subgroups. Nevertheless, IS PVCs were associated with 65 of the 77 (84.4%) catheter-related adverse events observed in this investigation. These included primarily extravasate infusions, dislocations, and occlusions, factors that are rather related to care and maintenance of the PVCs, and not the placement procedure itself.

The most commonly chosen site for cannulation in this surgical collective was the backside of the hand, which usually allows ease of access for anesthesiologists throughout the surgical procedure. Furthermore, should a cannulation failure occur, proximal alternative sites remain available. This remains the primary cannulation site despite evidence that placement in the forearm reduces catheter-associated complication rates [22].

Physicians and nurses achieved similar success rates. Nurse involvement increased when moderate or poor vein conditions were reported, whereas nurses were also significantly more experienced and placed a higher number of catheters per week in comparison to physicians, suggesting an advantage in terms of routine and experience. More experienced physicians are underrepresented, likely because they frequently supervise longer and more complex operations (often one per day), whereas less experienced colleagues usually accompany several short and simple procedures during a workday. When taking experience into consideration, there were no differences in success (or misuse) between physicians and nurses. Interestingly, the best success rates were achieved by students and interns, despite no differences in the subjective assessment of the patients’ venous status. This is possibly due to less time pressure and a selected patient group, but it must be noted that in the larger, more representative logistic regression (Supplemental Table 3), no significant differences were observed.

The PVCs were not always used as intended. The pervasive custom of moving the needle inside the plastic cannula prior to puncture took place in nearly a quarter of cases, being associated with an increase in the risk of failure from 16 to 64%. Likewise, partially withdrawing the needle through the cannula after a puncture, then reinserting it in a rescue attempt to redirect the cannula was seen in nearly 5% of cases, with an 88.5% likelihood of failure. The IFU specifically warn users not to reinsert needles through the cannula, and there are reports of intravenous catheter fractures and embolisms [23]. In this investigation, such maneuvers were related to a twofold increase in the rate of catheter damage in the laboratory analysis. While VS showed higher rates of damage, with or without needle reinsertion, IS appears to be more susceptible to damage by misuse, possibly due to the polyurethane material (see Supplemental Table 5). This information must be emphasized in training sessions in an effort to reduce such practices. It is also worthy of note that 25.7% of PVCs that were used properly still showed signs of damage.

It is important to note that an increasing number of tools have been developed for predicting difficult venous access, both for pediatric [24,25,26] and adult patients [19, 27,28,29]. Additionally, there has been a marked increase in the availability and acceptance of ultrasound-guided assessment and cannulation [30, 31], sparking the formation of vascular access teams [32]. The European Society of Anesthesiologist published guidelines recommending the early identification of difficult venous access through validated tools, as well as the use of ultrasound [33]. The implementation of such tools has the potential to improve patient care by adapting clinical processes and allowing an appropriate allocation of personnel and materials, as well as provide an objective basis for the comparison in clinical studies.

Limitations

Several limitations must be taken into account. Although an open-label investigation was unavoidable, there was an attempt to minimize bias through implementation of a randomization process and the use of CRFs. Nevertheless, an observation bias cannot be ruled out, and nor can a performance bias (involvement of more experienced colleagues by particularly difficult venous status). Although patients were accompanied by study personnel whenever possible, self-reported data from unsupervised operators may be an additional source of bias. The venous status of patients was subjective and may vary considerably, depending on the operator’s experience. Although the adult version of the Difficult Intravenous Access (DIVA) scale [27] was not available at the beginning of this trial, the lack of an objective scoring system is a clear limitation of this study. No identifiable information was collected from the operators, and no data is available regarding the number of operators involved nor the frequency of their participation. No follow-up attempt was made after discharge from the recovery room, and the short duration of observation may have affected the reported rate of complications. There might be additional factors affecting cannulation success that were not considered in this study, particularly in the multivariate analyses. Although both products were newly introduced, it is important to note that winged and ported catheters are far more common in Germany, so that operators were likely to be more familiar with this design type. Additionally, due to the high number of missing values, data on needle reinsertion and damage must be regarded with caution. Data generalizability is limited, as the study was conducted exclusively in a university setting.

Conclusions

First-attempt cannulation success was more likely with ported, winged VS catheters compared to non-ported, non-winged IS devices. The larger VS design allowed for improved handling and fixation, although this superiority could not be observed in younger patients or those with low BMI. The overall rate of successful puncture at first attempt in the operating room was 83%. Among clinicians, nurses were more experienced and placed the majority of catheters. The backside of the hand was the preferred site, and the most commonly reported reasons for failure were poor venous status and poor handling of the catheter. This trial identified the reinsertion of the needle stylet into the cannula, prior or during the cannulation attempt, as novel clinician-related markers that were associated with higher incidences of catheter damage and cannulation failure. This information must be highlighted in training sessions to limit these dangerous practices and improve patient safety. Standardized tools for the identification of difficult venous access are increasingly common, and the implementation of ultrasound techniques and dedicated vascular access teams have the potential to improve care and reduce catheter-related complications.

Availability of data and materials

The datasets analyzed during the current study are not publicly available due to constrains imposed in the consent forms, but an anonymized version is available from the corresponding author on reasonable request.

Abbreviations

PVC:

Peripheral venous catheter

BMI:

Body-mass index

VS:

Vasofix® Safety

IS:

Introcan® Safety

CBF:

Charité Campus Benjamin Franklin

CVK:

Charité Campus Virchow Klinikum

CCM:

Charité Campus Charité Mitte

UCT:

Universitätsklinikum Tübingen

CRF:

Case report form

ITT:

Intention-to-treat

PP:

Per protocol

ASA:

American Society of Anesthesiologists

OR:

Odds Ratio

CI:

Confidence Interval

References

  1. Waitt C, Waitt P, Pirmohamed M. Intravenous therapy. Postgrad Med J. 2004;80(939):1–6.

    Article  CAS  Google Scholar 

  2. Malach T, Jerassy Z, Rudensky B, Schlesinger Y, Broide E, Olsha O, et al. Prospective surveillance of phlebitis associated with peripheral intravenous catheters. Am J Infect Control. 2006;34(5):308–12.

    Article  Google Scholar 

  3. Pujol M, Hornero A, Saballs M, Argerich MJ, Verdaguer R, Cisnal M, et al. Clinical epidemiology and outcomes of peripheral venous catheter-related bloodstream infections at a university-affiliated hospital. J Hosp Infect. 2007;67(1):22–9.

    Article  CAS  Google Scholar 

  4. Alexandrou E, Ray-Barruel G, Carr PJ, Frost S, Inwood S, Higgins N, et al. International prevalence of the use of peripheral intravenous catheters: Prevalence of the Use of PIVCs. J Hosp Med. 2015;10(8):530–3.

    Article  Google Scholar 

  5. Rickard CM, Marsh N, Webster J, Playford EG, McGrail MR, Larsen E, et al. Securing All intraVenous devices Effectively in hospitalised patients–the SAVE trial: study protocol for a multicentre randomised controlled trial. BMJ Open. 2015;5(9):e008689.

    Article  Google Scholar 

  6. Moayedi S, Witting M, Hirshon JM, George N, Burke A, Schenkel S. Prospective, randomized controlled comparison of a flash-tip catheter and a traditional intravenous catheter in an urban emergency department. J Vasc Access. 2018;19(4):387–91.

    Article  Google Scholar 

  7. Castillo MI, Larsen E, Cooke M, Marsh NM, Wallis MC, Finucane J, et al. Integrated versus nOn-integrated Peripheral inTravenous catheter. Which is the most effective systeM for peripheral intravenoUs catheter Management? (The OPTIMUM study): a randomised controlled trial protocol. BMJ Open. 2018;8(5):e019916.

    Article  Google Scholar 

  8. Sabri A, Szalas J, Holmes KS, Labib L, Mussivand T. Failed attempts and improvement strategies in peripheral intravenous catheterization. Biomed Mater Eng. 2013;23(1–2):93–108.

    PubMed  Google Scholar 

  9. Lapostolle F, Catineau J, Garrigue B, Monmarteau V, Houssaye T, Vecci I, et al. Prospective evaluation of peripheral venous access difficulty in emergency care. Intensive Care Med. 2007;33(8):1452–7.

    Article  Google Scholar 

  10. Hess HA. A biomedical device to improve pediatric vascular access success. Pediatr Nurs. 2010;36(5):259–63.

    PubMed  Google Scholar 

  11. Carr PJ, Rippey JCR, Cooke ML, Trevenen ML, Higgins NS, Foale AS, et al. Factors associated with peripheral intravenous cannulation first-time insertion success in the emergency department. A multicentre prospective cohort analysis of patient, clinician and product characteristics. BMJ Open [Internet]. 2019 Apr 2 [cited 2020 Mar 22];9(4). https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6500093/Available from

  12. van Loon FH, Timmerman R, den Brok GP, Korsten EH, Dierick-van Daele AT, Bouwman AR. The impact of a notched peripheral intravenous catheter on the first attempt success rate in hospitalized adults: Block-randomized trial. J Vasc Access. 2021;2:1129729821990217.

    Google Scholar 

  13. Introcan Safety® IV Catheter Straight [Internet]. [cited 2021 Sep 14]. https://www.bbraunusa.com/en/products/b/introcan-safety-ivcatheterstraight.htmlAvailable from

  14. Vasofix® Safety [Internet]. [cited 2021 Sep 14]. https://www.bbraun.de/de/products/b0/vasofix-safety.htmlAvailable from

  15. McCoy CE. Understanding the Intention-to-treat Principle in Randomized Controlled Trials. West J Emerg Med. 2017;18(6):1075–8.

    Article  Google Scholar 

  16. Sebbane M, Claret P-G, Lefebvre S, Mercier G, Rubenovitch J, Jreige R, et al. Predicting peripheral venous access difficulty in the emergency department using body mass index and a clinical evaluation of venous accessibility. J Emerg Med. 2013;44(2):299–305.

    Article  Google Scholar 

  17. Jacobson AF, Winslow EH. Variables influencing intravenous catheter insertion difficulty and failure: an analysis of 339 intravenous catheter insertions. Heart Lung. 2005;34(5):345–59.

    Article  Google Scholar 

  18. Fields JM, Piela NE, Au AK, Ku BS. Risk factors associated with difficult venous access in adult ED patients. Am J Emerg Med. 2014;32(10):1179–82.

    Article  Google Scholar 

  19. van Loon FHJ, van Hooff LWE, de Boer HD, Koopman SSHA, Buise MP, Korsten HHM, et al. The Modified A-DIVA Scale as a Predictive Tool for Prospective Identification of Adult Patients at Risk of a Difficult Intravenous Access: A Multicenter Validation Study. J Clin Med. 2019;8(2):E144.

    Article  Google Scholar 

  20. Carr PJ, Higgins NS, Cooke ML, Rippey J, Rickard CM. Tools, Clinical Prediction Rules, and Algorithms for the Insertion of Peripheral Intravenous Catheters in Adult Hospitalized Patients: A Systematic Scoping Review of Literature. J Hosp Med. 2017;12(10):851–8.

    Article  Google Scholar 

  21. Hadaway L. Short peripheral intravenous catheters and infections. J Infus Nurs. 2012;35(4):230–40.

    Article  Google Scholar 

  22. Wallis MC, McGrail M, Webster J, Marsh N, Gowardman J, Playford EG, et al. Risk factors for peripheral intravenous catheter failure: a multivariate analysis of data from a randomized controlled trial. Infect Control Hosp Epidemiol. 2014;35(1):63–8.

    Article  Google Scholar 

  23. Glassberg E, Lending G, Abbou B, Lipsky AM. Something’s Missing: Peripheral Intravenous Catheter Fracture. The Journal of the American Board of Family Medicine. 2013;26(6):805–6.

    Article  Google Scholar 

  24. Yen K, Riegert A, Gorelick MH. Derivation of the DIVA score: a clinical prediction rule for the identification of children with difficult intravenous access. Pediatr Emerg Care. 2008;24(3):143–7.

    Article  Google Scholar 

  25. Riker MW, Kennedy C, Winfrey BS, Yen K, Dowd MD. Validation and refinement of the difficult intravenous access score: a clinical prediction rule for identifying children with difficult intravenous access. Acad Emerg Med. 2011;18(11):1129–34.

    Article  Google Scholar 

  26. Girotto C, Arpone M, Frigo AC, Micheletto M, Mazza A, Da Dalt L, et al. External validation of the DIVA and DIVA3 clinical predictive rules to identify difficult intravenous access in paediatric patients. Emerg Med J. 2020;37(12):762–7.

    PubMed  Google Scholar 

  27. van Loon FHJ, Puijn LAPM, Houterman S, Bouwman ARA. Development of the A-DIVA Scale: A Clinical Predictive Scale to Identify Difficult Intravenous Access in Adult Patients Based on Clinical Observations. Medicine (Baltimore). 2016;95(16):e3428.

    Article  Google Scholar 

  28. Angles E, Robin F, Moal B, Roy M, Sesay M, Ouattara A, et al. Pre-operative peripheral intravenous cannula insertion failure at the first attempt in adults: Development of the VENSCORE predictive scale and identification of risk factors. J Clin Anesth. 2021;75:110435.

    Article  Google Scholar 

  29. Civetta G, Cortesi S, Mancardi M, De Pirro A, Vischio M, Mazzocchi M, et al. EA-DIVA score (Enhanced Adult DIVA score): A new scale to predict difficult preoperative venous cannulation in adult surgical patients. J Vasc Access. 2019;20(3):281–9.

    Article  Google Scholar 

  30. O’Reilly-Shah VN, Franz A, Groenewald CB, Collins M, Patak LS. Training the Trainers in Ultrasound-guided Access to Improve Peripheral Intravenous Catheter Placement among Children Presenting for Anesthesia. Pediatr Qual Saf. 2021;6(3):e406.

    Article  Google Scholar 

  31. Gras S, Roy-Gash F, Bruneau B, Salvi N, Colas A-E, Skhiri A, et al. Reducing the time to successful intravenous cannulation in anaesthetised children with poor vein visibility using a near-infrared device: A randomised multicentre trial. Eur J Anaesthesiol. 2021;38(8):888–94.

    PubMed  Google Scholar 

  32. Bell JA, Spencer TR. Implementing an emergency department vascular access team: A quality review of training, competency, and outcomes. J Vasc Access. 2021;22(1):81–9.

    Article  Google Scholar 

  33. Lamperti M, Biasucci DG, Disma N, Pittiruti M, Breschan C, Vailati D, et al. European Society of Anaesthesiology guidelines on peri-operative use of ultrasound-guided for vascular access (PERSEUS vascular access). Eur J Anaesthesiol. 2020;37(5):344–76.

    Article  Google Scholar 

Download references

Acknowledgements

The authors thank Dr. Kathrin Scholz for the clinical monitoring and Prof. Dr.- Ing. Andreas Wittmann and his team from the Bergische Universität Wuppertal, Fakultät für Maschinenbau und Sicherheitstechnik, Fachbereich Technischer Infektionsschutz for the catheter damage assessment. The authors also thank the study nurses Beatrix Laetsch, Susanne Sonnemann and Tuba Gülmez, as well as Dr. Jörn Kiselev for the critical review of the manuscript.

Funding

Open Access funding enabled and organized by Projekt DEAL. This study was sponsored by B.Braun Melsungen AG under the direction of Dr. Heike Fries. The sponsorship did not influence the analysis or reporting of study outcomes.

Author information

Authors and Affiliations

Authors

Contributions

RM: writing-original draft, writing-review & editing, validation, formal analysis; KS: conceptualization, investigation, validation, supervision, writing-review & editing, project administration; TN: investigation, validation, supervision, formal analysis, writing-review & editing; JK: validation, formal analysis; UF: resources, supervision, project administration; PH: investigation, data curation; PR: resources, supervision, project administration; EF: investigation, data curation; WB: conceptualization, methodology, supervision; CS: conceptualization, methodology, resources, project administration, funding acquisition. All authors have read and approved the final version of the manuscript.

Corresponding author

Correspondence to Claudia Spies.

Ethics declarations

Ethics approval and consent to participate

Written informed consent was obtained from all participants or their proxies. The study was approved by the ethic committees of the Charité – Universitätsmedizin Berlin (EA2/104/14) and the Universitätsklinikum Tübingen (494/2014BO2), as well as by their respective data protection officers. The study was conducted in accordance with the Declaration of Helsinki and its amendments and was registered at ClinicalTrials.gov (NCT 02213965).

Consent for publication

B. Braun has granted permission to publish Fig. 1.

Competing interests

CS, UF, TN and PR have received material institutional support from B.Braun Melsungen AG for the conduction of this trial. The sponsorship did not influence the analysis or reporting of study outcomes. All other authors have no competing interests to report.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1:

 Supplemental Figure 1. Consort diagram, flowchart; *For one study site (Universitätsklinik Tübingen), only information about completed case report forms are available; # Other reasons include screening failure, urgent operations, and missing documentation; ## Other reasons include missing/incomplete data and drop-out participants.

Additional file 2:

 Supplemental Table 1. Patient characteristics by center. All data shown as frequencies and percentages. ASA  American Society of Anesthesiologists, CBF Campus Benjamin Franklin, CVK Campus Virchow Klinikum, CCM Campus Charité Mitte; UCT Universitätsklinikum Tübingen; P-values represent Chi-Square Test, or Fisher’s Exact Test when small cell exceptions were present; missing data is treated as such.

Additional file 3:

 Supplemental Table 2. Patient characteristics by center (continued). All data shown as frequencies and percentages. CBF Campus Benjamin Franklin, CVK Campus Virchow Klinikum, CCM Campus Charité Mitte, UCT Universitätsklinikum Tübingen. Missing data is treated as such. Vein status is a subjective variable assigned by the corresponding operator; P-values represent Chi-Square Test, or Fisher’s Exact Test when small cell exceptions were present; missing data is treated as such.

Additional file 4:

 Supplemental Table 3. Multivariate analysis without misuse markers; VS: Vasofix® Safety, IS: Introcan® Safety; adult participants only; All data analyzed as categorical variables.

Additional file 5:

 Supplemental Table 4. Catheter misuse in relation to risk of cannulation failure, detailed by type of catheter. Catheter misuse refers to needle movement, the practice of moving the needle in the cannula prior to puncture or during the catheterization attempt (the manufacturer’s use instructions explicitly warn against such maneuvers).

Additional file 6:

 Supplemental Table 5. Catheter misuse in relation to risk of catheter damage, detailed by type of catheter. Catheter misuse refers to needle movement, the practice of moving the needle in the cannula prior to puncture or during the catheterization attempt (the manufacturer’s use instructions explicitly warn against such maneuvers). Catheter damage refers to abnormalities in catheter structure, macroscopic or microscopic, such as cuts, tears, compressions or loss of material.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Mörgeli, R., Schmidt, K., Neumann, T. et al. A comparison of first-attempt cannulation success of peripheral venous catheter systems with and without wings and injection ports in surgical patients—a randomized trial. BMC Anesthesiol 22, 88 (2022). https://doi.org/10.1186/s12871-022-01631-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12871-022-01631-7

Keywords