Implementation of the TaperGuard™ endotracheal tube in an unselected surgical population to reduce postoperative pneumonia

Abstract

Background

Endotracheal tube (ETT) designs to decrease the risk of ventilator associated pneumonia (VAP) include supraglottic suctioning, and/or modifications of the cuff shape. The TaperGuard™ ETT has a tapered, polyvinylchloride cuff designed to reduce microaspiration around channels that form with a standard barrel-shaped cuff. We compared risk of postoperative pneumonia using the TaperGuard™ ETT and the standard ETT in surgical patients requiring general anesthesia with endotracheal intubation.

Methods

We used an interrupted time-series design to compare endotracheal intubation using the TaperGuard™ ETT (intervention cohort), and a historic cohort using the standard ETT (baseline cohort), among surgical patients requiring hospital admission. We compared the incidence of postoperative pneumonia in the intervention and baseline cohorts. Data were collected from the electronic health record and linked to patient-level data from National Surgical Quality Improvement Project. Additionally, we performed secondary analyses in a subgroup of patients at high risk of postoperative pneumonia.

Results

15,388 subjects were included; 6351 in the intervention cohort and 9037 in the baseline cohort. There was no significant difference in the incidence of postoperative pneumonia between the intervention cohort (1.62%) and the baseline cohort (1.79%). The unadjusted odds ratio (OR) of postoperative pneumonia was 0.90 (95% CI: 0.70, 1.16; p = 0.423) and the OR adjusted for patient characteristics and potential confounders was 0.90 (95% CI: 0.69, 1.19; p = 0.469), comparing the intervention and baseline cohorts. There was no a priori selected subgroup of patients for whom the use of the TaperGuard™ ETT was associated with decreased odds of postoperative pneumonia relative to the standard ETT. Hospital mortality was higher in the intervention cohort (1.5%) compared with the baseline cohort (1.0%; OR 1.46, 95% CI: 1.09, 1.95; p = 0.010).

Conclusions

The broad implementation of the use of the TaperGuard™ ETT for intubation of surgical patients was not associated with a reduction in the risk of postoperative pneumonia. In the setting of a low underlying postoperative pneumonia risk and the use of recommended preventative VAP bundles, further risk reduction may not be achievable by simply modifying the ETT cuff design in unselected or high-risk populations undergoing inpatient surgery.

Trial registration

ClinicalTrials.gov, ID NCT02450929.

Peer Review reports

Background

In high-risk surgical patient populations, the incidence of postoperative pneumonia has been reported to be as high as 21% [1,2,3]. Recent national efforts to decrease healthcare associated infections have resulted in the development of a bundle of interventions that have effectively decreased the incidence of VAP in high-risk patient populations [4].

Strategies to minimize microaspiration of supraglottic secretions include modification of the endotracheal tube (ETT) configuration and the utilization of supraglottic suctioning [5,6,7]. The TaperGuard™ ETT (Covidien, Boulder, CO) is designed to prevent microaspiration around channels that otherwise form with a barrel-shaped cuff (standard ETT) and may thereby reduce the incidence of VAP. This device has undergone multiple independent and manufacturer sponsored laboratory and clinical trials. Laboratory evaluations have demonstrated decreased passage of fluid or dye around the ETT compared with conventional barrel-shaped cuffs, but not in all experimental conditions [8,9,10,11,12,13,14].

Clinical evaluations in relatively small studies of unselected patient populations have failed to demonstrate a decrease in the incidence of postoperative pneumonia [15, 16]. Because the incidence of postoperative pneumonia in an otherwise unselected patient population is much lower than in those with prolonged ventilation or other risk factors, it is possible that prior trials were too small to detect a difference, or that selection of subgroups were suboptimal.

As part of a quality improvement initiative, our institution implemented a change of ETTs from the standard barrel-shaped design to the TaperGuard™ ETT for all surgical patients. The present study compares the incidence of postoperative pneumonia, before and after the implementation of the TaperGuard™ ETT in a large, unselected inpatient population undergoing surgery with general anesthesia and in several high-risk subgroups, to determine the efficacy of this device in reducing postoperative pneumonia.

Methods

Study design

This cohort study was conducted in the setting of the implementation of a perioperative quality improvement initiative within the Department of Anesthesiology and Perioperative Medicine at Oregon Health & Science University (OHSU) Hospital. On December 1, 2012, OHSU instituted a practice change to transition from ETTs with a barrel-shaped cuff design to the TaperGuard™ ETT for all surgical patients. We used an interrupted time-series to compare two cohorts of patients undergoing inpatient surgery with general anesthesia and the placement of an ETT, during a baseline period with the use of standard ETT and an intervention period with the use of the TaperGuard™ ETT. The baseline cohort included patients who had surgery between April 1, 2011 and November 30, 2012; the intervention cohort included patients who had surgery between December 1, 2012 and February 15, 2014. The collection and review of clinical information for this study was approved by the OHSU institutional review board, which waived the need for informed consent. The study was registered on clinicaltrials.gov as NCT02450929. This manuscript adheres to the applicable SQUIRE 2.0 guidelines.

During the intervention period, there were no active institutional changes to address postoperative pneumonia. Patients admitted to the ICU received a uniform pneumonia prevention bundle including oral care, head of bed elevation, daily sedation interruptions with spontaneous breathing trials, and appropriate stress-ulcer prophylaxis. There were no other institutional changes to operating room management during the two study periods, including default ventilator settings, aspiration prevention techniques, and oral care.

Patient population

All elective and emergency surgical patients undergoing procedures in the operating room that required endotracheal intubation followed by postoperative hospitalization were included in the study. We excluded patients younger than 18 years of age. For patients undergoing multiple surgeries during a single hospitalization, only the first surgical event of the hospitalization was described.

Outcomes and data collection

The primary outcome was postoperative pneumonia during the hospitalization, identified based on hospital discharge ICD-9 codes for bacterial and fungal pneumonia and included the following specific codes: 481.00–486.99 (pneumonia) and 997.31 (VAP). Centricity (General Electric, Fairfield, CT) and EPIC (Verona, WI) anesthesia information management systems were queried for clinical data. An internal OHSU perioperative patient database that included a cohort of these patients was also used to collect baseline demographics, characteristics of anesthetic and surgical perioperative care, and postoperative variables.

Statistical analysis

Summary statistics (means and standard deviations for quantitative characteristics and frequencies and percentages for categorical factors) were estimated for patients’ demographic characteristics (i.e., age, race, gender, body mass index), perioperative factors (i.e., American Society of Anesthesiologists [ASA] physical status classification, procedural classifications) and potential confounding factors (i.e., tidal volume, rapid sequence intubation, use of non-depolarizing neuromuscular blockade, positive end-expiratory pressure [PEEP]). We first compared the characteristics between the baseline and intervention cohorts using two-sample unequal variance t-test for the quantitative characteristics or chi-squared test statistics for the categorical ones. We also presented ETT cohort (unadjusted) summary statistics for the primary outcomes: (a) VAP, using a chi-square test, (b) duration of mechanical ventilation (min), and (c) hospital length of stay (days). The latter two comparisons were made using an unequal variance t-test. We then formally evaluated whether there were differences in these outcomes by adjusting (controlling) for potential confounding factors. For the postoperative pneumonia outcome and hospital mortality outcome, we performed multivariable logistic regression to test whether the odds of postoperative pneumonia or mortality in the intervention cohort were different than the odds of postoperative pneumonia or mortality in the baseline cohort. For the two quantitative secondary outcomes, hospital length of stay (LOS) and duration of mechanical ventilation (min), we performed multivariable linear regression using robust (sandwich) estimated standard errors to remedy possible violation of the model variance assumption. We tested for differences in the two cohorts using standard (adjusted) pairwise comparison tests.

Finally, we explored whether possible effects between postoperative pneumonia and the two cohorts were modified by the following potential effect modifiers: diabetes, hypertension, COPD, tobacco abuse, ischemic heart disease, GERD, heart failure, obesity, and intraoperative use of nondepolarizing neuromuscular blockade. These analyses were performed using multivariable logistic regression models. Separate models were fitted for each of the potential effect modifiers, similar to the primary adjusted analysis for postoperative pneumonia, but with the inclusion of one additional term for the effect modifier and its interaction term with the cohort predictor. The a priori selected confounding factors included in all adjusted analyses were ASA status (five categories), tidal Volume (ml/kg), Caucasian race (yes/no), age (in years), male gender (yes/no), and PEEP (=0 or > 0). All hypothesis tests, associated p-values and confidence intervals were two-sided. The statistical analyses were performed using Stata (ver. 15.1) and R (ver. 3.3.3) statistical packages.

Results

Demographics

The study flowchart is shown in Fig. 1. During the two study periods, 16,956 patients were potentially eligible. Of these patients, a total of 15,388 (91%) had complete data, with 9037 patients in the baseline cohort and 6351 in the intervention cohort (Table 1). Mean age (p <  0.001), Caucasian race (p = 0.004), ASA class (p <  0.001) and surgical category (p = 0.001) were significantly different among cohorts. However, the differences between these cohorts were not clinically relevant. Gender and mean body mass index were not different between the baseline and intervention cohorts. The frequency distribution of surgical procedures was also similar between the two cohorts in terms of clinical significance, even though they differed statistically (p = 0.001). The significant differences were due to the high degree of precision of the estimates because of the large sample size.

Fig. 1
figure 1

Study flowchart

Table 1 Demographic and intraoperative characteristics of patients intubated with standard barrel cuff endotracheal tube (baseline) and the TaperGuard™ endotracheal tube (intervention). Data are expressed as mean (standard deviation) unless otherwise specified

Intraoperative characteristics

Intraoperative characteristics are also shown in Table 1, stratified by ETT cohort. The use of rapid sequence intubation and the utilization of a non-depolarizing neuromuscular blockade were not significantly different in the two cohorts. The intervention cohort had a significantly lower median tidal volume of 7.9 ± 1.4 mL/kg compared to the baseline cohort (8.2 ± 1.6 mL/kg, p <  0.001), as well as a higher utilization of PEEP intraoperatively (intervention, 97.4% vs. baseline 95.2%, p <  0.001). The mean procedure duration was similar between the cohorts (mean difference = 2.0 min, 95% CI: − 6.0, 2.0; p = 0.333).

Study endpoints

Table 2 provides unadjusted and adjusted estimates for the study primary and secondary endpoints. The unadjusted estimates of the incidence of postoperative pneumonia were 1.62% in the intervention cohort and 1.79% in the baseline cohort (OR 0.9, 95% CI: 0.70, 1.16; p = 0.423). The estimated mean duration of mechanical ventilation was also not different between the two cohorts (mean difference = 2.0 min, 95% CI: − 2.0, 6.0; p = 0.333). Hospital length of stay was 0.6 days longer for the Intervention cohort (95% CI: 0.3, 0.9; p <  0.001). The associations remained unchanged after adjustment for potential confounders. The adjusted difference in the estimated hospital length of stay was slightly attenuated with a mean difference of 0.5 days longer for the intervention cohort (95% CI: 0.2, 0.8; p <  0.001) compared to the baseline cohort. Hospital mortality was significantly higher in the intervention group (1.5%) compared to the baseline cohort (1.0%; OR 1.46, 95% CI: 1.09, 1.95; p = 0.010), which remained unchanged after adjustment for potential confounders.

Table 2 Study primary and secondary endpoints among baseline (standard ETT) and intervention (TaperGuard™ ETT) cohorts. Data are expressed as mean (standard deviation) unless otherwise specified

The results of our investigation of possible effect modification of the relationship between ETT cohort and postoperative pneumonia are shown in Fig. 2. We did not find evidence that any of the disease or subclinical disease groups modified the association between ETT cohort and postoperative pneumonia (all p-values > 0.20).

Fig. 2
figure 2

Odds Ratios of VAP comparing the TaperGuard™ ETT with the standard ETT, among population subgroups

Discussion

To our knowledge, this study represents the largest evaluation of the effectiveness of TaperGuard™ ETTs for the prevention of postoperative pneumonia in a large, heterogeneous, surgical population. We found no difference in the odds of developing postoperative pneumonia during the use of the TaperGuard™ ETT relative to the use of the standard ETT. We also found no differences in postoperative pneumonia between the two cohorts among higher pneumonia risk subgroups. Considering previous work, these findings substantiate previous findings that the TaperGuard™ ETT does not have a role in preventing postoperative pneumonia.

The efficacy of the TaperGuard™ ETT in reducing incidence of VAP has been previously investigated. Bowton et al. conducted an observational, two-period study of ICU patients and found no reduction in VAP rate with hospital and community-wide implementation of the TaperGuard™ ETT [15]. Similar to ours, the VAP bundle adherence was high, resulting in relatively low incidence of VAP, suggesting that the study may not have had adequate power to detect a difference. Bowton et al. defined VAP based on National Healthcare Safety Network criteria [17], because all of their patients had ICU admission. We evaluated the TaperGuard™ ETT in a less selected patient population, who may have received tracheal intubation and mechanical ventilation only during the perioperative period. Due to the broader population, our ability to screen for and detect pneumonia events was limited to data collection performed on the least critically ill inpatients, meaning we were unable to apply National Healthcare Safety Network criteria. The highest quality data source for determination of postoperative pneumonia in our sample was hospital discharge ICD-9 codes for bacterial, fungal, and ventilator-associated pneumonia. It is possible that ICD-9 codes were not coded accurately, and the diagnostic processes leading to the application of codes were not available for us to review in aggregate. We also may have not captured a subset of hospitalized patients who were discharged from the hospital prior to onset of symptoms of pneumonia. However, our sample size was sufficiently large and there is no reason to suspect a difference in the duration or quality of postoperative mechanical ventilation practices or discharge behaviors between the two study periods. Due to the large sample, the estimates are precise, and the associated statistical tests have sufficiently high statistical power. Assuming an incidence of postoperative pneumonia of 1.8% for the standard ETT, the study would exceed 80% power to detect a reduction to 1.2% in the incidence of VAP in the TaperGuard™ ETT. At out institution, we also utilize a VAP prevention bundle as part of the standard of care for all intubated patients, and the components of the bundle did not change over the entire study period. However, it remains possible that due to unmeasured improvements in healthcare over time or unrecognized changes in care delivery between the cohorts, there are unmeasured confounding differences between the cohorts.

The rationale of the tapered cuff design of the TaperGuard™ ETT is to create a more complete seal around the tracheal wall, thus reducing micro-channels that allow the leakage of supraglottic secretions below the cuff. Multiple small laboratory studies have evaluated the degree to which the TaperGuard™ ETT reduces leakage of fluid around an inflated cuff in an experimental tracheal model. Most have found that in static conditions, the tapered cuff reduces the passage of fluid below the cuff [8,9,10,11,12,13,14]. Experimental studies that more closely mimic the physiologic conditions of tracheal intubation and ventilation suggest less sustained effects. One study found that the ability of TaperGuard™ ETT to reduce fluid leakage decreased significantly with PEEP < 10 cm H2O and with intubation times longer than 60 min [10]. Another study using microbial suspensions of Staphylococcus, Pseudomonas and Candida above the cuff revealed that the TaperGuard™ ETT failed to prevent inoculation of the space below the cuff [11]. The degree of inoculation also varied based on the diameter of the experimental tracheal model used, and the volume of fluid leakage.

The available evidence suggests that the TaperGuard™ ETT reduces measurable fluid leakage but does not prevent the passage of fluid or microorganisms, especially not over a broad set of conditions or for clinically relevant time periods. Measurement of microaspiration in vivo has not been thoroughly evaluated [14,15,16, 18]. In one study, 60 patients scheduled for lumbar spine surgery were intubated and equally randomized to receive a standard ETT or the TaperGuard™ ETT0 [18]. Dye was instilled into the supraglottic space, and bronchoscopy was performed to assess the degree of dye descent along the cuff up to 2 h. The TaperGuard™ ETT allowed dye leakage up to the second third of the cuff, but none into the subglottis. It is conceivable that the reduced but incomplete degree of protection from microaspiration of the tapered cuff design could contribute to the lack of effect on VAP prevention observed in our study. To our knowledge, the only other randomized trial that compared the standard ETT to the TaperGuard™ ETT was conducted in 109 high-risk patients undergoing major vascular surgery [16]. All of their patients were transferred to the ICU and screened daily for clinical suspicion of VAP until 5 days post extubation, up to 28 days. Tracheal aspirates were also sampled for pepsin and amylase to measure microaspiration. While the incidence of pneumonia was relatively high (42–44%) in that study, there was no difference in the incidence of VAP between groups. Additionally, there were no differences in amylase and pepsin concentrations from tracheal aspirates on postoperative days 1 and 2. Taken together, these findings suggest that the TaperGuard™ ETT does not appear to reliably prevent microaspiration or VAP even in high-risk groups.

Conclusion

The current study contributes to a growing body of literature suggesting that the TaperGuard™ ETT is not an effective device for broad implementation in surgical patients to reduce the risk of VAP in the perioperative setting. However, cuff shape is only one component of endotracheal ETT design modifications, and there are other possible configurations or material of cuffed ETTs that may be more effective. The inclusion of subglottic suctioning, or a polyurethane rather than polyvinylchloride cuff, are features that may be more important for VAP reduction, especially in high-risk patient groups. Based on collective evidence, TaperGuard™ ETT does not appear to prevent postoperative pneumonia. Utilization of this specialized ETT for the prevention of postoperative pneumonia in an unselected surgical patient population is not warranted.

Availability of data and materials

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

VAP:

Ventilator associated pneumonia

ETT:

Endotracheal tube

OR:

Odds ratio

OHSU:

Oregon Health & Science University

NSQIP:

National Surgical Quality Improvement Project

ASA:

American Society of Anesthesiologists

LOS:

Length of stay

PEEP:

Positive end-expiratory pressure

References

  1. Kalil AC, Metersky ML, Klompas M, et al. Management of adults with hospital-acquired and ventilator-associated pneumonia: 2016 clinical practice guidelines by the Infectious Diseases Society of America and the American Thoracic Society. Clin Infect Dis. 2016;63(5):e61–e111.

    Article  PubMed  PubMed Central  Google Scholar 

  2. Rello J, Ollendorf DA, Oster G, et al. Epidemiology and outcomes of ventilator-associated pneumonia in a large US database. Chest. 2002;122:2115–21.

    Article  PubMed  Google Scholar 

  3. Stevens JP, Kachniarz B, Wright SB, et al. When policy gets it right: variability in US Hospitals' diagnosis of ventilator-associated pneumonia*. Crit Care Med. 2014;42:497–503.

    Article  PubMed  Google Scholar 

  4. Eom JS, Lee MS, Chun HK, et al. The impact of a ventilator bundle on preventing ventilator-associated pneumonia: a multicenter study. Am J Infect Control. 2014;42:34–7.

    Article  PubMed  Google Scholar 

  5. Muscedere J, Rewa O, McKechnie K, Jiang X, Laporta D, Heyland DK. Subglottic secretion drainage for the prevention of ventilator-associated pneumonia: a systematic review and meta-analysis. Crit Care Med. 2011;39:1985–91.

    Article  PubMed  Google Scholar 

  6. Mahmoodpoor A, Hamishehkar H, Hamidi M, et al. A prospective randomized trial of tapered-cuff endotracheal tubes with intermittent subglottic suctioning in preventing ventilator-associated pneumonia in critically ill patients. J Crit Care. 2017;38:152–6.

    Article  PubMed  Google Scholar 

  7. Deem S, Treggiari MM. New endotracheal tubes designed to prevent ventilator-associated pneumonia: do they make a difference? Respir Care. 2010;55:1046–55.

    PubMed  Google Scholar 

  8. Chenelle CT, Itagaki T, Fisher DF, Berra L, Kacmarek RM. Performance of the PneuX system: a bench study comparison with 4 other endotracheal tube cuffs. Respir Care. 2017;62:102–12.

    Article  PubMed  Google Scholar 

  9. Matsunami S, Komasawa N, Fujiwara S, Nakao K, Tatsumi S, Minami T. Comparison of TaperGuard tube and the Portex Softseal for prevention of vomitus leakage in an in vitro simulation airway model. Masui. 2014;63:586–9.

    PubMed  Google Scholar 

  10. Zanella A, Scaravilli V, Isgro S, et al. Fluid leakage across tracheal tube cuff, effect of different cuff material, shape, and positive expiratory pressure: a bench-top study. Intensive Care Med. 2011;37:343–7.

    Article  PubMed  Google Scholar 

  11. Mariyaselvam MZ, Marsh LL, Bamford S, Smith A, Wise MP, Williams DW. Endotracheal tubes and fluid aspiration: an in vitro evaluation of new cuff technologies. BMC Anesthesiol. 2017;17:36.

    Article  PubMed  PubMed Central  Google Scholar 

  12. Lau AC, Lam SM, Yan WW. Benchtop study of leakages across the Portex, TaperGuard, and microcuff endotracheal tubes under simulated clinical conditions. Hong Kong Med J. 2014;20:7–15.

    Article  PubMed  Google Scholar 

  13. Shiotsuka J, Lefor AT, Sanui M, Nagata O, Horiguchi A, Sasabuchi Y. A quantitative evaluation of fluid leakage around a polyvinyl chloride tapered endotracheal tube cuff using an in-vitro model. HSR Proc Intensive Care Cardiovasc Anesth. 2012;4:169–75.

    CAS  PubMed  PubMed Central  Google Scholar 

  14. Gaszynska E, Ratajczyk P, Wieczorek A, Szewczyk T, Gaszyński T. Comparison of microaspiration around TaperGuard tube cuffs and standard tracheal tubes in obese patients subjected to surgery under general anesthesia. Pol Przegl Chir. 2014;86:107–10.

    Article  PubMed  Google Scholar 

  15. Bowton DL, Hite RD, Martin RS, Sherertz R. The impact of hospital-wide use of a tapered-cuff endotracheal tube on the incidence of ventilator-associated pneumonia. Respir Care. 2013;58:1582–7.

    Article  PubMed  Google Scholar 

  16. Monsel A, Lu Q, Le Corre M, et al. Tapered-cuff endotracheal tube does not prevent early postoperative pneumonia compared with spherical-cuff endotracheal tube after major vascular surgery: a randomized controlled trial. Anesthesiology. 2016;124:1041–52.

    Article  PubMed  Google Scholar 

  17. Rosenthal VD, Al-Abdely HM, El-Kholy AA, et al. International nosocomial infection control consortium report, data summary of 50 countries for 2010-2015: device-associated module. Am J Infect Control. 2016;44:1495–504.

    Article  PubMed  Google Scholar 

  18. D'Haese J, De Keukeleire T, Remory I, Van Rompaey K, Umbrain V, Poelaert J. Assessment of intraoperative microaspiration: does a modified cuff shape improve sealing? Acta Anaesthesiol Scand. 2013;57:873–80.

    Article  CAS  PubMed  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

This study was funded by a grant from Covidien, and performed at Oregon Health & Science University. The funder had no role in the data acquisition, analysis or reporting.

Author information

Authors and Affiliations

Authors

Contributions

RPM: This author made substantial contributions to the conception and design of the work, data acquisition, analysis, and interpretation, drafting and revising the work, and final approval of the manuscript. NDY: This author made substantial contributions to the data acquisition, analysis, and interpretation, drafting and revising the work, and final approval of the manuscript. MMT: This author made substantial contributions to the data analysis, and interpretation, drafting and revising the work, and final approval of the manuscript. PT: This author made substantial contributions to the data acquisition, analysis, and interpretation, and final approval of the manuscript. CS: This author made substantial contributions to the conception and design of the work, revising the work, and final approval of the manuscript. MFA: This author made substantial contributions to the conception and design of the work, revising the work, and final approval of the manuscript. All authors have read and approved this manuscript.

Authors’ information

Not applicable.

Corresponding author

Correspondence to Miriam M. Treggiari.

Ethics declarations

Ethics approval and consent to participate

The collection and review of clinical information for this study was approved by the Oregon Health & Science University Institutional Review Board, with reference #10160.

The Oregon Health & Science University Institutional Review Board waived the need for informed consent.

Consent for publication

Not Applicable.

Competing interests

This was an investigator-initiated study. The sponsor (Covidien, Mansfield, Massachusetts) provided the tapered-cuff ETTs at a discounted price and provided salary support for data collection. The sponsor was not involved in study design, data collection, or data analysis. The sponsor had no access to study data, nor any role in formulating the conclusions for this report but was allowed to review the manuscript prior to submission for publication.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Martini, R.P., Yanez, N.D., Treggiari, M.M. et al. Implementation of the TaperGuard™ endotracheal tube in an unselected surgical population to reduce postoperative pneumonia. BMC Anesthesiol 20, 211 (2020). https://doi.org/10.1186/s12871-020-01117-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12871-020-01117-4

Keywords