A retrospective study of the effects of a vasopressor bolus on systolic slope (dP/dt) and dynamic arterial elastance (Eadyn)

Abstract

Background

To enhance the utility of functional hemodynamic monitoring, the variables systolic slope (dP/dt) and dynamic arterial elastance (Eadyn) are calculated by the Hypotension Prediction Index (HPI) AcumenĀ® Software. This study was designed to characterize the effects of phenylephrine and ephedrine on dP/dt and Eadyn.

Methods

This was a retrospective, non-randomized analysis of data collected during two clinical studies. All patients required intra-operative controlled mechanical ventilation and had an indwelling radial artery catheter connected to an Acumen IQ sensor. Raw arterial pressure waveform data was downloaded from the patient monitor and all hemodynamic measurements were calculated off-line. The anesthetic record was reviewed for bolus administrations of either phenylephrine or ephedrine. Cardiovascular variables prior to drug administration were compared to those following vasopressor administrations. The primary outcome was the difference for dP/dt and Eadyn at baseline compared with the average after the bolus administration. All data sets demonstrated non-normal distributions so statistical analysis of paired and unpaired data followed the Wilcoxon matched pairs signed-rank test or Mann-Whitney U test, respectively.

Results

201 doses of phenylephrine and 100 doses of ephedrine were analyzed. All data sets are reported as median [95% CI]. Mean arterial pressure (MAP) increased from 62 [54,68] to 78 [76,80] mmHg following phenylephrine and from 59 [55,62] to 80 [77,83] mmHg following ephedrine. Stroke volume and cardiac output both increased. Stroke volume variation and pulse pressure variation decreased. Both drugs produced significant increases in dP/dt, from 571 [531, 645] to 767 [733, 811] mmHg/sec for phenylephrine and from 537 [509, 596] to 848 [779, 930] mmHg/sec for ephedrine. No significant changes in Eadyn were observed.

Conclusion

Bolus administration of phenylephrine or ephedrine increases dP/dt but does not change Eadyn. dP/dt demonstrates potential for predicting the inotropic response to phenylephrine or ephedrine, providing guidance for the most efficacious vasopressor when treating hypotension.

Trial registration

Data was collected from two protocols. The first was deemed to not require written, informed consent by the Institutional Review Board (IRB). The second was IRB-approved (Effect of Diastolic Dysfunction on Dynamic Cardiac Monitors) and registered on ClinicalTrials.gov (NCT04177225).

Peer Review reports

Background

Intraoperative hypotension is common in patients undergoing general anesthesia [1, 2]. Maintaining optimal intraoperative hemodynamics is paramount, as strong correlations exist between hypotension and the incidence of postoperative myocardial infarction, acute kidney injury, and mortality [3,4,5]. Hypotension can be broadly considered as pathology related to inadequate fluid resuscitation, vasomotor tone, and/or cardiac function. As such, the treatment of hypotension includes identifying the cause and administering intravenous fluids, vasoconstrictors, and/or inotropes, respectively [6]. Determining the most efficacious treatment for a hypotensive episode can be elusive as blood pressure is determined by complex interactions between cardiac stroke volume and arterial tone [7, 8].

The HemoSphere patient monitor with Hypotension Prediction Index (HPI) AcumenĀ® software (Edwards Lifesciences, Irvine, CA) is designed to guide intraoperative therapy and minimize hypotension. Machine learning was used to develop an algorithm that can predict an impending hypotensive event based on the integration of selected arterial waveform characteristics [9]. The HPI software has been validated as a reliable predictive index in both cardiac [10] and non-cardiac [11] surgical procedures and has been shown to decrease the total duration of intraoperative hypotension [12]. This software provides hemodynamic variables that characterize contractility, (systolic slope, dP/dt) and vascular tone (dynamic arterial elastance, Eadyn). These variables may provide therapeutic guidance for treatment of hypotension. Characterization of their response to commonly used vasoconstrictors may improve patient management. This study examined specifically how these variables respond to phenylephrine and ephedrine.

Methods

Data collected from two protocols were included for this single center study. The first protocol was an evaluation of non-invasive cardiac output monitors in patients undergoing elective cardiac surgery. This study was deemed to not require written, informed consent by the Institutional Review Board (IRB) as part of a quality improvement project. The primary results from this evaluating the utility of the HPI in cardiac surgical patients has been published [10]. The second protocol was an IRB-approved prospective evaluation of the effects of diastolic function on the dynamic monitors of cardiac function (Effect of Diastolic Dysfunction on Dynamic Cardiac Monitors, NCT04177225, first registered 26/11/2019). The primary data from this protocol is still being evaluated.

All patients in this second protocol provided written, informed consent. All patients required controlled mechanical ventilation and a radial artery pressure monitoring catheter connected to Acumen IQ sensor and a physiological monitor (EV-1000 or HemoSphere) that contained the HPI software (Edwards Lifesciences LLC, Irvine, CA). Post-operatively, the raw waveform data was downloaded from the physiological monitor. All hemodynamic parameters were calculated off-line at twenty second intervals. The following parameters were extracted and evaluated: mean arterial pressure (MAP), heart rate (HR), stroke volume (SV), cardiac output (CO), stroke volume variation (SVV), pulse pressure variation (PPV), systolic slope (dP/dt), and dynamic arterial elastance (Eadyn).

Each patientā€™s electronic medical record was reviewed for bolus administrations of either phenylephrine or ephedrine. Medications administered during cardiopulmonary bypass were excluded. Given the potential for imprecise charting for medication administration, drug administration time was defined as the time with the lowest blood pressure occurring within 3Ā min of the electronic medical record documented drug administration. Baseline cardiovascular variables were determined as the average over three minutes prior to drug administration. The primary outcome was the difference for the cardiovascular variables dP/dt and Eadyn at baseline compared with the average taken over three minutes beginning two minutes after the bolus vasopressor administration.

Statistical analysis

Initial review of the scatter plots for each hemodynamic variable revealed apparent outliers. For each value associated with a changeā€‰ā‰„ā€‰3 standard deviations of the average values for that variable, the primary data were reviewed. If the primary data values appeared to be physiologically unlikely, probably reflecting a recording artifact, all variables associated with that drug administration were removed from the final analysis. For all measured hemodynamic variables, the distribution of each final data set was assessed using the Dā€™Agostino Pearson normality test. All data sets demonstrated non-normal distributions. Statistical analysis of paired and unpaired data followed the Wilcoxon matched pairs signed-rank test or Mann-Whitney U test, respectively using GraphPad Prism version 10.2.1 for Windows, GraphPad Software, San Diego, California USA, www.graphpad.com. An alpha of 0.05 was applied to determine statistical significance.

Results

Data from 101 patients undergoing elective, non-laparoscopic cardiac, vascular, intra-abdominal oncologic, or head and neck cancer surgery collected between January 2019 and April 2020 at the University of California, Davis Medical Center were included in this study. 72 were male, 29 were female with an average age of 69ā€‰Ā±ā€‰10.8 years, average weight of 82ā€‰Ā±ā€‰18.0Ā kg, and average height of 173ā€‰Ā±ā€‰9.3Ā cm. Four vasopressor doses were removed from analysis as outliers or artifacts. The MAP response for different doses of phenylephrine (50, 80, and 100 mcg) were not statistically different (12ā€‰Ā±ā€‰12, 22ā€‰Ā±ā€‰12, 18ā€‰Ā±ā€‰14 mmHg, respectively). The MAP response for different doses of ephedrine (5 and 10Ā mg), were not statistically different (23ā€‰Ā±ā€‰16, 23ā€‰Ā±ā€‰16 mmHg, respectively). Thus, all doses of phenylephrine were combined to comprise the data set for each hemodynamic parameter extracted; likewise, for all doses of ephedrine. A total of 201 doses of phenylephrine (doses/patient: mean 2.73, median 2, maximum 9) and 100 doses of ephedrine (doses/patient: mean 1.98, median 2, maximum 8) were included for analysis.

The changes in the absolute values of each hemodynamic variable are provided in TableĀ 1. The bolus administration of phenylephrine was associated with an average increase of 28ā€‰Ā±ā€‰22% in MAP (median:23%, 95% CI:20.6ā€“27.3). There was no significant change in heart rate (average 1ā€‰Ā±ā€‰8%, median 0.1%, 95% CI: -0.6ā€“1.2). The stroke volume significantly increased by an average of 6ā€‰Ā±ā€‰15.7% (median: 3%, 95% CI: 1.8ā€“4.6) with a consequent average increase in the cardiac output of 7ā€‰Ā±ā€‰16% (median:5%, 95%CI:3.5ā€“6.1). The stroke volume variation significantly decreased by an average 14ā€‰Ā±ā€‰30% (median: -15%, 95% CI: -19.5 to -11.3) as did the pulse pressure variation by an average 12ā€‰Ā±ā€‰32.5% (median: -14%, 95% CI: -18.8 to -7.5). (Fig.Ā 1)

Table 1 Hemodynamic Variables - Absolute Values
Fig. 1
figure 1

Hemodynamic responses to phenylephrine and ephedrine as absolute values. Paired comparisons to baseline values, ā€œ*ā€ indicates pā€‰<ā€‰0.05

The systolic slope (dP/dt) increased significantly by 41ā€‰Ā±ā€‰51% (median:26%, 95% CI: 20.8ā€“31.9) but there was no significant change in the dynamic arterial elastance (Eadyn), average 4ā€‰Ā±ā€‰20% (median: 3%, 95% CI -0.4ā€“5.8). (Fig.Ā 2) For ephedrine, the bolus administration was associated with an average increase of 40ā€‰Ā±ā€‰30% in MAP (median:31%, 95% CI:26.2ā€“43.6). Heart rate increased significantly by an average of 8ā€‰Ā±ā€‰12% (median:6%, 95% CI: 3.5ā€“9.4) The stroke volume also increased significantly by an average of 14ā€‰Ā±ā€‰31.8% (median: 4%, 95% CI: 2.8ā€“8.1), combining with the increased heart rate to produce an average increase of 21ā€‰Ā±ā€‰32% in the cardiac output (median:12%, 95% CI:7.8ā€“16.5). The stroke volume variation significantly decreased by an average of 20ā€‰Ā±ā€‰24% (median: -22%, 95% CI: -26 to -15) with a similarly significant decrease of 15ā€‰Ā±ā€‰35.7% (median: -24%, 95% CI: -30.6 to -11.2) in the pulse pressure variation. (Fig.Ā 1) The systolic slope (dP/dt) increased significantly by 63ā€‰Ā±ā€‰54% (median:52%, 95% CI: 36.5ā€“66.1) but there was no significant change in the dynamic arterial elastance (Eadyn), average 9ā€‰Ā±ā€‰36% (median: -1%, 95% CI -5.6ā€“3.4). (Fig.Ā 2)

Fig. 2
figure 2

Hemodynamic variable responses to phenylephrine or ephedrine in absolute values. Paired comparisons to baseline values, ā€œ*ā€ indicates pā€‰<ā€‰0.05

When the responses to phenylephrine and ephedrine were compared, the average increase in MAP (17ā€‰Ā±ā€‰13 mmHg vs. 23ā€‰Ā±ā€‰16 mmHg) was comparable. However, ephedrine produced a significant increase in heart rate (0.5ā€‰Ā±ā€‰5 vs. 4ā€‰Ā±ā€‰6 beats per minute) and while the increases in stroke volume were comparable (3ā€‰Ā±ā€‰10 vs. 7ā€‰Ā±ā€‰13Ā ml), ephedrine produced a greater increase in cardiac output (0.2ā€‰Ā±ā€‰0.8 vs. 0.7ā€‰Ā±ā€‰1.0Ā L per minute). The effects on SVV and PPV were comparable for both drugs (-2ā€‰Ā±ā€‰3.8 vs. -3ā€‰Ā±ā€‰3.3% and āˆ’ā€‰2ā€‰Ā±ā€‰4.9 vs. -3ā€‰Ā±ā€‰4.3%, respectively). The increase in dP/dt (198ā€‰Ā±ā€‰202 vs. 330ā€‰Ā±ā€‰245 mmHg/sec) was greater for ephedrine in comparison to phenylephrine, while there was no difference in the changes for Eadyn (0.02ā€‰Ā±ā€‰0.23 vs. 0.04ā€‰Ā±ā€‰0.28) between the two drugs. (Fig.Ā 3).

Fig. 3
figure 3

Hemodynamic variable response to phenylephrine or ephedrine, change in absolute values. Group comparisons between drugs, ā€œ*ā€ indicates pā€‰<ā€‰0.05

Secondary analysis was then performed to evaluate the influence of the baseline intravascular volume status, as characterized by the SVV, on the observed responses. Patients with an SVVā€‰ā‰„ā€‰13% were considered to be volume responsive. Weak correlations were observed between the baseline SVV and the increase in MAP, SV and dP/dt following vasopressor administration. These correlations were slightly stronger following ephedrine. No correlations were observed for the changes in HR or Eadyn. (TableĀ 2) For within group comparisons following the administration of phenylephrine, only the change in dP/dt (pā€‰=ā€‰0.02) was greater in the volume responsive patients as compared to the non-volume responsive patients. Following ephedrine, the increases in CO (pā€‰=ā€‰0.0004), SV (pā€‰=ā€‰0.003), MAP (pā€‰=ā€‰0.003) and dP/dt (pā€‰<ā€‰0.0001) were all greater in the volume responsive patients. For comparisons between groups, in the non-volume responsive patients, the increases in CO (pā€‰=ā€‰0.001), HR (pā€‰<ā€‰0.0001) and dP/dt (pā€‰=ā€‰0.001) were all greater following administration of ephedrine. In the volume responsive patients, the increases in CO (pā€‰=ā€‰0.001), SV (pā€‰=ā€‰0.045), HR (pā€‰<ā€‰0.001), MAP (pā€‰=ā€‰0.002) and dP/dt (pā€‰=ā€‰0.0005) were also greater following the administration of ephedrine.

Table 2 Correlations Between SVV and Hemodynamic Variables

Discussion

The hemodynamic variables of maximal systolic slope (dP/dt) and dynamic arterial elastance (Eadyn) are not new [13,14,15], however, with the HPI AcumenĀ® software, they are readily available to potentially guide intra-operative blood pressure management. dP/dt provides a correlate of myocardial contractility [16,17,18] while Eadyn provides a characterization of vascular resistance [7, 15]. Because clinical experience with these values is limited, we evaluated their response to bolus administrations of phenylephrine or ephedrine. The relationships between these vasopressors and the additional hemodynamic variables provided by the HPI software were also observed. Both phenylephrine and ephedrine increased MAP, CO and dP/dt, while neither had any significant effect on Eadyn. Both drugs decreased PPV and SVV.

The use of rate of pressure change (dP/dt) measurements to characterize myocardial contractility initially tracked left ventricular (LV) pressure during isovolumetric contraction [13]. This measurement requires direct LV catheterization. Studies which administered assorted inotropes demonstrated dP/dt is very sensitive to changes in contractility in isolated tissues, animals and humans [16]. Further, dP/dt was shown to be more accurate when trending cardiac contractility changes within individual patients in response to acute interventions as opposed to a comparative measure for assessing absolute values of contractility among patients [17]. To expand the clinical utility of this measurement, dP/dt derived from the arterial pressure has been shown to correlate with both direct LV pressure measurements [17, 19] and echocardiographic surrogates [18]. Although the absolute values differ between LV and arterial dP/dt, the response to inotropes and vasodilators within a given patient correlate [17]. The HPI AcumenĀ® software makes peripheral arterial dP/dt continuously available. Characterizing the impact of commonly used vasopressors on this variable will improve the use of this LV contractility measure to guide patient care. However, the limitations of radial arterial pressure measurement in comparison to direct LV pressure measurement should be emphasized, given the multiple potential confounding influences on the radial pressure tracing.

Clinical strategies for resuscitating hypotensive patients rely on understanding fluid responsiveness and vasomotor tone [20]. The ratio of pulse pressure (PP) to stroke volume (SV) was proposed as a reliable estimate of total arterial compliance based on the Windkessel model, which integrates LV SV and the stiffness of an arterial vessel to predict change in arterial pressure [21, 22]. If arterial elastance is held constant, PP is directly proportional to SV, and greater PPV positively correlates with increases in CO following intravenous fluid administration [23]. Expansion upon this model reveals Eadyn, the ratio of PPV to SVV, as a characterization of vasomotor tone. Studies have focused on this parameterā€™s utility for predicting changes in MAP among patients in the intensive care setting in the response to fluid boluses or vasopressors [7, 24]. Garcia et al. demonstrated that Eadyn was sufficient to predict arterial pressure change in a study of hypotensive patients under controlled mechanical ventilation after receiving fluids, while neither the ratio PP/SV nor systemic vascular resistance (SVR) predicted a pressure response [22]. Eadyn assessed arterial tone in resuscitated septic patients was also shown to predict lower MAP after decreased norepinephrine doses were given [24]. Identifying relationships among vasopressors and Eadyn may strengthen its utility as an indicator of arterial tone perioperatively.

When comparing the effects of the vasopressors on the primary hemodynamic variables for this study, ephedrine demonstrated a greater impact than phenylephrine on MAP, HR, CO and dP/dt. Ephedrineā€™s greater effect on MAP could be explained by its Ī±1-agonist properties which increase SVR in combination with its Ī²-agonist actions and concomitant increase in HR and contractility. In contrast, phenylephrine, as a pure Ī±1-agonist, has no corresponding direct effects on HR or contractility. Both drugs were, however, associated with an increase in CO and contractility as assessed by the dP/dt. These observations were not expected for phenylephrine but are consistent with previous observations. The Ī±1-mediated vasoconstriction increases both arterial and venous vascular tone. This increase in venous tone may increase left ventricular end-diastolic volume with consequent increased contractility [25]. In animal studies, phenylephrine increased MAP in both preload independent and preload dependent (volume-responsive) states. However, in the volume-responsive state, phenylephrine increased venous return and increased cardiac output [26]. Human studies in clinical settings have not been as consistent. Rebet et al. found that in general surgical patients, phenylephrine consistently increased MAP but decreased CO in patients who were preload independent and had no effect on CO in patients who were preload dependent [27]. In contrast, Kalmar et al. studied the effects of phenylephrine for the treatment of hypotension in patients undergoing elective sigmoidectomy [28]. Nearly all patients were volume-responsive with an average baseline SVV of 19% and CO increased by an average of 26% following administration of phenylephrine [28]. In addition, the increased arterial tone and associated increased afterload might also increase the dP/dt, independent of a change in contractility [13, 29]. Ephedrine will produce all these same actions and, in addition, increase contractility through Ī²1-agonist effects resulting in the observed greater effects on HR, CO and contractility as measured by dP/dt.

In contrast, neither phenylephrine nor ephedrine had a significant effect on Eadyn. Eadyn= PPV/SVV. Both drugs decreased SVV and PPV by a similar magnitude. If Eadyn was a simple measure of vascular tone, vasoconstriction should increase Eadyn. The most common measure of vascular tone, systemic vascular resistance (SVR), calculates the average resistance to blood flow over time that is primarily a function of arteriolar tone. In contrast, Eadyn uses the dynamic variables of SVV and PPV to provide a more comprehensive characterization of vascular impedance by incorporating the oscillatory nature of blood flow, arterial compliance and blood pressure wave propagation and reflection [22]. Our results emphasize that Eadyn is a more nuanced value that may predict the blood pressure response to a fluid bolus but is not a simple measure of systemic vascular resistance. For this study, conclusions regarding the comparisons of these two variables are limited because central venous pressure measurements were not available for the calculation of SVR in all patients.

The generalizability of our findings to other surgical populations may be limited since the etiology of intraoperative hypotension is often multifactorial. The efficacy of either dP/dt or Eadyn as therapeutic guides to treat intraoperative hypotension needs further exploration. We retrospectively analyzed hemodynamic parameters after vasopressor administration in surgeries conducted under general anesthesia. To strengthen the study design and more completely characterize the effects, additional information including the amount of concurrent fluid given to a patient during the hypotensive episode, ventilation mode, and primary anesthetic could be standardized. Prospective, blinded vasopressor treatments would provide more robust comparisons. In addition, in this study patient population of cardiac and general surgery patients, their variable comorbidities, physiologic hemostasis, and susceptibility to hypotension were not controlled.

Conclusions

To our best knowledge, this study is the first to specifically measure the responses of dP/dt and Eadyn to vasopressor boluses. Our understanding of how dP/dt and Eadyn respond to phenylephrine and ephedrine is quantifiably characterized. dP/dt demonstrates potential for characterizing the inotropic response after a bolus of phenylephrine or ephedrine while Eadyn does not appear to be affected by bolus administration of phenylephrine or ephedrine.

Data availability

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

CO:

Cardiac Output

dP/dt:

Systolic Slope

Eadyn :

Dynamic arterial elastance

HPI:

Hypotension Prediction Index

HR:

Heart rate

IRB:

Institutional Review Board variation

LV:

Left Ventricular

MAP:

Mean Arterial Pressure

PP:

Pulse Pressure

PPV:

Pulse Pressure Variation

SVR:

Systemic Vascular Resistance

SV:

Stroke Volume

SSV:

Stroke Volume Variation

References

  1. van BijkerJB WA, Kappen TH, et al. Incidence of Intraoperative Hypotension as a function of the Chosen Definition. Literature definitions Applied to a Retrospective Cohort using Automated Data Collection. Anesthesiology. 2007;107:213ā€“20.

    ArticleĀ  Google ScholarĀ 

  2. Saasouh W, Christensen AL, Xing F, et al. Incidence of intraoperative hypotension during nonā€“cardiac surgery in community anesthesia practice: a retrospective observational analysis. Perioperative Med. 2023;12:29.

    ArticleĀ  Google ScholarĀ 

  3. Sun LY, Wijeysundera DN, Tait GA, et al. Association of intraoperative hypotension with acute kidney injury after elective noncardiac surgery. Anesthesiology. 2015;123(3):515ā€“23.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  4. Salmasi V, Maheshwari K, Yang D, et al. Relationship between intraoperative hypotension, defined by either reduction from baseline or Absolute Thresholds, and Acute kidney and myocardial Injury after noncardiac surgery. Retrospective Cohort Anal Anesthesiology. 2017;126:47ā€“65.

    Google ScholarĀ 

  5. Wesselink EM, Kappen TH, Torn HM, et al. Intraoperative hypotension and the risk of postoperative adverse outcomes: a systematic review. Br J Anaesth. 2018;121(4):706ā€“21.

    ArticleĀ  CASĀ  PubMedĀ  Google ScholarĀ 

  6. Vincent JL, De Backer D, Circulatory Shock. N Engl J Med. 2013;369(18):1726ā€“34.

    ArticleĀ  CASĀ  PubMedĀ  Google ScholarĀ 

  7. Holder AL, Pinsky MR. Applied physiology at the bedside to drive resuscitation algorithms. J Cardiothorac Vasc Anesth. 2014;28(6):1642ā€“59.

    ArticleĀ  PubMedĀ  PubMed CentralĀ  Google ScholarĀ 

  8. Magder S. The meaning of blood pressure. Crit Care. 2018;22(1):257.

    ArticleĀ  CASĀ  PubMedĀ  PubMed CentralĀ  Google ScholarĀ 

  9. Hatib F, Jian Z, Buddi S, et al. Machine-learning Algorithm to Predict Hypotension based on high-fidelity arterial pressure Waveform Analysis. Anesthesiology. 2018;129(4):663ā€“74.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  10. Shin B, Maler SA, Reddy K, et al. Use of the Hypotension Prediction Index during Cardiac surgery. J Cardiothorac Vasc Anesth. 2021;35(6):1769ā€“75.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  11. Davies SJ, Vistisen ST, Jian Z, et al. Ability of an arterial waveform analysis-derived hypotension Prediction Index to predict future hypotensive events in Surgical patients. Anesth Analg. 2020;130(2):352ā€“9.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  12. Wijnberge M, Geerts BF, Hol L, et al. Effect of a machine learning-derived early warning system for Intraoperative Hypotension vs Standard Care on depth and duration of intraoperative hypotension during elective noncardiac surgery: the HYPE randomized clinical trial. JAMA. 2020;323(11):1052ā€“60.

    ArticleĀ  PubMedĀ  PubMed CentralĀ  Google ScholarĀ 

  13. Mason DT. Usefulness and limitations of the rate of rise of intraventricular pressure (dp-dt) in the evaluation of myocardial contractility in man. Am J Cardiol. 1969;23(4):516ā€“27.

    ArticleĀ  CASĀ  PubMedĀ  Google ScholarĀ 

  14. Remington JW, Noback CR, Hamilton WF, et al. Volume elasticity characteristics of the human aorta and prediction of the stroke volume form the pressure pulse. Am J Physiol. 1948;153(2):298ā€“308.

    ArticleĀ  CASĀ  PubMedĀ  Google ScholarĀ 

  15. Pinsky MR. Functional hemodynamic monitoring: Applied physiology at the bedside. In: Vincent JL, editor. Yearbook of emergency and intensive Care Medicine. Berlin, GDR: Springer-; 2001. pp. 537ā€“52.

    Google ScholarĀ 

  16. Mason D, Braunwald E, Covell JW, et al. Assessment of cardiac contractility. Circulation. 1971;44(1):47ā€“58.

    ArticleĀ  CASĀ  PubMedĀ  Google ScholarĀ 

  17. De Hert SG, Robert D, Cromheecke S, et al. Evaluation of left ventricular function in anesthetized patients using femoral artery dP/dtmax. J Cardiothorac Vasc Anesth. 2006;20(3):325ā€“30.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  18. Tartiere JM, Logeart D, Beauvais F, et al. Non-invasive radial pulse wave assessment for the evaluation of left ventricular systolic performance in heart failure. Eur J Heart Fail. 2007;9:477ā€“83.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  19. Morimont P, Lambermont B, Desaive T, et al. Arterial dP/dtmax accurately reflects left ventricular contractility during shock when adequate vascular filling is achieved. BMC Cardiovasc Disord. 2012;12:13.

    ArticleĀ  PubMedĀ  PubMed CentralĀ  Google ScholarĀ 

  20. Pinsky MR. Defining the boundaries of bedside pulse contour analysis: dynamic arterial elastance. Crit Care. 2011;15(1):120.

    ArticleĀ  PubMedĀ  PubMed CentralĀ  Google ScholarĀ 

  21. Chemla D, HĆ©bert JL, Coirault C, et al. Total arterial compliance estimated by stroke volume-to-aortic pulse pressure ratio in humans. Am J Physiol. 1998;274(2):H500ā€“5.

    CASĀ  PubMedĀ  Google ScholarĀ 

  22. GarcĆ­a MI, Cano AG, Romero MG. Dynamic arterial elastance to predict arterial pressure response to volume loading in preload-dependent patients. Crit Care. 2011;15(1):R15.

    ArticleĀ  Google ScholarĀ 

  23. Teboul JL, Monnet X, Chemla D, et al. Arterial pulse pressure variation with mechanical ventilation. Am J Respir Crit Care Med. 2019;199(1):22ā€“31.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  24. Guinot PG, Bernard E, Levrard M, et al. Dynamic arterial elastance predicts mean arterial pressure decrease associated with decreasing norepinephrine dosage in septic shock. Crit Care. 2015;19(1):14.

    ArticleĀ  PubMedĀ  PubMed CentralĀ  Google ScholarĀ 

  25. Jacobs R, Lochy S, Malbrain M. Phenylephrine-induced recruitable preload from the venous side. J Clin Monit Comput. 2019;33:373ā€“6.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  26. Cannesson M, Jian Z, Chen G, et al. Effects of phenylephrine on cardiac output and venous return depend non the position of the heart on the Frank-Starling relationship. J Appl Physiol. 2012;113:281ā€“9.

    ArticleĀ  CASĀ  PubMedĀ  Google ScholarĀ 

  27. Rebet O, Andremont O, Gerard JL, et al. Preload dependency determines the effects of phenylephrine on cardiac output in anaesthetised patients. A prospective observational study. Eur J Anaesthesiol. 2016;33:638ā€“44.

    ArticleĀ  CASĀ  PubMedĀ  Google ScholarĀ 

  28. Kalmar AF, Allaert S, Pletinckx P, et al. Phenylephrine increases cardiac output by raising cardiac preload in patients with anesthesia induced hypotension. J Clin Monit Comput. 2018;32:969ā€“76.

    ArticleĀ  CASĀ  PubMedĀ  PubMed CentralĀ  Google ScholarĀ 

  29. Vaquer S, Chemla D, Teboul JL, et al. Influence of changes in ventricular systolic function and loading conditions on pulse contour analysis-derived femoral dP/dtmax. Ann Intensive Care. 2019;9(1):61.

    ArticleĀ  PubMedĀ  PubMed CentralĀ  Google ScholarĀ 

Download references

Acknowledgements

Not applicable.

Funding

No external funding was provided for the planning and conduct of the research or preparation of the manuscript.

Author information

Authors and Affiliations

Authors

Contributions

AA: Assisted with data collection, collation and analysis and manuscript preparation.SS: Assisted with study design, data collection, collation and analysis and manuscript preparation.NF: Assisted with study design, data collection, collation and analysis and manuscript preparation.

Corresponding author

Correspondence to Neal W. Fleming.

Ethics declarations

Ethics approval and consent to participate

For this study, data was collated from two protocols. The first protocol was an evaluation of non-invasive cardiac output monitors in patients undergoing elective cardiac surgery. This study was deemed to not require written, informed consent by the University of California, Davis Institutional Review Board (IRB) as part of a quality improvement project. The second protocol was an IRB-approved prospective evaluation of the effects of diastolic function on the dynamic monitors of cardiac function (Effect of Diastolic Dysfunction on Dynamic Cardiac Monitors, NCT04177225). All patients in this second protocol provided written, informed consent.

Consent for publication

Not applicable.

Competing interests

Neal Fleming has supported contracted research for Acacia Pharma, Tsumera Pharmaceuticals, Haisco Pharmaceuticals, Masimo, Inc. and Edwards LifeSciences and received honoraria from Edwards LifeSciences for invited presentations.

Additional information

Publisherā€™s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the articleā€™s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the articleā€™s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Abdallah, A.C., Song, S.H. & Fleming, N.W. A retrospective study of the effects of a vasopressor bolus on systolic slope (dP/dt) and dynamic arterial elastance (Eadyn). BMC Anesthesiol 24, 257 (2024). https://doi.org/10.1186/s12871-024-02574-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12871-024-02574-x

Keywords