Table 2 Comparison of values for mean arterial pressure, systolic blood pressure and diastolic blood pressure obtained from the non-invasive measurements arm cuffs (NIBP) compared with invasive arterial line (INV) in 29 participants (359 pairs of NIBP-INV measurements). Summary of the calculated bias and 95% limits of agreement with 95% CIs calculated and used to create the modified Bland-Altman plots shown in Additional Figure 3, INV – NIBP

From: Correction: ClearSight™ finger cuff versus invasive arterial pressure measurement in patients with body mass index above 45 kg/m2

INV-NIBP

Bias (SD) mmHg

Lower LOA (95% CI) mmHg

Upper LOA (95% CI) mmHg

Systolic blood pressure

-0.6 (16.8)

-33.5 (-43.8 – -26.4)

32.4 (25.2 – 42.7)

Diastolic blood pressure

-1.7 (10.6)

-22.5 (-28.5 – -18.2)

19.1 (14.9 – 25.2)

Mean arterial pressure

-3.6 (10.8)

-24.9 (-30.8 – -20.7)

17.6 (13.5 – 23.5)

  1. INV invasive measurements, NIBP non-invasive blood pressure measurements, LOA limits of agreement