Fig. 3 | BMC Anesthesiology

Fig. 3

From: Correction: ClearSight™ finger cuff versus invasive arterial pressure measurement in patients with body mass index above 45 kg/m2

Fig. 3

A. Modified Bland-Altman plot for systolic blood pressure (SBP). The plot shows the agreement between measurements from invasive arterial monitoring (SBP-INV) and the non-invasive arm cuffs (SBP-NIBP). B. Modified Bland-Altman plot for diastolic blood pressure (DBP). The plot shows the agreement between measurements from invasive arterial monitoring (DBP-INV) and the non-invasive arm cuffs (DBP-NIBP). C. Modified Bland-Altman plot for mean arterial pressure (MAP). The plot shows the agreement between measurements from invasive arterial monitoring (MAP-INV) and the non-invasive arm cuffs (MAP-NIBP). Participants’ multiple measurements are presented individually as participant number. The middle horizontal line indicates the bias, the bottom and top lines are the 95% limits of agreement and shaded regions represent the 95% CIs for the lower and upper limits of agreement