Skip to main content

Comparison of the third-generation streamlined liner of the pharynx airway (SLIPA-3G) with the laryngeal mask airway supreme for laparoscopic cholecystectomy: a randomized prospective study

Abstract

Background

The third-generation streamlined liner of the pharynx airway (SLIPA-3G) is a new-generation supraglottic airway device (SAD) that is non-cuffed and disposable, with a sealing pressure that varies dynamically with the airway pressure. This study compared the SLIPA-3G with the laryngeal mask airway supreme (LMAS) in patients undergoing laparoscopic cholecystectomy.

Methods

Two hundred and twenty patients scheduled for laparoscopic cholecystectomy were randomly allocated to either the SLIPA-3G group or the LMAS group. Data were collected on the patients’ hemodynamic parameters at different time points, ease of insertion, fiberoptic view, oropharyngeal leak pressure (OLP) at different time points and SAD-related complications.

Results

The mean OLP immediately after device placement in the LMAS group was significantly higher than that of the SLIPA-3G group (31.34 ± 6.99 cmH2O vs.28.94 ± 6.01 cmH2O, P = 0.008, 95% CI 0.62–4.17). The OLPs of the two groups were not significantly different after the induction of a pneumoperitoneum until the end of surgery. The OLP increased gradually through the course of the operation in the SLIPA-3G group (P value = 0.035) but not in the LMAS group (P value = 0.945). There was no significant difference between the two groups in hemodynamic parameters, insertion time and success rate, fiberoptic view and complication rate.

Conclusions

The SLIPA-3G and LMAS were associated with comparable OLPs, hemodynamic parameters, ease of insertion, fiberoptic views and complication rates when used during laparoscopic cholecystectomy. The SLIPA-3G can be used as an effective alternative to the LMAS in patients undergoing laparoscopic surgeries.

Peer Review reports

Background

Supraglottic airway devices (SADs) have been widely used as minimally invasive alternatives to endotracheal tubes (ETTs). SADs possess a number of advantages over ETTs, including ease of insertion, minimal tracheal injury, hemodynamic stability and fewer complications such as sore throat [1]. There were initial concerns over the risk of aspiration associated with the use of SADs during laparoscopic surgeries, as the rise in intra-abdominal pressure may cause difficulties with ventilation and an increased risk of gastric regurgitation [2]. Numerous studies have demonstrated that second-generation SADs with esophageal drainage tubes can be safely used in patients undergoing elective laparoscopic surgeries without an elevated risk of aspiration [3,4,5,6,7].

The OLP is an important indicator of the efficacy of positive-pressure ventilation and airway protection for an SAD [8]. A meta-analysis looking at studies that compared the SLIPA with other conventional laryngeal mask airways (LMAs; Proseal LMA, classic LMA and SoftSeal LMA) found no differences between the two groups in terms of ease of insertion, oropharyngeal leak pressure (OLP) and quality of the fiberoptic view of the larynx [9].

The third-generation SLIPA (SLIPA-3G) is the latest-generation SAD with a sealing pressure that varies dynamically with the airway pressure, i.e. a ‘self-energizing sealing device’ according to the Miller classification system [10]. It has a non-inflatable silicone cuff that is designed to conform to the human pharynx [11], creating a close perilaryngeal seal. The cuff has a thick and firm base and a thin tip. During positive pressure ventilation, the airway pressure equilibrates with the cuff pressure, resulting in tight oropharyngeal sealing. The new features in the SLIPA-3G include an additional passage for the gastric tube to allow for gastric decompression as well as an anti-reflux window on the posterior surface of the device to allow for rapid evacuation of regurgitated material from the oral cavity (Fig. 1). These unique features may provide additional protection against aspiration and airway obstruction due to regurgitated gastric content. In addition, it is manufactured with an in-built bite block and a reinforced shaft that allows for fiberoptic examinations. This is one of the first studies to evaluate the clinical application of the SLIPA-3G in laparoscopic surgery by comparing its OLP and other clinical parameters with those for the LMAS.

Fig. 1
figure 1

Design and features of the SLIPA-3G. A Illustration of features of the SLIPA-3G. 1, Self-sealing silicone cuff. 2, Silicone airway tube. 3, Airway connector. 4, Gastric tube outlet port. 5, Gastric tube inlet port. 6, Ventilating aperture. B-E Views of different parts of the SLIPA-3G

Methods

Study setting and population

The study was approved by the PLA Northern Theater Command General Hospital Ethics Committee. All methods were carried out in accordance with relevant guidelines and regulations. Written informed consent was obtained from all participants. A total of 220 adults (American Society of Anesthesiologists physical status I-III) aged 18–70 years who were scheduled for elective laparoscopic cholecystectomy under general anesthesia were enrolled from August 2020 to December 2020. The exclusion criteria were (1) body mass index (BMI) > 30 kg/m2; (2) inadequate fasting (< 8 h); (3) suspected difficult airway; (4) oropharyngeal infection or other pathology; (5) latex allergy; and (6) emergency surgery.

Study protocol

The participants were randomly allocated into one of two equal groups, the SLIPA-3G group and the LMAS group, based on numbers generated from a random number table. The random numbers were kept in sealed envelopes that were opened in the operating room by the anesthesiologist, just before the induction of general anesthesia. Both the investigators and the patients were blinded to the randomization and allocation.

Standard monitoring (pulse oximetry, electrocardiography, non-invasive blood pressure) and bispectral index (BIS) monitoring were applied in the operating room. All patients were premedicated with dexmedetomidine infusion at 1 µg/kg/hr for 10 min. Following preoxygenation with 100% oxygen for 3 min, anesthesia was induced with sufentanil 0.4–0.6 µg/kg, propofol 1–2 mg/kg and rocuronium 0.6 mg/kg. After loss of the lash reflex and a BIS value less than 50, the lubricated SLIPA-3G or LMAS was inserted and mechanical ventilation was initiated. In the SLIPA-3G group, the size of the device was selected by matching the width of the patient’s thyroid cartilage with that of the SLIPA-3G [12]. The anesthesiologist tilted the patient’s head using the left hand while inserting the toe of the SLIPA-3G towards the posterior pharynx using the right hand. With firm pressure applied on the chamber against the patient’s hard palate, the device was pushed down at the junction of the stem and chamber until resistance was felt. The entire induction process was performed by a single attending anesthesiologist.

Effective ventilation was confirmed with the presence of bilateral chest expansion and a square capnography waveform. If effective ventilation could not be confirmed, the SAD position was adjusted by rotating the device or changing its depth of insertion, or a difference size was used. After three unsuccessful attempts at placing either SAD, endotracheal intubation was performed and failed SAD placement was documented. A fiberoptic assessment of the glottic view was performed after the confirmation of effective ventilation.

Anesthesia was maintained using a balanced technique with inhaled sevoflurane at an end-tidal concentration of 2–4% in 60% oxygen and air, combined with propofol infusion at 50–150 µg/kg/min and remifentanil infusion at 0.1–0.2 µg/kg/min. Additional rocuronium was given as necessary to maintain adequate surgical relaxation. All patients were mechanically ventilated using the volume-controlled mode, with a tidal volume of 6–8 ml/kg and respiratory rate of 10–12 breaths per minute to maintain an end-tidal carbon dioxide reading in the range of 35–45 mmHg. At the end of the operation, neuromuscular blockade was reversed with neostigmine 2.0 mg and atropine 1.0 mg. The SAD was removed when the patient was awake and able to obey commands.

Data collection

The following data were recorded:

  1. 1.

    Patient demographics: age, gender, height, weight, BMI.

  2. 2.

    Hemodynamic parameters: mean arterial blood pressure (MAP), heart rate (HR), and BIS before anesthesia induction (T0), immediately after SAD insertion (T1), five minutes after induction of the pneumoperitoneum (T2), five minutes after cessation of the pneumoperitoneum (T3), when the patient regained consciousness (T4) and immediately after SAD removal (T5).

  3. 3.

    Ease of insertion: success rate of first insertion, number of attempts, time taken for adequate placement (from jaw opening to effective supraglottic ventilation).

  4. 4.

    Fiberoptic view: The best views assessed from the tip of the orifice of the SLIPA-3G or LMAS were scored from 1 to 4 as described by Brimacombe and Berry as follows: 4, only vocal cords seen; 3, vocal cords and posterior epiglottis seen; 2, vocal cords and anterior epiglottis seen; 1, vocal cords not seen [13].

  5. 5.

    OLP – primary outcome: The OLP was measured using the “audible noise” method that was first described by Keller et al. [8]. When the LMAS was used, the cuff was inflated to a pressure of 60 cmH2O using a cuff manometer. With the adjustable pressure limiting valve of the anesthetic circuit closed at 70 cmH2O (manual ventilation mode) and a fresh gas flow rate of 3 L/min, the pressure at which an air leak was heard over the patient’s mouth was measured using (1) a calibrated aneroid cuff manometer attached to the proximal end of the SAD via a connecting tube and (2) the ventilator (i.e. airway pressure displayed on the anesthetic machine). The average of the two values measured with (1) and (2) was recorded as the OLP. The OLP was measured during T1, T2, T3 and T4.

  6. 6.

    Post-removal complications: The SAD was examined for the presence of blood stains once removed. The participants were interviewed by an independent, blinded investigator on the following day about any SAD-related adverse effects including sore throat and dysphagia.

Statistical analyses

Sample size was calculated based on a pilot study conducted at our center, which found that the OLP of the LMAS was 28.1 ± 6.3 cmH2O and that of the SLIPA-3G was 31.1 ± 6.6 cmH2O. In order to detect an inter-device OLP difference of 3 cmH2O with a power of 0.9 and a type I error of 0.05, a sample size of 99 subjects per group would be required. A total of 110 patients were enrolled in each group to allow for potential dropouts. The Statistical Package for Social Sciences (SPSS) software (version 26 for Windows, SPSS Inc., Chicago, IL, USA) was used for all statistical analyses. Data were expressed as mean ± standard deviation, unless stated otherwise. Unpaired t-test was used to compare numerical data. Categorical variables were compared using the Chi-squared or Fisher’s exact test. Statistical significance was defined as a P value < 0.05.

Results

Sample

We recruited 220 patients for the study, among which six in the SLIPA-3G group and three in the LMAS group had endotracheal intubation performed after three unsuccessful attempts to place the SAD. These patients were excluded in the final analyses. There were no significant differences between groups in demographic data (Table 1).

Table 1 Demographic, anesthetic, and surgical data

Primary outcome

Immediately after insertion, the mean OLP of the LMAS was significantly higher than that of the SLIPA-3G by 2.4 cmH2O (P value = 0.008). However, there was no significant difference in OLP between the two groups throughout the rest of the surgery (Table 2). Within-group analysis revealed that the OLP of the SLIPA-3G increased significantly as the surgery progressed from T1 to T4 (P value = 0.035). Such variation in OLP was not observed in the LMAS group (P value = 0.945).

Table 2 Oropharyngeal leak pressure (cmH2O)

Secondary outcomes

There were no significant differences between the SLIPA-3G and LMAS with regard to the patients’ hemodynamic parameters (Table 3), ease of insertion and fiberoptic views (Table 4). Placement of the SLIPA-3G was successful on the first attempt 67.3% of the time, compared to 71.0% for insertion of the LMAS. It took an average of 18.83 s and 16.46 s respectively to properly position a SLIPA-3G or LMAS.

Table 3 Hemodynamic parameters
Table 4 Insertion parameters

After device removal, blood stains were noted in 7.7% of cases in the SLIPA-3G group and in 7.5% of cases in the LMAS group. Thirteen patients in the SLIPA-3G group and 11 in the LMAS group complained of a sore throat. Two patients in the LMAS group had dysphagia, which was not experienced by any patient in the SLIPA-3G group. There was no significance difference in the overall complication rates between the groups (Table 5). Gastric aspiration was not observed in either group.

Table 5 Post-removal complications

Discussion

Owing to its unique design, the SLIPA has a few potential advantages over conventional SADs. The device conforms to the hypopharynx, negating the need for an inflatable cuff. Without a sealed cuff, the interior of the bowl of the SLIPA is in continuous with the patient’s airway. As a result, the pressure exerted by the device on the upper airway mucosa equilibrates with the patient’s airway pressure. Based on the same principle, the SLIPA’s OLP may vary with the patient’s airway pressure. This was indeed observed in our study where the SLIPA-3G’s OLP increased gradually after induction of the pneumoperitoneum. Although the LMAS’s group had higher OLP immediately after insertion, there was no significant variation in the OLP throughout the course of the surgery. Pneumoperitoneum decreases chest wall compliance through the lower side. There is, as a consequence, an increase in airway pressure. To attain a similar transpulmonary pressure, this “new” (higher) pressure therefore fills the SLIPA cuff with extra volume, resulting in a stronger seal [14]. Another proposed explanation for the rise in SLIPA’s OLP over-time is the softening of the device at body temperature, thereby producing a more effective perilaryngeal seal [15]. Such dynamic variation in the SLIPA’s OLP may provide unique benefits in situations where the patient’s airway pressure may increase intraoperatively e.g. during laparoscopic surgery.

SAD insertion is known to cause less hemodynamic disturbances compared to endotracheal intubation [16]. In this study, we found no significant difference in the patients’ hemodynamic parameters at different time points between the two groups. We also observed a general drop in the patients’ MAP, HR and BIS values after induction and placement of SAD in both groups. This study’s results agree with previous study results that the SLIPA-3G, like the LMAS, does not cause significant hemodynamic responses during insertion.

The SLIPA is designed based on the human pharyngeal anatomy [11]. Combined with a dynamically equilibrating pressure between the device and the upper airway mucosa, which is unique to the SLIPA-3G, upper airway injuries caused by over-inflation of SAD cuffs such as sore throat [17] and dysphagia may potentially be reduced. In this study, we observed similar complication rates between the SLIPA-3G and LMAS groups. However, our study has insufficient power to detect any difference in postoperative complications. Further studies will be required to evaluate the safety of the device. The device is smaller than LMAS, potentially facilitating easy insertion. Our study showed that the SLIPA-3G had a similar insertion time, success rate and fiberoptic views as the LMAS in the hands of experienced anesthesiologists. Previous studies have found that the SLIPA is easier to insert for novices compared to other conventional SADs [18,19,20]. The SLIPA may therefore be useful as a rescue airway in resuscitation scenarios, where the operator may not be an anesthesiologist or have much experience with the insertion of SADs.

Conclusions

In our study, we found that the SLIPA-3G, in comparison to the LMAS, provide for a similar OLP after induction of the pneumoperitoneum and stable patient hemodynamic parameters during laparoscopic cholecystectomy. The SLIPA-3G could be easily inserted with an overall success rate of 94.5% after three attempts. It was associated with similar fiberoptic views and complication rates as the LMAS. Therefore, the SLIPA-3G is an effective SAD that can be used in mechanically ventilated patients undergoing laparoscopic surgeries.

Availability of data and materials

The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.

Abbreviations

SLIPA-3G:

Streamlined Pharynx Airway Liner Airway-3 g

LMAS:

Laryngeal Mask Airway Supreme

OLP:

Oropharyngeal Leakage Pressure

SAD:

Supraglottic Airway Device

ETT:

Endotracheal tube

BMI:

Body Mass Index

MAP:

Mean Arterial Pressure

HR:

Heart Rate

BIS:

Bispectral Index

References

  1. Karaaslan E, Akbas S, Ozkan AS, Colak C, Begec Z. A comparison of laryngeal mask airway-supreme and endotracheal tube use with respect to airway protection in patients undergoing septoplasty: a randomized, single-blind, controlled clinical trial. BMC Anesthesiol. 2021;21:5.

    Article  CAS  Google Scholar 

  2. Cajander P, Edmark L, Ahlstrand R, Magnuson A, de Leon A. Effect of positive end-expiratory pressure on gastric insufflation during induction of anaesthesia when using pressure-controlled ventilation via a face mask: A randomised controlled trial. Eur J Anaesthesiol. 2019;36:625–32.

    Article  CAS  Google Scholar 

  3. Kim D, Park S, Kim JM, Choi GS, Kim GS. Second generation laryngeal mask airway during laparoscopic living liver donor hepatectomy: a randomized controlled trial. Sci Rep. 2021;11:3532.

    Article  CAS  Google Scholar 

  4. Parikh SS, Parekh SB, Doshi C, Vyas V. ProSeal Laryngeal Mask Airway versus Cuffed Endotracheal Tube for Laparoscopic Surgical Procedures under General Anesthesia: A Random Comparative Study. Anesth Essays Res. 2017;11:958–63.

    Article  Google Scholar 

  5. Cook TM, Kelly FE. Time to abandon the ‘vintage’ laryngeal mask airway and adopt second-generation supraglottic airway devices as first choice. Br J Anaesth. 2015;115:497–9.

    Article  CAS  Google Scholar 

  6. Beleña JM, Ochoa EJ, Núñez M, Gilsanz C, Vidal A. Role of laryngeal mask airway in laparoscopic cholecystectomy. World J Gastrointest Surg. 2015;7:319–25.

    Article  Google Scholar 

  7. Schmutz A, Bohn E, Spaeth J, Heinrich S. Comprehensive evaluation of manikin-based airway training with second generation supraglottic airway devices. Ther Clin Risk Manag. 2019;15:367–76.

    Article  Google Scholar 

  8. Keller C, Brimacombe JR, Keller K, Morris R. Comparison of four methods for assessing airway sealing pressure with the laryngeal mask airway in adult patients. Br J Anaesth. 1999;82:286–7.

    Article  CAS  Google Scholar 

  9. Choi GJ, Kang H, Baek CW, Jung YH, Woo YC, Kim SH, et al. Comparison of streamlined liner of the pharynx airway (SLIPA ™) and laryngeal mask airway: a systematic review and meta-analysis. Anaesthesia. 2015;70:613–22.

    Article  CAS  Google Scholar 

  10. Michálek P, Miller DM. Airway management evolution - in a search for an ideal extraglottic airway device. Prague Med Rep. 2014;115:87–103.

    Article  Google Scholar 

  11. Hein C, Plummer J, Owen H. Evaluation of the SLIPA (streamlined liner of the pharynx airway), a single use supraglottic airway device, in 60 anaesthetized patients undergoing minor surgical procedures. Anaesth Intensive Care. 2005;33:756–61.

    Article  CAS  Google Scholar 

  12. Miller DM, Camporota L. Advantages of ProSeal and SLIPA airways over tracheal tubes for gynecological laparoscopies. Can J Anaesth. 2006;53:188–93.

    Article  Google Scholar 

  13. Brimacombe J, Berry A. A proposed fiber-optic scoring system to standardize the assessment of laryngeal mask airway position. Anesth Analg. 1993;76:457.

    CAS  PubMed  Google Scholar 

  14. Sorbello M, Micaglio M, Zdravkovic I, Gaçonnet C, Skinner M. Pressure, volume and temperature: Boyle’s law rules airways. Minerva Anestesiol. 2018;84:1112–4.

    PubMed  Google Scholar 

  15. Kang H, Kim DR, Jung YH, Baek CW, Park YH, In OhJ, et al. Pre-warming the Streamlined Liner of the Pharynx Airway (SLIPA) improves fitting to the laryngeal structure: a randomized, double-blind study. BMC Anesthesiol. 2015;15:167.

    Article  Google Scholar 

  16. Jarineshin H, Kashani S, Vatankhah M, Abdulahzade Baghaee A, Sattari S, Fekrat F. Better Hemodynamic Profile of Laryngeal Mask Airway Insertion Compared to Laryngoscopy and Tracheal Intubation. Iran Red Crescent Med J. 2015;17:e28615.

    Article  Google Scholar 

  17. Lange M, Smul T, Zimmermann P, Kohlenberger R, Roewer N, Kehl F. The effectiveness and patient comfort of the novel streamlined pharynx airway liner (SLIPA) compared with the conventional laryngeal mask airway in ophthalmic surgery. Anesth Analg. 2007;104:431–4.

    Article  Google Scholar 

  18. Oh SK, Lim BG, Kim H, Lim SH. Comparison of the clinical effectiveness between the streamlined liner of pharyngeal airway (SLIPA) and the laryngeal mask airway by novice personnel. Korean J Anesthesiol. 2012;63:136–41.

    Article  Google Scholar 

  19. Hein C, Owen H, Plummer J. Randomized comparison of the SLIPA (Streamlined Liner of the Pharynx Airway) and the SS-LM (Soft Seal Laryngeal Mask) by medical students. Emerg Med Australas. 2006;18:478–83.

    PubMed  Google Scholar 

  20. Chen Y, Bai J, Wang R. Insertion methods comparison for The Streamlined Liner of the Pharynx Airway in children by novice. Acta Anaesthesiol Scand. 2018;62:319–27.

    Article  CAS  Google Scholar 

Download references

Acknowledgements

Not Applicable.

Funding

This study was supported by Natural Science Foundation of Liaoning Province (No. 2019-ZD-1063). The funding bodies had no role in the design of the study, the collection, analysis, or interpretation of the data, or writing the manuscript.

Author information

Authors and Affiliations

Authors

Contributions

HF designed and performed the study, he also drafted the manuscript. LL and YD analyzed the data and explained the results. LZ and ZY uploaded the data. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Yugang Diao.

Ethics declarations

Ethics approval and consent to participate

The study was approved by the PLA Northern Theater Command General Hospital Ethics Committee. All methods were carried out in accordance with relevant guidelines and regulations. Written informed consent was obtained from all participants.

Consent for publication

Not Applicable.

Competing interests

None declared.

Additional information

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Fan, H., Li, L., Zhu, L. et al. Comparison of the third-generation streamlined liner of the pharynx airway (SLIPA-3G) with the laryngeal mask airway supreme for laparoscopic cholecystectomy: a randomized prospective study. BMC Anesthesiol 22, 97 (2022). https://doi.org/10.1186/s12871-022-01638-0

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12871-022-01638-0

Keywords