The effect of Ventricular Assist Devices on cerebral autoregulation: A preliminary study

Abstract

Background

The insertion of Ventricular Assist Devices is a common strategy for cardiovascular support in patients with refractory cardiogenic shock. This study sought to determine the impact of ventricular assist devices on the dynamic relationship between arterial blood pressure and cerebral blood flow velocity.

Methods

A sample of 5 patients supported with a pulsatile ventricular assist device was compared with 5 control patients. Controls were matched for age, co-morbidities, current diagnosis and cardiac output state, to cases. Beat-to-beat recordings of mean arterial pressure and cerebral blood flow velocity, using transcranial Doppler were obtained. Transfer function analysis was performed on the lowpass filtered pressure and flow signals, to assess gain, phase and coherence of the relationship between mean arterial blood pressure and cerebral blood flow velocity. These parameters were derived from the very low frequency (0.02-0.07 Hz), low frequency (0.07-0.2 Hz) and high frequency (0.2-0.35 Hz).

Results

No significant difference was found in gain and phase values between the two groups, but the low frequency coherence was significantly higher in cases compared with controls (mean ± SD: 0.65 ± 0.16 vs 0.38 ± 0.19, P = 0.04). The two cases with highest coherence (~0.8) also had much higher spectral power in mean arterial blood pressure.

Conclusions

Pulsatile ventricular assist devices affect the coherence but not the gain or phase of the cerebral pressure-flow relationship in the low frequency range; thus whether there was any significant disruption of cerebral autoregulation mechanism was not exactly clear. The augmentation of input pressure fluctuations might contribute in part to the higher coherence observed.

Peer Review reports

Background

Ventricular assist devices (VAD) are mechanical pumps that replace or augment left and/or right ventricular function in cases of refractory cardiogenic shock. A number of approaches are currently taken related to the indications of these devices: VAD can be used as a bridge to heart transplantation, as a bridge to myocardial recovery leading in some cases to their prolonged use with meaningful survival and improved quality of life [1]. Recently VAD have also begun to be used as a "bridge to destination" that is, they are the final plan for the patient, being used for many years, until the patient succumbs.

Fundamental differences regarding cardiac output and systemic circulation distinguish two main types of VAD: pulsatile and continuous-flow VAD. The main advantages of continuous-flow VAD being the self-contained nature, not requiring a pneumatic driver, longevity, lack of bearing contacting with blood and absence of artificial valves with theoretically smaller thrombogenic surface [2]. However, the effects of non-pulsatile perfusion on end-organ function remain controversial [35]. Pulsatile circulation and its effects on systemic vascular resistances have been related to the improvement of microcirculation and endothelial integrity [6, 7]; reduction in splanchnic perfusion and reduction of intestinal edema [8]; improvement of the cerebral haemodynamics and cerebrospinal fluid drainage [2] and the maintenance of neuro-endocrine cascades, specifically within the renin-angiotensine system and catecholamine release [5].

Despite the use of pulsatile VADs, non-homogeneous output is often generated as pulsatile VADs eject once the pre-established filling volume (stroke volume) has been reached. Therefore, the VAD ejection rate varies depending on preload and systemic resistance. Frequently there is a variable degree of persistent native cardiac contractibility, leading to asynchrony, and irregularities in arterial blood pressure waveform (Figure 1). In such situations of circulatory irregularity, end-organ perfusion such as cerebral blood flow may require an intact autoregulation to ensure stable microcirculation.

Figure 1
figure 1

Real time, beat-to-beat traces of arterial blood pressure (BP) and cerebral blood flow velocity (CBFV) with a ventricular assist device (VAD). Upper channel: arterial BP waveform in a patient supported with a VAD, showing irregular fluctuations; middle channel: CBFV (insonated at the level of middle cerebral artery) with fluctuations transmitted from arterial BP; lower channel: electrocardiogram (ECG).

Cerebral autoregulation is the mechanism by which cerebral blood flow (CBF) is maintained despite changes in cerebral perfusion pressure (CPP). Cerebral autoregulation mediates states of hyperemia and ischemia to avoid vasogenic edema or infarction respectively [9]. Impaired autoregulation has been regarded as a risk factor associated with adverse neurological outcome after cardiac surgery [10, 11]. As a dynamic phenomenon, cerebral autoregulation may respond to spontaneous and induced changes in arterial blood pressure (BP) such as those occurring with pulsatile VADs [12, 13]. Cerebral autoregulation has been extensively studied using transcranial Doppler (TCD) which measures cerebral blood flow velocities (CBFV) as a surrogate of CBF [14, 15] using a variety of methods [16]. From all described methods, transfer function analysis (TFA) enables the analysis of phase shift, gain and coherence between two signals (arterial BP as input and CBFV as output) at a range of frequencies, and has the advantage of being applicable for continuous and non-invasive testing of cerebral autoregulation at the bedside.

Rider and coworkers assessed cerebral autoregulation in patients supported with non-pulsatile VADs, by exposing them to dynamic maneuvers such as head-up tilting and measuring the change in CBFV. They found that cerebral autoregulation was impaired, suggesting that circulatory pulsatility is crucial for the maintenance of cerebral autoregulation [17]. However, their study occurred during the acute phase of the disease, after the insertion of a non-pulsatile VAD and prior to any myocardial "modeling" [18] could have occurred. Some authors have demonstrated that even with the use of non-pulsatile VAD, if a recovery time is allowed, CBF shows recovery of its pulsatility [2], attributing this finding to overall myocardial recovery and specifically right ventricular recovery. Whilst previous study examined the effects of non-pulsatile VAD on the regulation of steady state CBF, this study is the first to investigate the effects of pulsatile VAD, which generates irregular pressure waveform patterns, on the dynamic cerebral pressure-flow relationship by applying the cross-spectral TFA technique.

Methods

Institutional Ethics Committee approval for the performance of the study was granted. All patients or their next of kin gave informed consent prior to enrolment in the study.

A convenience sample of five patients supported with a pulsatile Thoratec VAD (Thoratec corporation, Pleasenton, CA, US) was compared with five control patients, matched for age, comorbidities, current diagnosis and cardiac output state (Table 1). All cases were supported with a left ventricular pulsatile VAD and inotropic drugs for an average of 7 days. All patients were in their acute phase of their disease. Control subjects were in a low output state requiring inotropic or vasopressor support but without the support of VAD. Although their mean arterial blood pressure (MAP) was similar to the VAD cases, their native left ventricular ejection fraction (LV EF) was better. All patients in the control group survived, whereas 2 of the VAD cases died (Table 2).

Table 1 Demographics and patients' characteristics
Table 2 Therapy and clinical variables

We recorded at least 5 minutes of data under resting conditions in all subjects. Simultaneous beat-to-beat recordings of BP and cerebral blood flow velocity (CBFV) waveforms were sampled using a data acquisition unit (ADInstruments, Australia). The BP waveform was acquired from an intra-arterial catheter; CBFV of middle cerebral artery (MCA) was measured using a transcranial Doppler device with a 2 MHz probe and a power of 100 mW/cm2 (DWL, Germany). CBFV of middle cerebral arteries (MCAs) were measured using TCD following referenced criteria at the temporal acoustic window [15]. Both MCAs were insonated and the side with best acoustic characteristics chosen for study. Intra-patient variability was minimized by using only one investigator formally trained in TCD [16]. Stability of the insonated vessel diameter was assumed by maintaining a stable partial pressure of arterial carbon dioxide (pCO2) during measurements.

Therapeutic and clinical variables were recorded at the moment of data acquisition. This study was merely observational and did not interfere with the treating physician's management plan.

Spectral Analysis

For the assessment of cerebral autoregulation, this study used TFA based on frequency domain cross-spectral analysis. TFA assesses the relationship between two signals in the frequency domain and yields three interpretable parameters (i.e. gain, phase, and coherence). Gain is the indicator of the magnitude with which the change of output signal (i.e. CBFV) is caused by the change of input signal (i.e. BP). In the context of cerebral autoregulation analysis, a small gain indicates that cerebral blood flow does not change significantly when blood pressure changes, indicating that the cerebral autoregulatory mechanisms are intact. Phase shift relates to the temporal lag between BP and CBFV at each frequency. Zero phase lag signifies synchronous fluctuations, whilst positive phase suggests CBFV leading BP, and negative phase suggests BP leading CBFV.

The gain and phase metrics, however, need to be interpreted in the context of the cross-spectral coherence, which is an estimation of the linear correlation between the input and output signals at particular frequencies. Coherence varies between 0 and 1; where 0 indicates no linear relationship and 1 indicates perfect linear relationship. It has been suggested that an increase in coherence may be indicative of a blunted cerebral autoregulation [21]. A low coherence, however, can be interpreted as presence of external noise/input, or nonlinear/lack of relationship between input and output.

In this study, spectral analysis was performed on 5 min artifact-free segments of continuous CBFV and BP signals. Signals were downsampled to 1 Hz after appropriate anti-aliasing lowpass filtering, with any slow trend removed by cubic spline detrending. The frequency spectra and transfer function were obtained using the Welch method [21]. This involved subdividing the signal into 120s segments with 75% overlap (resulting in 7 segments), multiplying each segment with a Hanning window, then performing a Fast Fourier Transform (FFT), and finally averaging to give the spectra. Defining the autospectra of BP and CBFV as S xx(f) and S yy(f) (with f denoting frequency), the cross-spectrum of BP and CBFV, S xy(f), was computed as the product of S xx*(f) and S yy(f) (asterisk denotes the complex conjugate). The transfer function from BP to CBFV was computed as H(f) = S xy(f)/S xx(f), and the gain magnitude and phase angle of the transfer function was obtained accordingly. The magnitude-squared coherence function was computed as γ2(f) = |S xy(f)|2/S xx(f)S yy(f), for detecting linear correlation between the spectral components in the two signals. Coherence ranged from 0 (lack of linear correlation) to 1 (perfect linear relationship).

The spectral powers of BP and CBFV and the mean values of the transfer function gain, phase and coherence were calculated in the very low frequency (VLF, 0.02-0.07 Hz), low frequency (LF, 0.07-0.20 Hz) and high frequency (HF, 0.20-0.35 Hz) ranges as previously defined [22]. Unpaired Student's t-test was performed to compare the variables between the VAD and the control groups. P < 0.05 was considered statistically significant.

Results

The patient characteristics are presented in table 1 and 2. No significant difference in MAP, pCO2 and LVEF was found between the VAD and the control groups. The levels of pCO2 were maintained within normal ranges and stable throughout the study, thus the effect of CO2 on cerebral vessel was minimised. LVEF was generally lower for the VAD cases, whichwas expected as these were patients with baseline refractory cardiogenic shock who required a VAD for life support. However the difference did not reached statistical significance.

The results from spectral and cross-spectral transfer function analysis of MAP and CBFV were presented in table 3 and 4. Display of gain, phase and coherence for a representative case and control are shown in figures 2 and 3 respectively. No significant difference was found between the VAD and the control groups, apart from a significantly higher LF coherence between MAP and CBFV in the VAD cases (P = 0.04).

Table 3 Power spectrum analysis of mean arterial pressure (MAP) and mean cerebral blood flow velocity (CBFV) in ventricular assist device (VAD) cases and controls
Table 4 Transfer function analysis (TFA) of mean arterial pressure (MAP) and mean cerebral blood flow velocity (CBFV) in ventricular assist device (VAD) cases and controls
Figure 2
figure 2

Transfer function analysis (TFA) of mean arterial pressure (MAP) and cerebral blood flow velocity (CBFV) in a patient with ventricular assist device (VAD). The gain, phase and coherence spectra of a representative case with VAD were shown.

Figure 3
figure 3

Transfer function analysis (TFA) of mean arterial pressure (MAP) and cerebral blood flow velocity (CBFV) in a control patient. The gain, phase and coherence spectra of a representative control were shown.

Discussion

In this study, the cross-spectral transfer function analysis technique was applied to study the dynamic relationship between systemic BP and CBFV in patients using pulsatile VAD. The rationale was to describe any potential alteration of cerebral autoregulation function associated with the use of VAD, as the long term use of VAD may lead to impaired cerebral autoregulation and worse neurological outcomes.

The key finding of the study was the higher coherence between MAP and CBFV in the VAD patients compared with the controls, at the LF range. A low coherence between MAP and CBFV (<0.5)indicates a lack of linear relationship between pressure and flow at the particular frequency range, and can be attributed to the presence of an intact cerebral autoregulation that introduces nonlinearity relationship [21, 22]. It has been suggested that the complex nonlinear behavior of the cerebral vasculature might be responsible for the low coherences at the VLF and LF ranges [2426]. The augmented LF coherence in the VAD patients, on the other hand, might suggest a lower degree of cerebral autoregulation, possibly due to disruption of autoregulatory mechanisms by the use of VAD. However, one potential limitation to this interpretation was that, although no significant difference in the MAP power was observed between the two groups, the two VAD patients with the highest coherence (~0.8) also had much higher spectral power in MAP than the rest of the group. It has been suggested that an increased input pressure change might lead to an increase in coherence, via an improved "signal-to-noise" ratio [27]. This effect might contribute in part to the higher coherence in the VAD group.

The lack of differences in TFA gain and phase between the VAD and the control group also raised questions whether there was significant disruption of cerebral autoregulation by the use of pulsatile VAD. Alterations in cerebral autoregulation function by pathological conditions (such as stroke and autonomic failure [28, 29]) are typically associated with changes in gain and/or phase, which were not observed in the current study. Neverthelss, it appeared that the highpass filtering property of the cerebral circulation, characterised by smaller gain at the lower frequencies (VLF) and an increase in gain towards the higher frequencies (HF) [21, 27], was more apparent in the control group compared with the VAD group. It would therefore still be possible that gain properties of cerebral autoregulation might have changed in the VAD patients, although the interpretation of the gain parameter would have been limited somewhat by the low coherences in the control patients.

Methodological considerations and limitations

In this study, direct assessment of CBF was not feasible as the use of non-imaging TCD does not facilitate the measurement of the cerebral vessel cross-sectional area. Instead, there is a global consensus supporting the use of CBFV as a surrogate for CBF, provided the vessel diameter remains stable during the study [30]. Among all factors intervening in changes of vessel diameter and therefore determining CBF [23], pCO2 is directly related with vessel diameter and was maintained stable and within normal values, during patient recruitment.

For TCD recordings, only the MCA with better acoustic properties was recorded and analyzed. Although spatial heterogeneity of cerebral perfusion as well as interhemispheric differences has been described [30]; the endpoint in this study was to ensure the best transcranial Doppler recordings in order to minimize the signal-to-noise ratio and increase data reliability [30].

No significant change in gain and phase was found between the two groups in this study, but the small population recruited could have contributed to the lack of statistical significance, thus further studies with larger sample size would be desirable.

Conclusion

The use of pulsatile VAD affected the coherence but not the gain or phase of the cerebral pressure-flow relationship in the low frequency range, thus whether there was any significant disruption of cerebral autoregulation mechanism was not clear. The augmentation of input pressure fluctuations might contribute in part to the higher coherence observed. Given the absence of all conditions that define autoregulation, these results should be regarded as preliminary data, and further studies, employing bigger samples, are warranted.

References

  1. Rose EA, Gelijns AC, Moskowitz AJ: Long-term mechanical left ventricular assistance for end-stage heart failure. N Engl J Med. 2001, 345: 1435-43. 10.1056/NEJMoa012175.

    Article  CAS  PubMed  Google Scholar 

  2. Potapov EV, Loebe M, Nasseri BA: Pulsatile flow in patients with a novel nonpulsatile implantable ventricular assist device. Circulation. 2000, 102: III183-7.

    Article  CAS  PubMed  Google Scholar 

  3. Taylor KM, Bain WH, Davidson KG, Turner MA: Comparative clinical study of pulsatile and non-pulsatile perfusion in 350 consecutive patients. Thorax. 1982, 37: 324-30. 10.1136/thx.37.5.324.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  4. Shevde K, DeBois WJ: Pro: pulsatile flow is preferable to nonpulsatile flow during cardiopulmonary bypass. J Cardiothorac Anesth. 1987, 1: 165-8. 10.1016/0888-6296(87)90012-3.

    Article  CAS  PubMed  Google Scholar 

  5. Canivet JL, Larbuisson R, Damas P: Plasma renin activity and urine beta 2-microglobulin during and after cardiopulmonary bypass: pulsatile vs non-pulsatile perfusion. Eur Heart J. 1990, 11: 1079-82.

    CAS  PubMed  Google Scholar 

  6. Fukae K, Tominaga R, Tokunaga S, Kawachi Y, Imaizumi T, Yasui H: The effects of pulsatile and nonpulsatile systemic perfusion on renal sympathetic nerve activity in anesthetized dogs. J Thorac Cardiovasc Surg. 1996, 111: 478-84. 10.1016/S0022-5223(96)70459-2.

    Article  CAS  PubMed  Google Scholar 

  7. Busse R, Fleming I: Pulsatile stretch and shear stress: physical stimuli determining the production of endothelium-derived relaxing factors. J Vasc Res. 1998, 35: 73-84. 10.1159/000025568.

    Article  CAS  PubMed  Google Scholar 

  8. Gaer JA, Shaw AD, Wild R: Effect of cardiopulmonary bypass on gastrointestinal perfusion and function. Ann Thorac Surg. 1994, 57: 371-5. 10.1016/0003-4975(94)90999-7.

    Article  CAS  PubMed  Google Scholar 

  9. Aaslid R, Lindegaard KF, Sorteberg W, Nornes H: Cerebral autoregulation dynamics in humans. Stroke. 1989, 20: 45-52.

    Article  CAS  PubMed  Google Scholar 

  10. Newman MF, Kirchner JL, Phillips-Bute B: Longitudinal assessment of neurocognitive function after coronary-artery bypass surgery. N Engl J Med. 2001, 344: 395-402. 10.1056/NEJM200102083440601.

    Article  CAS  PubMed  Google Scholar 

  11. Roach GW, Kanchuger M, Mangano CM: Adverse cerebral outcomes after coronary bypass surgery. Multicenter Study of Perioperative Ischemia Research Group and the Ischemia Research and Education Foundation Investigators. N Engl J Med. 1996, 335: 1857-63. 10.1056/NEJM199612193352501.

    Article  CAS  PubMed  Google Scholar 

  12. Czosnyka M, Piechnik S, Richards HK, Kirkpatrick P, Smielewski P, Pickard JD: Contribution of mathematical modelling to the interpretation of bedside tests of cerebrovascular autoregulation. J Neurol Neurosurg Psychiatry. 1997, 63: 721-31. 10.1136/jnnp.63.6.721.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  13. Giller CA: The frequency-dependent behavior of cerebral autoregulation. Neurosurgery. 1990, 27: 362-8. 10.1097/00006123-199009000-00004.

    Article  CAS  PubMed  Google Scholar 

  14. Newman MF, Croughwell ND, Blumenthal JA: Effect of aging on cerebral autoregulation during cardiopulmonary bypass. Association with postoperative cognitive dysfunction. Circulation. 1994, 90: II243-9.

    CAS  PubMed  Google Scholar 

  15. Aaslid R, Markwalder TM, Nornes H: Noninvasive transcranial Doppler ultrasound recording of flow velocity in basal cerebral arteries. J Neurosurg. 1982, 57: 769-74. 10.3171/jns.1982.57.6.0769.

    Article  CAS  PubMed  Google Scholar 

  16. Bellapart J, Fraser JF: Transcranial Doppler assessment of cerebral autoregulation. Ultrasound Med Biol. 2009, 35: 883-93. 10.1016/j.ultrasmedbio.2009.01.005.

    Article  PubMed  Google Scholar 

  17. Ridder D, Markham DW, Fu Q, Pacini E, Bhella P, Minniefield N, Metzemaekers LF, Galbreath MM, Palmer D, Shibata S, Zhang R, Bethea BT, Meyer DM, Patel PC, Drazner MH, Levine BD: Non-pulsatile Left Ventricular Assist Device (LVAD) patients show impaired cerebral autoregulation in response to upright tilt. Circulation. 2009, 120: S725-S726. 10.1161/CIRCULATIONAHA.108.846501.

    Article  Google Scholar 

  18. Zafeiridis A, Jeevanandam V, Houser SR, Margulies KB: Regression of cellular hypertrophy after left ventricular assist device support. Circulation. 1998, 98: 656-62.

    Article  CAS  PubMed  Google Scholar 

  19. Pagani M, Somers V, Furlan R: Changes in autonomic regulation induced by physical training in mild hypertension. Hypertension. 1988, 12: 600-10.

    Article  CAS  PubMed  Google Scholar 

  20. Saul JP, Berger RD, Chen MH, Cohen RJ: Transfer function analysis of autonomic regulation. II. Respiratory sinus arrhythmia. Am J Physiol. 1989, 256: H153-61.

    CAS  PubMed  Google Scholar 

  21. Zhang R, Zuckrman JH: Transfer Function Analysis of dynamic cerebral autoregulation in humans. Am J Physiol. 1998, 274: H233-H241.

    CAS  PubMed  Google Scholar 

  22. Zhang R, Benbehani K: Dynamic pressure-flow relationship of the cerebral circulation during acute increase in arterial pressure. J Physiol. 2009, 587: 2567-2577. 10.1113/jphysiol.2008.168302.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  23. Huber P, Jyoji H: Effect of Contrast material, Hypercapnia, Hyperventilation, Hypertonic Glucose and Papaverine on the Diameter of the Cerebral Arteries: Angiographiic Determination in Man. Investigate Radiology. 1967, 2 (1):

  24. Giller CA: The frequency-dependant behaviour of cerebral autoregulation. Neurosurgery. 1990, 27: 362-368. 10.1097/00006123-199009000-00004.

    Article  CAS  PubMed  Google Scholar 

  25. Panerai RB: Assessment of cerebral pressure autoregulation in humans, a review of measurement methods. Physiol Meas. 1998, 19: 305-338. 10.1088/0967-3334/19/3/001.

    Article  CAS  PubMed  Google Scholar 

  26. Panerai RB, Eames PJ, Potter JF: Multiple coherence of cerebral blood flow velocity in humans. Am J Physiol Heart Circ Physiol. 2006, 291: H251-H259. 10.1152/ajpheart.01348.2005.

    Article  CAS  PubMed  Google Scholar 

  27. Claassen JA, Levine BD, Zhang R: Dynamic cerebral autoregulation during repeated squat-stand maneuvers. J Appl Physiol. 2009, 106: 153-160. 10.1152/japplphysiol.90822.2008.

    Article  PubMed  Google Scholar 

  28. Blaber AP, Bondar RL, Stein F, Dunphy PT, Moradshahi P: Transfer Function Analysis of cerebral autoregulation dynamics in Autonomic failure patients. Stroke. 1997, 28: 1686-1692.

    Article  CAS  PubMed  Google Scholar 

  29. Reinhard M, Roth M, Guschlbauer B, Harloff A, Timmer J, Czosnyka M, Hetzel A: Dynamic cerebral autoregulation in acute ischemic stroke assessed from spontaneous blood pressure fluctuations. Stroke. 2005, 36: 1684-1689. 10.1161/01.STR.0000173183.36331.ee.

    Article  CAS  PubMed  Google Scholar 

  30. Lindegaard KF, Lundar T, Wiberg J, Sjoberg D, Aaslid R, Nornes H: Variations in middle cerebral artery blood flow investigated with noninvasive Transcranial blood velocity measurements. Stroke. 1987, 18: 1025-1030.

    Article  CAS  PubMed  Google Scholar 

Pre-publication history

Download references

Acknowledgements

We acknowledge Dr Daniel Mullany for ensuring patients' availability during study recruitment.

This research was supported by a research grant from the Royal Brisbane Hospital Research Foundation (Protocol 2007/076). Additional funding was provided by the Prince Charles Hospital Research Foundation.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Judith Bellapart.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors' contributions

JB and JFF conceived and designed the study; JB undertook patient screen and data acquisition; JB drafted the manuscript which was reviewed and amended by all other authors. GC and Y-C T undertook data analysis and contributed to its interpretation. AGB undertook statistical analysis. KRD conceived and designed the technical components of the study, also reviewed revised the manuscript. RB and PA reviewed and amended the manuscript. All authors read and approved the final manuscript.

Authors’ original submitted files for images

Below are the links to the authors’ original submitted files for images.

Authors’ original file for figure 1

Authors’ original file for figure 2

Authors’ original file for figure 3

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Cite this article

Bellapart, J., Chan, G.S., Tzeng, YC. et al. The effect of Ventricular Assist Devices on cerebral autoregulation: A preliminary study. BMC Anesthesiol 11, 4 (2011). https://doi.org/10.1186/1471-2253-11-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/1471-2253-11-4

Keywords