Skip to main content

Table 1 Accuracy statistics for Masimo Radical: Rincon et al.results versus Masimo results

From: Correspondence to: performance of the new SmartCardia wireless, wearable oximeter: a comparison with arterial SaO2 in healthy volunteers

 

Ricon et al. BMC

Anesthesiology

(SmartCardia study)

2022 [1]

Masimo

(IRB Approved, FDA Cleared Validation Study on health adults) 2017

Bias (Mean Error)

2.02

-0.01

Precisiom (Standard deviation of error)

4.60

1.16

Arms (root-mean–square of the differences)

5.00

1.16

Number of data pairs

286

1493

Number of Subjects

12

25