Skip to main content

Table 3 Rates of accidental epidural catheter removal by type of catheter connection

From: Accidental epidural catheter removal rates and strength required for disconnection: a retrospective cohort and laboratory study

Removal method

Old group

New group

Taping group

(n = 360)

(n = 182)

(n = 378)

Scheduled

 Count

347 (96.4%)

171 (94.0%)

374 (98.9%)

 Adjusted residual

-0.804

-2.631

2.927

Dislodgement

 Count

6 (1.7%)

1 (0.5%)

3 (0.8%)

 Adjusted residual

1.360

-0.781

-0.717

Disconnection

 Count

7 (1.9%)

10 (5.5%)

1 (0.3%)

 Adjusted residual

-0.021

3.848

-3.094

  1. Values are presented as counts (%) or adjusted residuals
  2. Residual analysis was based on Fisher's exact test
  3. Adjusted residuals smaller than − 1.96 or greater than 1.96 indicate a statistically significant difference at p < 0.05