From: Comparisons of the segments of left-sided double-lumen tracheobronchial tubes as industrial products
Product name | 35 Fr | 37 Fr | ||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Sample number | Lot No | Tracheal lumen | Bronchial lumen | Sample number | Lot No | Tracheal lumen | Bronchial lumen | |||||||||||||
3.0 mm | 3.5 mm | 4.0 mm | 4.5 mm | 3.0 mm | 3.5 mm | 4.0 mm | 4.5 mm | 3.0 mm | 3.5 mm | 4.0 mm | 4.5 mm | 3.0 mm | 3.5 mm | 4.0 mm | 4.5 mm | |||||
Bronch-Cath™ | ① | 201210608X | 〇 | 〇 | 〇 | × × | 〇 | 〇 | 〇 | × × | ⑪ | 201411237X | 〇 | 〇 | 〇 | 〇 | 〇 | 〇 | 〇 | 〇 |
② | 201503168X | 〇 | 〇 | 〇 | × × | 〇 | 〇 | 〇 | × × | ⑫ | 201410348X | 〇 | 〇 | 〇 | 〇 | 〇 | 〇 | 〇 | × × | |
Blue Line™ | ③ | 3227850 | 〇 | 〇 | 〇 | × × | 〇 | 〇 | × × | × × | ⑬ | 3227862 | 〇 | 〇 | 〇 | × × | 〇 | 〇 | 〇 | × × |
④ | 3227851 | 〇 | 〇 | 〇 | × × | 〇 | 〇 | 〇 | × × | ⑭ | 3227865 | 〇 | 〇 | 〇 | × × | 〇 | 〇 | 〇 | × × | |
Bronchopart® | ⑤ | 15GE29J | 〇 | × × | × × | × × | 〇 | 〇 | × × | × × | ⑮ | 16HE33J | 〇 | 〇 | 〇 | × × | 〇 | 〇 | 〇 | × × |
⑥ | 15DE16J | 〇 | × 〇 | × × | × × | 〇 | 〇 | 〇 | × × | ⑯ | 15GE29J | 〇 | 〇 | 〇 | × × | 〇 | 〇 | 〇 | × × | |
SHER-I-BRONCH® | ⑦ | 73L1500302 | 〇 | 〇 | 〇 | × × | 〇 | 〇 | 〇 | 〇 | ⑰ | 73L1500369 | 〇 | 〇 | 〇 | 〇 | 〇 | 〇 | 〇 | 〇 |
⑧ | 73L1500168 | 〇 | 〇 | 〇 | × × | 〇 | 〇 | 〇 | 〇 | ⑱ | 73K1600076 | 〇 | 〇 | 〇 | × × | 〇 | 〇 | 〇 | 〇 | |
Coopdech | ⑨ | B161013X | 〇 | 〇 | 〇 | × × | 〇 | 〇 | 〇 | × × | ⑲ | B161205X | 〇 | 〇 | 〇 | 〇 | 〇 | 〇 | 〇 | 〇 |
⑩ | B161017X | 〇 | 〇 | 〇 | × × | 〇 | 〇 | 〇 | × × | ⑳ | B161209X | 〇 | 〇 | 〇 | 〇 | 〇 | 〇 | 〇 | 〇 |