Comparison | r (95% CI) sample 1 (n = 37) | r (95% CI) sample 2 (n = 35) | r diff sample 1 to sample 2 (95% CI) | r diff p-value* |
---|---|---|---|---|
EXTEM A10 (hPa) vs Quantra CS (hPa) | 0.93 (0.86 to 0.96) | 0.94 (0.88 to 097) | − 0.01 (− 0.08 to 0.05) | 1.00 |
FIBTEM A10 (hPa) vs Quantra FCS (hPa) | 0.92 (0.85 to 0.96) | 0.96 (0.92 to 0.98) | − 0.04 (− 0.11 to 0.01) | 0.60 |
INTEM CT (sec) vs Quantra CT (sec) | 0.71 (0.51 to 0.84) | 0.75 (0.55 to 0.86) | −0.03 (− 0.28 to 0.20) | 1.00 |
HEPTEM CT (sec) vs Quantra CTH (sec) | 0.59 (0.33 to 0.77) | 0.77 (0.59 to 0.88) | −0.18 (− 0.47 to 0.08) | 0.68 |
(EXTEM A10 – FIBTEM A10) (hPa) vs Quantra PCS (hPa) | 0.15 (−0.18 to 0.45) | 0.80 (0.64 to 0.90) | −0.65 (− 0.99 to − 0.33) | 0.0002 |
Platelet count (×1000/μL) vs (EXTEM A10 – FIBTEM A10) (hPa) | 0.70 (0.48 to 0.83) | 0.69 (0.46 to 0.83) | 0.01 (−0.23 to 0.25) | 1.00 |
Platelet count (×1000/μL) vs Quantra PCS (hPa) | 0.37 (0.05 to 0.62) | 0.69 (0.46 to 0.83) | −0.32 (− 0.65 to − 0.01) | 0.28 |