Skip to main content

Table 3 Population kinetic parameters

From: Can perioperative hemodilution be monitored with non-invasive measurement of blood hemoglobin?

Kinetic parameter

Measurement route

Best estimate

95% CI

CV%

Vc (L)

Invasive

3.91

3.27–4.56

8.5

Non-invasive

3.11

2.51–3.71

9.8

k12 (10− 3 min− 1)

Invasive

38.3

30.9–46.0

10.0

Non-invasive

47.0

22.0–72.0

27.1

k21 (10−3 min− 1)

Invasive

23.5

14.0–32.9

20.5

Non-invasive

11.2

(−18.5)–41.0

135.2

k10 (10−3 min−1)

Invasive

9.1

3.7–14.3

29.8

Non-invasive

9.4

2.7–16.1

36.5

  1. Data are based on 30 infusion experiments (Studies 1, 2 and 3)
  2. CI confidence interval, CV between-subject coefficient of variation