Skip to main content

Table 3 Comparison of weight-standardized expiratory tidal volumes in patients not initially easy to mask ventilate

From: The Articulated Oral Airway as an aid to mask ventilation: a prospective, randomized, interventional, non-inferiority study

Variable Guedel vs. AOA p-value
Expiratory tidal volume Mean diff = 0.397 (ml/kg)
CI = (0.134 to 0.660)
 
Inspiratory tidal volume Mean diff = − 0.041 (ml/kg)
CI = (− 0.321 to 0.240)
 
Ease of placement aOdds ratio = 0.312
CI = (0.012 to 8.285)
> 0.999
Normal expiratory waveform Odds ratio = 1.536
CI = (0.238 to 9.926)
0.65
Blood on 1st device (Guedel 1st vs. 2nd) aOdds ratio = 5.769
CI = (0.232 to 143.371)
0.47
  1. a = corrected odds ratio for 0-cell count