Skip to main content

Table 3 Summary of correlation and Bland-Altman analyses according to each pre- and intraoperative measurement time point

From: Perioperative non-invasive versus semi-invasive cardiac index monitoring in patients with bariatric surgery – a prospective observational study

Measurement time point

Data pairs CINexfin vs. CIFloTrac

N

Correlation coeffcient rs

P value

Bias, l/min/m2

SD of bias, l/min/m2

LOA, l/min/m2

PE,

%

1 Baseline

51

0.56

< 0.001

−0.08

0.90

− 1.85-1.68

54.42

2 Anesthesia induction + ATP

52

0.56

< 0.001

0.08

0.60

− 1.09-1.26

55.63

3 Baseline Anesthesia

53

0.38

0.01

−0.05

1.02

−2.06-1.96

89.64

4 Anesthesia + PLR

52

0.74

< 0.001

−0.20

0.50

−1.18-0.79

47.85

5 Anesthesia + Fluid bolus

53

0.82

< 0.001

−0.33

0.42

−1.17-0.50

46.34

6 Baseline PP

54

0.66

< 0.001

−0.29

0.72

−1.69-1.12

68.64

7 PP + ATP

52

0.64

< 0.001

−0.21

0.49

−1.16-0.75

48.88

8 PP + ATP + Fluid bolus

54

0.77

< 0.001

−0.18

0.42

−1.00-0.63

35.95

9 End PP

53

0.65

< 0.001

−0.34

0.47

−1.26-0.57

36.54

10 End Anesthesia + ATP

50

0.64

< 0.001

−0.03

0.78

−1.55-1.49

48.14

  1. ATP Anti-Trendelenburg posture, LOA Limits of agreement, PE Percentage error, PLR Passive leg raising, PP Pneumoperitoneum, SD Standard deviation