Skip to main content

Table 2 Concordance between the TEG6S and TEG5000 systems

From: Assessment of agreement and interchangeability between the TEG5000 and TEG6S thromboelastography haemostasis analysers: a prospective validation study

Parameter

Lin’s CC

RMAR Slope

RMAR Intercept

Intraclass correlation coefficient

Reaction time

0.782 (0.640–0.924)

0.755

0.923

0.787

Kinetic time

0.816 (0.690–0.941)

1.298

−0.934

0.823

Alpha angle

0.791 (0.635–0.946)

1.039

−1.429

0.798

Maximum amplitude

0.896 (0.828–0.964)

0.992

−4.977

0.896

LY30%

0.336 (0.097–0.575)

0.427

0.035

0.335