Skip to main content

Table 1 Optimal balloon volume (ml) in the bench experiment and clinical study

From: Use of esophageal balloon pressure-volume curve analysis to determine esophageal wall elastance and calibrate raw esophageal pressure: a bench experiment and clinical study

 

At atmosphere

In chambers

In patients

VMIN

all 0.2

0.2 (0.2–0.4) [0.2–0.4]

0.3 (0.2–0.4) [0.2–0.6]

VMAX

all 2.0

2.0 (2.0–2.2) [1.6–2.4]

1.6 (1.4–1.8) [1.4–2.2]

VWORK

all 1.8

1.8 (1.8–2.0) [1.4–2.0]

1.2 (1.1–1.4) [0.8–2.0]

VBEST

0.4 (0.4–0.6) [0.4–0.6]

0.4 (0.4–0.6) [0.4–0.8]

1.0 (0.8–1.2) [0.6–1.4]

  1. V MIN and V MAX : the minimal and maximal balloon volume representing the optimal filling volume with the least influence of balloon recoil pressure, V WORK : the difference between VMIN and VMAX, V BEST the best balloon filling volume
  2. Data are shown as median (interquartile range) [range]