Skip to main content

Table 2 Respiratory parameter changes in supine and pneumoperitoneum and Trendelenburg position

From: I-Gel is a suitable alternative to endotracheal tubes in the laparoscopic pneumoperitoneum and trendelenburg position

Ā 

Supine

Pneumoperitoneum and Trendelenburg

i-gel

ETT

p-valuea

i-gel

ETT

p-valueb

Leak Fraction (%)

7.01 [3.73]

5.76 [2.67]

0.341

6.20 [3.49]

6.38 [3.71]

0.883

Leak Volume (ml)

31.99 [14.54]

26.06 [9.62]

0.327

27.66 [1.73]

28.35 [18.18]

0.968

Resistance (cmH2O/l/s)

5.82 [1.18]

10.25 [1.03]

<0.01

7.95 [1.57]

13.13 [1.71]

<0.01

Peak Airway Pressure (cmH2O)

11.44 [1.43]

12.53 [1.68]

0.03

20.75 [2.75]

21.60 [3.24]

0.461

Mean Airway Pressure (cmH2O)

4.67 [0.71]

4.95 [0.46]

0.049

6.73 [1.33]

6.65 [0.98]

0.841

Compliance (ml/cmH2O)

48.45 [7.98]

46.34 [8.13]

0.62

23.02 [4.12]

23.19 [4.64]

0.779

  1. ETT endotracheal tube
  2. Data are median [IQR]
  3. a p-value in comparison with i-gel and ETT in supine
  4. b p-value in comparison with i-gel and ETT in Pneumoperitoneum and Trendelenburg position