Table 2 Logistic regression analysis for right bronchial misplacement of left-sided double-lumen tubes

From: Misplacement of left-sided double-lumen tubes into the right mainstem bronchus: incidence, risk factors and blind repositioning techniques

 

Regression coefficient (SE)

Odds ratio (95 % CI)

P-value

Age

0.01 (0.01)

1.01 (0.99 to 1.04)

0.182

Female

1.33 (0.32)

3.78 (2.03 to 7.04)

<0.001

Height

−0.07 (0.02)

0.93 (0.89 to 0.97)

<0.001

Weight

−0.01 (0.01)

1.00 (0.97 to 1.03)

0.884

Laryngoscopic gradea

   

 2

−0.15 (0.48)

0.86 (0.34 to 2.22)

0.760

 3

0.19 (0.74)

1.21 (0.28 to 5.21)

0.795

Double-lumen tube sizeb

   

 32-Fr

1.59 (0.37)

4.88 (2.35 to 10.17)

<0.001

 35-Fr

0.88 (0.38)

2.42 (1.16 to 5.05)

0.018

RC distance

−0.08 (0.05)

0.93 (0.84 to 1.02)

0.136

LC distnace

−0.08 (0.05)

0.92 (0.83 to 1.02)

0.122

Tracheal diameter

−0.22 (0.06)

0.81 (0.71 to 0.91)

0.001

Right tracheobronchial angle (°)

0.10 (0.07)

1.01 (0.98 to 1.04)

0.456

Left tracheobronchial angle (°)

0.01 (0.08)

1.01 (0.98 to 1.04)

0.495

Right bronchial diameter (mm)

−0.21 (0.07)

0.81 (0.70 to 0.93)

0.003

Left bronchial diameter (mm)

−0.20 (0.08)

0.82 (0.70 to 0.97)

0.017

  1. RC and LC distances: horizontal distances between the carina and the right or left internal borders of the trachea at 3 cm above the carina; Tracheal diameter: sum of the RC and LC distances
  2. aLaryngoscopic grade 1 and bSize of 37-Fr were considered as the reference value for the analysis