Skip to main content

Table 3 Device insertion characteristics, gastric tube insertion characteristics and fiber-optic view

From: Comparisons of clinical performance of guardian laryngeal mask with laryngeal mask airway ProSeal

Variables

Group P (P-LMA)

Group G (G-LMA)

Pvalue

(n = 40)

(n = 40)

 

Successful insertion attempt

First

38 (95%)

39 (97.5%)

1.000

Second

2 (5%)

1 (2.5%)

 

Over all

100%

100%

 

Device insertion time, seconds

15.95 ± 1.84**

13.20 ± 1.82**

<0.0001**

Ease of insertion

   

1

28

30

 

2

10

9

 

3

2

1

 

4

0

0

 

Manipulation

6/40

4/40

0.373

Maneuvers

6/40

4/40

0.373

Successful gastric tube insertion attempt

First

35 (87.5%)

38 (95%)

0.431

Second

5 (12.5%)

2 (5%)

 

Overall

40 (100%)

40 (100%)

 

Gastric tube insertion time, seconds

13.75 ± 1.74**

10.05 ± 1.35**

<0.0001**

Fiber-optic view

   

(1/2/3/4)

14/21/3/2

17/20/2/1

0.836

  1. **P < 0.01, *p < 0.05, P-LMA: ProSeal laryngeal mask airway; G-LMA: Guardia laryngeal mask airway.