Skip to main content

Table 2 Airway sealing pressure, cuff volume, intracuff pressure and pharyngolaryngeal morbidity

From: Comparisons of clinical performance of guardian laryngeal mask with laryngeal mask airway ProSeal

Variables

Group P (P-LMA)(n = 40)

Group G (G-LMA)(n = 40)

Pvalue

Volume of air for cuff inflation to 60 cmH 2 O, ml

13.90 ± 1.19**

22.60 ± 3.27**

<0.0001**

Airway sealing pressure, cmH 2 O

29.41 ± 4.33*

32.36 ± 4.93*

0.04*

Intracuff pressure measurement, cmH 2 O

30 min

58.83 ± 1.14

58.90 ± 1.15

0.823

60 min

58.67 ± 1.09

58.27 ± 1.20

0.182

90 min

58.40 ± 1.23

58.33 ± 1.35

0.870

120 min

58.27 ± 1.09

58.44 ± 1.03

0.658

Blood staining on removal of device

9/40 (22.5%)

8/40 (20%)

1.000

Postoperative pharyngolaryngeal morbidity

Sore throat

1 h

7/40 (17.5%)

6/40 (15%)

1.000

2 h

7/40

6/40

 

24 h

0

0

 

Dysphagia

1 h

5/40 (12.5%)

4/40 (10%)

1.000

2 h

5/40

4/40

 

24 h

0

0

 

Dysphonia

1 h

0

0

 

2 h

0

0

 

24 h

0

0

 
  1. **P < 0.01, *p < 0.05, P-LMA: ProSeal laryngeal mask airway; G-LMA: Guardian laryngeal mask airway; h: hours, min: minutes, ml: milliliter.