Skip to main content

Table 2 Measurements of force required to displace the catheter from the needle in Cannula P and I as an indirect measure of static resistance

From: Functional evaluation and practice survey to guide purchasing of intravenous cannulae

 

Cannula P

Cannula I

Difference

p-value

22G static resistance; N

1.80 (25.2)

0.39 (28.7)

1.41 [1.09 to 1.73]

2 × 10-6

20G static resistance; N

0.73 (40.0)

0.55 (37.3)

0.19 [-0.04 to 0.41]

0.12

18G static resistance; N

2.40 (37.4)

0.44 (13.3)

1.96 [1.40 to 2.52]

7 × 10-5

  1. Values are mean with (relative standard deviation %) or [95% confidence interval].