Skip to main content

Table 1 Measurements of resistance through the needle in Cannula P and I

From: Functional evaluation and practice survey to guide purchasing of intravenous cannulae

 

Cannula P

Cannula I

Difference

p-value

22G resistance; mmHg.L-1.hr-1

728 (3.7)

421 (3.6)

307 [289 to 325]

2 × 10-14

20G resistance; mmHg.L-1.hr-1

287 (6.3)

152 (3.1)

135 [125 to 144]

3 × 10-10

18G resistance; mmHg.L-1.hr-1

71 (2.4)

44 (4.7)

27 [26 to 28]

1 × 10-16

  1. Values are mean with (relative standard deviation %) or [95% confidence interval].