Evaluation of a new arterial pressure-based cardiac output device requiring no external calibration

Abstract

Background

Several techniques have been discussed as alternatives to the intermittent bolus thermodilution cardiac output (COPAC) measurement by the pulmonary artery catheter (PAC). However, these techniques usually require a central venous line, an additional catheter, or a special calibration procedure. A new arterial pressure-based cardiac output (COAP) device (FloTrac™, Vigileo™; Edwards Lifesciences, Irvine, CA, USA) only requires access to the radial or femoral artery using a standard arterial catheter and does not need an external calibration. We validated this technique in critically ill patients in the intensive care unit (ICU) using COPAC as the method of reference.

Methods

We studied 20 critically ill patients, aged 16 to 74 years (mean, 55.5 ± 18.8 years), who required both arterial and pulmonary artery pressure monitoring. COPAC measurements were performed at least every 4 hours and calculated as the average of 3 measurements, while COAP values were taken immediately at the end of bolus determinations. Accuracy of measurements was assessed by calculating the bias and limits of agreement using the method described by Bland and Altman.

Results

A total of 164 coupled measurements were obtained. Absolute values of COPAC ranged from 2.80 to 10.80 l/min (mean 5.93 ± 1.55 l/min). The bias and limits of agreement between COPAC and COAP for unequal numbers of replicates was 0.02 ± 2.92 l/min. The percentage error between COPAC and COAP was 49.3%. The bias between percentage changes in COPAC (ΔCOPAC) and percentage changes in COAP (ΔCOAP) for consecutive measurements was -0.70% ± 32.28%. COPAC and COAP showed a Pearson correlation coefficient of 0.58 (p < 0.01), while the correlation coefficient between ΔCOPAC and ΔCOAP was 0.46 (p < 0.01).

Conclusion

Although the COAP algorithm shows a minimal bias with COPAC over a wide range of values in an inhomogeneous group of critically ill patients, the scattering of the data remains relative wide. Therefore, the used algorithm (V 1.03) failed to demonstrate an acceptable accuracy in comparison to the clinical standard of cardiac output determination.

Peer Review reports

Background

Accurate evaluation of cardiac performance is an important goal in the treatment of critically ill patients. Usually a pulmonary artery catheter (PAC) is placed for assessment of cardiac output (CO), but as a result of the current discussion about the usefulness and risks of the PAC [1, 2], several less invasive techniques have been discussed as alternatives to the intermittent bolus thermodilution cardiac output (COPAC) measurement by the PAC [3], which still represents the clinical standard.

Particularly techniques using arterial waveform analysis for CO assessment (PiCCOplus ®, Pulsion Medical Systems, Munich, Germany and LiDCO™plus, LiDCO Ltd, Cambridge, UK) have been tested as alternatives to the PAC, partly with excellent results [410]. However, these techniques usually require an additional catheter, a central venous line, or a special calibration procedure such as transpulmonary thermodilution (PiCCOplus ®) or lithium dilution (LiDCO™plus). Furthermore, the inability of these devices to compensate for changes in individual aortic input impedance or central aortic compliance, which both may alter the calculated CO by the pulse contour technique in hemodynamic instability, can lead to erroneous assessment of CO [11, 12].

A new arterial pressure-based cardiac output (COAP) device (FloTrac™, Vigileo™; Edwards Lifesciences, Irvine, CA, USA) only requires access to the radial or femoral artery using a standard arterial catheter and does not need an external calibration. To facilitate COAP assessment by arterial pulse waveform analysis without an external calibration mode, the system estimates individual arterial compliance according to Langewouters five component model [13] and continuously compensates for changes in vascular tone by detecting characteristic alterations in the arterial pressure waveform. But until now there has been only limited information about the value of this new device [1416].

Therefore, this study was performed to evaluate the accuracy of FloTrac™-derived COAP assessment in an inhomogeneous group of critically ill patients in a neurosurgical intensive care unit (ICU) using COPAC as the method of reference.

Methods

After obtaining approval of the Institutional Ethics Committee of the University of Regensburg Medical Centre (Regensburg, Germany) and with written informed consent from the patient or their relative, we studied 20 critically ill patients (10 male), aged 16 to 74 years (mean, 55.5 ± 18.8 years) in a neurosurgical ICU, who due to the severity of their illness required both, arterial and pulmonary artery pressure monitoring. Patients with intracardiac shunts or peripheral vascular disease were excluded from the study.

Analgesia based sedation was maintained with an infusion of fentanyl of 1.5–4.5 mg·kg-1·h-1 and midazolam of 0.15–0.35 mg·kg-1·h-1, supplemented with an infusion of ketamine of 1.5–4.0 mg·kg-1·h-1 in some patients. Pressure controlled ventilation (BiLevel ®-mode, Bennett 840 ™ Ventilator System, Puritan Bennett, Pleasanton, CA, USA) with a positive end expiratory pressure of 5–15 mmHg and a tidal volume of 6–8 ml/kg to an end tidal pCO2 of 32–38 mmHg was maintained throughout the study.

All patients received a radial arterial line for continuous monitoring of arterial blood pressure (Siemens monitor SC 9000, Erlangen, Germany). A 7.5F pulmonary artery catheter (Baxter Healthcare Corporation, Irvine, CA, USA) was inserted via an 8.5 F introducer into the internal jugular or subclavian vein for intermittent thermodilution cardiac output (COPAC) measurement (Siemens monitor SC 9000, Erlangen, Germany). COPAC measurements were performed at least every 4 hours by injection of 10 ml iced saline solution via the CVP port of the PAC and subsequent detection by the thermistor embedded into the PAC. The average of 3 measurements, all measured within a 15% range randomly distributed over the respiratory cycle, was calculated according to the Stewart-Hamilton formula. If there was more than 15% variation between the values, five measurements were performed, the highest and lowest values from CO calculation were excluded, and the remaining three values averaged.

The FloTrac™sensor was attached to the existing arterial line and connected to the Vigileo™monitor for arterial pressure-based COAP assessment. Following initiation of the Vigileo™monitor by entering patient's age, gender, height and weight, the system computes stroke volume (SV) from the patients arterial pressure signal and displays COAP continuously.

The methodology of arterial pressure-based COAP assessment by the FloTrac™system has been previously described by Manecke [17] and involves the calculation of SV regarding the proven relationship between pulse pressure (PP, the difference between systolic and diastolic blood pressure) and SV [18, 19]. As demonstrated by Boulain, aortic PP is proportional to SV and is inversely related to aortic compliance for a given SV [18]. Based upon this physiological principle, the FloTrac™device assesses the arterial pulse waveform at a sampling rate of 100 Hz over a 20 second period, which generates approximately 2000 data points, and calculates the standard deviation (SDAP) of each measured beat to provide a robust assessment of key PP characteristics. Calculation of SDAP should be more precisely related to SV, in that multiple values are measured to determine the variability, rather than a simple single PP measurement [20]. SDAP of the arterial pressure waveform is computed on a beat-to-beat basis using the following equation:

S D A P = 1 N 1 k = 0 N 1 [ A P ( k ) A P m e a n ] 2 MathType@MTEF@5@5@+=feaafiart1ev1aaatCvAUfKttLearuWrP9MDH5MBPbIqV92AaeXatLxBI9gBaebbnrfifHhDYfgasaacPC6xNi=xI8qiVKYPFjYdHaVhbbf9v8qqaqFr0xc9vqFj0dXdbba91qpepeI8k8fiI+fsY=rqGqVepae9pg0db9vqaiVgFr0xfr=xfr=xc9adbaqaaeGacaGaaiaabeqaaeqabiWaaaGcbaGaem4uamLaemiraq0aaSbaaSqaaiabdgeabjabdcfaqbqabaGccqGH9aqpdaGcaaqaaKqbaoaalaaabaGaeGymaedabaGaemOta4KaeyOeI0IaeGymaedaaOWaaabCaeaacqGGBbWwcqWGbbqqcqWGqbaucqGGOaakcqWGRbWAcqGGPaqkcqGHsislcqWGbbqqcqWGqbaudaWgaaWcbaGaemyBa0MaemyzauMaemyyaeMaemOBa4gabeaakiabc2faDnaaCaaaleqabaGaeGOmaidaaaqaaiabdUgaRjabg2da9iabicdaWaqaaiabd6eaojabgkHiTiabigdaXaqdcqGHris5aaWcbeaaaaa@50B3@

where AP(k) is the kth pulse pressure sample in the current beat, N is the total number of samples, and APmean is the mean of arterial pressure [21]. To continuously compensate for changes in vascular compliance and peripheral resistance, a scale factor χ is calculated on the basis of biometric data [13] and the analysis of characteristic alterations of the individual arterial pressure waveform, such as skewness or kurtosis, reflecting changes in vascular tone. In the used version of software (V 1.03), χ was recalculated every 10 minutes. Thus, COAP is computed as:

CO AP = HRSD APχ

where HR is the heart rate and χ a scale factor proportional to vascular compliance and peripheral resistance.

FloTrac™-derived COAP values were taken immediately at the end of bolus determinations, representing an average over the last minutes. Measurements were completed when the PAC or the arterial catheter was removed, or the patient was weaned from mechanical ventilation.

Statistical Analysis

The bias between COPAC and COAP for all values and for percentage changes (Δ = trend analysis) between consecutive CO determinations was calculated as the mean difference between measurements and expresses the agreement between methods. To avoid a systematical calculation error in the case of repeated measurements per subject, the bias between methods was calculated according to a modified statistical approach by Bland and Altman for unequal numbers of replicates [22]. The upper and lower limits of agreement (LOA), defining the range in which 95% of the differences between methods are expected to lie, were calculated as bias ± 1.96 SD. The bias and the LOA are reported as 95% confidence interval. The percentage error, defined as the standard deviation of the bias (± 1.96 SD) divided by the mean CO, was calculated according to Critchley and Critchley [23]. The acceptable limit of variability between techniques was determined as a percentage error below ± 30%. Linear regression analysis was performed between the absolute values of COPAC and COAP and between percentage changes in COPAC and COAP. A p < 0.05 was regarded as significant.

Results

Demographic data, patient characteristics and the number of measurement repetitions per patient are presented in Table 1.

Table 1 Demographic data and patient characteristics

A total of 164 coupled measurements were obtained. Absolute values of COPAC ranged from 2.80 to 10.80 l/min (mean 5.93 ± 1.55 l/min), while absolute values of COAP ranged from 3.40 to 9.80 l/min (mean 5.91 ± 1.15 l/min). The modified Bland-Altman analysis for an unequal number of replicates between COPAC and COAP showed a mean bias and LOA of 0.02 ± 2.92 l/min (Figure 1A). The percentage error between COPAC and COAP was 49.3%. COPAC and COAP showed a correlation coefficient of 0.58 (p < 0.01) as displayed in Figure 1B.

Figure 1
figure 1

Bland-Altman plot (A) and regression analysis (B) for comparison between the intermittent thermodilution-derived cardiac output (COPAC) by the pulmonary artery catheter (PAC) and the arterial pressure-based cardiac output (COAP) by the FloTrac™system for unequal numbers of replicates. The solid line represents the mean difference (bias) and the dashed lines represent the limits of agreement (bias ± 1.96 SD).

A total of 144 coupled measurements were obtained for a trend analysis of percentage changes in CO. The bias between percentage changes in COPAC (ΔCOPAC) and percentage changes in COAP (ΔCOAP) for consecutive measurements was -0.70% with LOA of ± 32.28% (Figure 2A). ΔCOPAC and ΔCOAP revealed a correlation coefficient of 0.46 (p < 0.01) as shown in Figure 2B.

Figure 2
figure 2

Bland-Altman plot (A) and regression analysis (B) for comparison between the percentage changes in intermittent thermodilution-derived cardiac output (ΔCOPAC) by the pulmonary artery catheter (PAC) and arterial pressure-based cardiac output (ΔCOAP) by the FloTrac™system for consecutive measurements. The solid line represents the mean difference (bias) and the dashed lines represent the limits of agreement (bias ± 1.96 SD).

The results of the analysis of agreement, assessed by bias, and the distribution of the observed differences, indicated by bias ± 1.96 SD as upper and lower LOA including 95% confidence intervals are presented in Table 2.

Table 2 Mean difference (bias) between intermittent thermodilution cardiac output (COPAC) and arterial pressure-based cardiac output (COAP) for all values (A) and for percentage changes between consecutive measurements of COPAC (ΔCOPAC) and COAP (ΔCOAP) (B) according to the modified approach by Bland and Altman for unequal numbers of replicates with upper and lower limits of agreement (bias ± 1.96 SD), together with 95% confidence intervals (in parentheses)

Discussion

This should be the first investigation evaluating a new arterial pressure-based CO-device in an inhomogeneous group of critically ill patients on a neurosurgical ICU in comparison to the clinical standard of CO determination. The fact that all patients received pulmonary artery pressure monitoring due to the severity of their illness differentiates our trial from previous studies, which all evaluated FloTrac-system in elective cardiac surgical patients, representing a relatively homogeneous, hemodynamically stable and artificial group of patients in which pulmonary artery pressure monitoring is not necessarily performed.

The results of the present study demonstrate a minimal bias between the absolute values of COPAC and COAP over a large range of values in an inhomogeneous group of critically ill patients. However, a relative wide scattering of the data could be observed, possibly due to the calculation mode with which the algorithm compensates for changes in vascular tone.

Several studies have shown that less invasive devices for CO assessment based on arterial waveform analysis, such as PiCCOplus ® or LiDCO™plus, are valuable alternatives to the intermittent thermodilution technique by the PAC [410]. However, all these techniques require an additional catheter, a central venous line, or an external calibration procedure. Furthermore, the inability of these techniques to continuously compensate for changes in vascular tone during hemodynamic instability [12] with the consequential need of frequent recalibrations invalidates strictly speaking the claim of a continuously measuring device. Therefore, finding a minimal invasive device for accurate and continuous CO assessment, which is simple to handle and independent of an additional catheter and an external calibration mode, is still of constant clinical interest.

The recently introduced FloTrac™system should be an interesting alternative for CO assessment and may have potential advantages due to its relative non-invasiveness and simplicity, but until now there has been only limited information about the value of this new device [1416]. Opdam and colleagues recently studied six patients after elective cardiac surgery and stated that the cardiac index (CI) values obtained with the FloTrac™system were imprecise compared with the PAC because of an inconsistent bias (0.21 l·min-1·m2) and a wide scattering of data (± 1.02 l·min-1·m2) [15]. However, major limitations of that study were the small number of patients included (6 patients), the extremely varying replicates of CO determinations obtained for each patient (8 vs. 158) and an inadequate statistical method. Not using the modified statistical approach by Bland and Altman for unequal numbers of replicates in comparison studies may falsify any result in an unpredictable manner. Variability of measurements made on different subjects is usually much greater than the variability between measurements on the same subject, and this phenomenon should be taken into account. In the worst case the findings may merely reflect the results of the patient with the majority of measuring repetitions.

Sander and co-workers compared 30 CABG patients at four different time points and demonstrated that CO measurements by the FloTrac™system showed a high bias (0.6 l/min, with a percentage error of 54%) and a wide range of LOA (± 2.8 l/min) in comparison with the COPAC measurement [16]. The authors suggest that CO assessment with pulse contour analysis techniques in a setting after cardiac surgery basically might not be the ideal method. Similar results were recently reported by Mayer et al, who found a bias and LOA of 0.46 ± 1.15 l·min-1·m2 with a percentage error of 46% for CI comparisons between methods in 40 cardiac surgical patients [14].

Our results are in accordance with the findings of the very limited number of publications regarding the accuracy of COAP assessment and also revealed a relative wide scattering of data as shown in Figure 1A. All these investigations were performed with the same version of FloTrac™system software (V 1.03) in which the algorithm compensates for changes in vascular resistance using a 10 minute moving average. Obviously, this method may have difficulty responding to rapid changes in vascular tone during hemodynamic instability. Improvement regarding the response time to changes in vascular tone has been accomplished in a subsequent version of software (V 1.07). The 10 minute moving window for assessing vascular tone was set to a 1 minute moving window.

The estimation of COPAC has become the clinical standard, despite the fact that the method has limited accuracy and may be affected by many factors such as ventilation, volume and temperature of injectate, and technique of indicator injection [24, 25], which may have contributed to the variations found between methods in the present study. Differences of at least 15% in COPAC must be achieved under clinical conditions for clinical relevance when using the triplicate method [24, 26]. Critchley and Critchley demonstrated that errors of both test and reference method should be combined when assessing comparative CO, which results in a percentage error of ± 30% to be clinically acceptable [23]. The percentage error of ± 49.3% between COPAC and COAP found in the present study exceeded the recommended limit to accept a new technique that has been compared to the reference method.

Accuracy of CO assessment is an important issue and should certainly be comparable between methods, but the more relevant question for a clinician, operating with the FloTrac™system in an ICU setting, concerns the value of COAP assessment to reasonably detect the direction of changes in CO under different hemodynamic conditions. Trend analysis of percentage changes in CO between consecutive measurements (Figure 2A) demonstrates a minimal bias between ΔCOPAC and ΔCOAP and LOA slightly above the value suggested by Critchley and Critchley [23]. While the absolute values of CO did not show an exact agreement between methods, the direction of changes in COPAC assessment was obviously identified more precisely by the FloTrac™system. However, to demonstrate that the COAP algorithm correctly identifies changes in vascular tone, interventional studies are required in which the FloTrac™system has to be compared to a continuous measuring gold standard.

The range of cardiac outputs in the present investigation is large, which should add power to our findings as all often comparable studies were performed on hypodynamic groups of patients. However, analysing the Bland-Altman plot (Figure 1A), a slight trend for the FloTrac™system to underestimate high CO values and to overestimate low CO values could be observed. This should be taken into account for further improvements of the algorithm.

Even though the FloTrac™system was designed as a screening tool in a segment of patients, who might have been under-monitored before, any new CO device has to be compared to the clinical gold standard of CO determination by the PAC in varying clinical situations before it could be recommended for a broader application. Once attached to a pre-existing arterial line, the device provides additional information within 20 seconds without any external calibration procedure, which may be valuable in the care of an endangered patient. The advantages of the device compared to other monitoring systems using the arterial pressure for CO assessment are rationalized by its simplicity and its lack of necessity for a time-consuming placement of an additional catheter.

Limitations

The results of the present study are limited by the small and varying number of CO estimations obtained for each patient. However, even in this small group of patients, the findings clearly demonstrate the limitations of the new device and the need for improvements regarding the response time to rapid changes in vascular tone. Furthermore, as changes in vascular tone have not been induced in this investigation, additional interventional studies are required to compare the latest version of FloTrac™software to a continuous measuring gold standard regarding its ability to correctly identify changes in arterial compliance.

Conclusion

FloTrac™system has the potential to be a promising alternative for cardiac output measurement, but the used algorithm (V 1.03) failed to demonstrate an acceptable accuracy in comparison to the clinical standard of CO determination. Further studies are required to evaluate the accuracy of this new device in various experimental and clinical settings using the latest version of software.

Abbreviations

AP:

arterial pressure

BMI:

body mass index

CABG:

coronary artery bypass grafting

CO:

cardiac output

COAP :

arterial pressure-based cardiac output

COPAC :

intermittent bolus thermodilution cardiac output

CVP:

central venous pressure

HR:

heart rate

ICB:

intracerebral bleeding

ICU:

intensive care unit

LOA:

limits of agreement

MVR:

mitral valve replacement

PAC:

pulmonary artery catheter

PAH:

pulmonary arterial hypertension

PP:

puls pressure

s:

seconds

SAH:

subarachnoid haemorrhage

SD:

standard deviation

SDAP :

standard deviation of the arterial pressure

SV:

stroke volume

χ:

scale factor proportional to vascular compliance and peripheral resistance.

References

  1. Connors AF, Speroff T, Dawson NV, Thomas C, Harrell FE, Wagner D, Desbiens N, Goldman L, Wu AW, Califf RM, Fulkerson WJ, Vidaillet H, Broste S, Bellamy P, Lynn J, Knaus WA: The effectiveness of right heart catheterization in the initial care of critically ill patients. JAMA. 1996, 276: 889-897.

    Article  PubMed  Google Scholar 

  2. Sandham JD, Hull RD, Brant RF, Knox L, Pineo GF, Doig CJ, Laporta DP, Viner S, Passerini L, Devitt H, Kirby A, Jacka M, Canadian Critical Care Clinical Trials Group: A randomized, controlled trial of the use of pulmonary-artery catheters in high-risk surgical patients. N Engl J Med. 2003, 348: 5-14.

    Article  PubMed  Google Scholar 

  3. Cholley BP, Payen D: Noninvasive techniques for measurements of cardiac output. Curr Opin Crit Care. 2005, 11: 424-429.

    Article  PubMed  Google Scholar 

  4. Bein B, Worthmann F, Tonner PH, Paris A, Steinfath M, Hedderich J, Scholz J: Comparison of esophageal Doppler, pulse contour analysis, and real-time pulmonary artery thermodilution for the continuous measurement of cardiac output. J Cardiothorac Vasc Anesth. 2004, 18: 185-189.

    Article  PubMed  Google Scholar 

  5. Della Rocca G, Costa MG, Coccia C, Pompei L, Di Marco P, Vilardi V, Pietropaoli P: Cardiac output monitoring: aortic transpulmonary thermodilution and pulse contour analysis agree with standard thermodilution methods in patients undergoing lung transplantation. Can J Anaesth. 2003, 50: 707-711.

    Article  PubMed  Google Scholar 

  6. Felbinger TW, Reuter DA, Eltzschig HK, Moerstedt K, Goedje O, Goetz AE: Comparison of pulmonary arterial thermodilution and arterial pulse contour analysis: evaluation of a new algorithm. J Clin Anesth. 2002, 14: 296-301.

    Article  PubMed  Google Scholar 

  7. Hamilton TT, Huber LM, Jessen ME: PulseCO: a less-invasive method to monitor cardiac output from arterial pressure after cardiac surgery. Ann Thorac Surg. 2002, 74: 1408-1412.

    Article  Google Scholar 

  8. Jonas MM, Tanser SJ: Lithium dilution measurement of cardiac output and arterial pulse waveform analysis: an indicator dilution calibrated beat-by-beat system for continuous estimation of cardiac output. Curr Opin Crit Care. 2002, 8: 257-261.

    Article  PubMed  Google Scholar 

  9. Pittman J, Bar Yosef S, SumPing J, Sherwood M, Mark J: Continuous cardiac output monitoring with pulse contour analysis: A comparison with lithium indicator dilution cardiac output measurement. Crit Care Med. 2005, 33: 2015-2021.

    Article  PubMed  Google Scholar 

  10. Rauch H, Muller M, Fleischer F, Bauer H, Martin E, Bottiger BW: Pulse contour analysis versus thermodilution in cardiac surgery patients. Acta Anaesthesiol Scand. 2002, 46: 424-429.

    Article  CAS  PubMed  Google Scholar 

  11. Rödig G, Prasser C, Keyl C, Liebold A, Hobbhahn J: Continuous cardiac output measurement: pulse contour analysis vs thermodilution technique in cardiac surgical patients. Br J Anaesth. 1999, 82: 525-530.

    Article  PubMed  Google Scholar 

  12. Wiesenack C, Fiegl C, Keyser A, Prasser C, Keyl C: Assessment of fluid responsiveness in mechanically ventilated cardiac surgical patients. Eur J Anaesth. 2005, 22: 658-665.

    Article  CAS  Google Scholar 

  13. Langewouters GJ, Wesseling KH, Goedhard WJ: The pressure dependent dynamic elasticity of 35 thoracic and 16 abdominal human aortas in vitro described by a five component model. J Biomech. 1985, 18: 613-620.

    Article  CAS  PubMed  Google Scholar 

  14. Mayer J, Boldt J, Schöllhorn T, Röhm KD, Mengistu AM, Suttner S: Semi-invasive monitoring of cardiac output by a new device using arterial pressure waveform analysis: a comparison with intermittent pulmonary artery thermodilution in patients undergoing cardiac surgery. Br J Anaesth. 2007, 98: 176-182.

    Article  CAS  PubMed  Google Scholar 

  15. Opdam HI, Wan L, Bellomo R: A pilot assessment of the FloTrac™ cardiac output monitoring system. Intensive Care Med. 2006 Oct 25.

  16. Sander M, Spies CD, Grubitzsch H, Foer A, Müller M, von Heymann C: Comparison of uncalibrated arterial waveform analysis in cardiac surgery patients with thermodilution cardiac output measurements. Crit Care. 2006, 10: R164-

    Article  PubMed  PubMed Central  Google Scholar 

  17. Manecke GR: Edwards FloTrac™sensor and Vigileo™monitor: easy, accurate, reliable cardiac output assessment using the arterial pulse wave. Expert Rev Med Devices. 2005, 2: 523-527.

    Article  PubMed  Google Scholar 

  18. Boulain T, Achard J-M, Teboul J-L, Richard C, Perrotin D, Ginies G: Changes in BP induced by passive leg raising predict response to fluid loading in critically ill patients. Chest. 2002, 121: 1245-1252.

    Article  PubMed  Google Scholar 

  19. Guyton AC, Hall JE: Vascular distensibility and functions of the arterial and venous systems. Textbook of medical physiology. Edited by: Guyton AC, Hall JE. 2000, Philadelphia: Saunders, 152-161.

    Google Scholar 

  20. Headley JM: Arterial pressure-based technologies: a new trend in cardiac output monitoring. Crit Care Nurs Clin N Am. 2006, 18: 179-187.

    Article  Google Scholar 

  21. Cecconi M, Wilson J, Rhodes A: Pulse Pressure Analysis. I Yearbook of intensive care and emergency medicine. Edited by: Vincent JL. 2006, Berlin, Heidelberg, New York: Springer, 176-183.

    Chapter  Google Scholar 

  22. Bland JM, Altman DG: Measuring agreement in method comparison studies. Stat Methods Med Res. 1999, 8: 135-160.

    Article  CAS  PubMed  Google Scholar 

  23. Critchley LA, Critchley JA: A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput. 1999, 15: 85-91.

    Article  CAS  PubMed  Google Scholar 

  24. Taylor SH, Silke B: Is the measurement of cardiac output useful in the clinical practice?. Br J Anaesth. 1988, 60: 90-98.

    Article  Google Scholar 

  25. Nishikawa T, Dohi S: Errors in the measurement of cardiac output by thermodilution. Can J Anaesth. 1993, 40: 142-153.

    Article  CAS  PubMed  Google Scholar 

  26. Stetz CW, Miller RG, Kelly GE, Raffin TA: Reliability of the thermodilution method in the determination of cardiac output in clinical practice. Am Rev Resp Dis. 1982, 126: 1001-1004.

    CAS  PubMed  Google Scholar 

Pre-publication history

Download references

Acknowledgements

This study was supported in part by a research grant from Edwards Lifesciences, Irvine, CA, USA.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Christoph Wiesenack.

Additional information

Competing interests

CW received a travel grant from Edwards Lifesciences (Irvine, USA) to present this data at an international meeting. The authors declare that there are no further competing interests.

Authors' contributions

CP designed the study and performed the statistical analysis.

SB collected the clinical data and participated in the design of the study.

CK extensively revised the manuscript.

SS collected the clinical data and participated in the analysis of the data.

BT collected and processed the clinical data.

MA collected the clinical data and participated in the analysis of the data.

JW collected the clinical data and participated in the analysis of the data.

CW designed the study, processed the data and wrote the manuscript.

All author(s) read and approved the final manuscript.

Authors’ original submitted files for images

Below are the links to the authors’ original submitted files for images.

Authors’ original file for figure 1

Authors’ original file for figure 2

Rights and permissions

Open Access This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution License ( https://creativecommons.org/licenses/by/2.0 ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Cite this article

Prasser, C., Bele, S., Keyl, C. et al. Evaluation of a new arterial pressure-based cardiac output device requiring no external calibration. BMC Anesthesiol 7, 9 (2007). https://doi.org/10.1186/1471-2253-7-9

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/1471-2253-7-9

Keywords