From: Continuous Non-invasive finger cuff CareTaker® comparable to invasive intra-arterial pressure in patients undergoing major intra-abdominal surgery

Fig. 2

Sketch of the aorta/arm complex arterial system and its effect on the arterial pressure pulse line shape that is observed at the radial/digital artery. Two reflection sites, one at the height of the renal arteries, the other one in the vicinity of the iliac bifurcation, give rise to the reflected pulses (gray) that trail the primary left ventricular ejection (black). Amplitudinal changes between the left ventricular ejection pulse P1 and the renal reflection pulse P2 as well as timing changes between P1 and the iliac reflection P3 are used to track blood pressure. An arterial stiffness measure is derived from the inversion profile of the pulse envelope