Table 4 The error grid with four zones demonstrating the level of agreement in cardiac index when comparing both Starling SV and LiDCOrapid with bolus thermodilution technique with a pulmonary artery catheter. The trending ability is optimal in zone 1, whereas in zone 4 the trending ability is poor

From: Reliability of bioreactance and pulse power analysis in measuring cardiac index during cytoreductive abdominal surgery with hyperthermic intraperitoneal chemotherapy (HIPEC)

 

Starling SV

LiDCO rapid

Error grid zone

All

n = 416

Phase 1

n = 12

Phase 2

n = 270

Phase 3

n = 64

Phase 4

n = 70

All

n = 402

Phase 1

n = 8

Phase 2

n = 262

Phase 3

n = 63

Phase 4

n = 69

Zone 1

32.5%

16.7%

40.7%

18.8%

15.7%

35.2%

37.5%

40.5%

20.6%

27.5%

Zone 2

14.4%

33.3%

13.7%

9.4%

18.6%

14.9%

12.5%

13.7%

17.5%

17.4%

Zone 3

34.4%

16.7%

31.5%

40.6%

42.9%

34.3%

37.5%

28.6%

49.2%

42.0%

Zone 4

18.8%

33.3%

14.1%

31.3%

22.9%

15.7%

12.5%

17.2%

12.7%

13.0%

  1. Phase 1 was before the initiation of cytoreductive surgery (CRS) in the operation room, phase was 2 during CRS, phase 3 was during hyperthermic intraperitoneal chemotherapy (HIPEC) and phase 4 was after HIPEC before leaving to the intensive care unit