Method 1 | Method 2 | Method 3 | Method 4 | |
---|---|---|---|---|
Workstation | Atlan A350 | Both workstations | Carestation 650 | Carestation 650 |
Procedure origin | Dräger (draft) | Dynasthetics (ACF instruction manual) | GE (technical report) | GE (technical report plus breathing system and bellows changed) |
1 | Standby mode | Vaporizer removed | Vaporizer removed | Vaporizer removed |
2 | Vaporizer removed | Manual ventilation with 10 l⋅min− 1 FGF for 90 s | Breathing circuits, breathing bag, bacterial filter and CO2 sampling line changed. | Breathing circuits, breathing bag, bacterial filter and CO2 sampling line changed |
3 | Breathing circuits, breathing bag and CO2 sampling line removed | ACF attached to both inspiratory and expiratory limps | Washout ventilation: O2 = 100%, FGF = 15 l⋅min− 1, frequency = 12⋅min− 1, VT = 700 ml, I:E = 1:2, PEEP = off. | sodium lime canister, water trap, breathing system and bellow changed. |
4 | Sodium lime canister and adapter removed | Breathing circuits, breathing bag, bacterial filter and CO2 sampling line changed | Ventilation stopped when concentration of volatile anaesthetics dropped to < 5 ppm. | Washout ventilation: O2 = 100%, FGF = 15 l⋅min− 1, frequency = 12⋅min− 1, VT = 700 ml, I:E = 1:2, PEEP = off. |
5 | Piston diaphragm and breathing system changed | System leak test | Bag-vent switch to vent position | Ventilation stopped when concentration of volatile anaesthetics dropped to < 5 ppm. |
6 | New sodium lime adapter without canister attached | Test lung ventilated with 10 l⋅min− 1 FGF | breathing circuit, breathing bag, sodium lime canister, bacterial filter and CO2 sampling line changed. | Bag-vent switch to vent position |
7 | 30 s O2-flush | System leak test | breathing circuit, breathing bag, sodium lime canister, water trap, bacterial filter and CO2 sampling line changed. | |
8 | New breathing circuits, breathing bag and CO2 sampling line attached | 10 s O2-flush | System leak test | |
9 | System leak test | Test lung ventilated with 15 l⋅min− 1 FGF | 10 s O2-flush | |
10 | APL-valve to 30cmH2O | Test lung ventilated with 15 l⋅min− 1 FGF | ||
11 | Test lung attached | |||
12 | Breathing bag removed | |||
13 | Sodium lime adapter removed | |||
14 | 60 min washout: ventilation mode = volume controlled, O2 = 100%, FGF 15 l⋅min− 1, frequency = 40⋅min− 1, Pmax = 70cmH2O, VT = 900 ml, PEEP = off, Tinsp = 0.7 s, Tplat = 20% | |||
15 | Standby mode | |||
16 | Piston diaphragm changed | |||
17 | Breathing circuits, breathing bag and CO2 sampling line removed, test lung | |||
18 | New sodium lime adapter and canister attached | |||
19 | 30 s O2-flush | |||
20 | New breathing circuits, breathing bag and CO2 sampling line attached | |||
21 | System leak test | |||
22 | Breathing bag removed ventilation mode = Man/Spon APL-valve to 30cmH2O FGF = 4 l⋅min−1 wait for “patient” | |||
23 | FGF = 10 l⋅min− 1 30 s O2-flush breathing bag re-attached attach bacterial filter | |||
24 | Test lung ventilated with a FGF of 12 l⋅min1 |