Table 1 Experimental Setup

From: Preparation of Dräger Atlan A350 and General Electric Healthcare Carestation 650 anesthesia workstations for malignant hyperthermia susceptible patients

 

Method 1

Method 2

Method 3

Method 4

Workstation

Atlan A350

Both workstations

Carestation 650

Carestation 650

Procedure origin

Dräger

(draft)

Dynasthetics

(ACF instruction manual)

GE

(technical report)

GE (technical report plus breathing system and bellows changed)

1

Standby mode

Vaporizer removed

Vaporizer removed

Vaporizer removed

2

Vaporizer removed

Manual ventilation with 10 lmin− 1 FGF for 90 s

Breathing circuits, breathing bag, bacterial filter and CO2 sampling line changed.

Breathing circuits, breathing bag, bacterial filter and CO2 sampling line changed

3

Breathing circuits, breathing bag and CO2 sampling line removed

ACF attached to both inspiratory and expiratory limps

Washout ventilation:

O2 = 100%, FGF = 15 lmin− 1, frequency = 12min− 1, VT = 700 ml, I:E = 1:2, PEEP = off.

sodium lime canister, water trap, breathing system and bellow changed.

4

Sodium lime canister and adapter removed

Breathing circuits, breathing bag, bacterial filter and CO2 sampling line changed

Ventilation stopped when concentration of volatile anaesthetics dropped to < 5 ppm.

Washout ventilation:

O2 = 100%, FGF = 15 lmin− 1, frequency = 12min− 1, VT = 700 ml, I:E = 1:2, PEEP = off.

5

Piston diaphragm and breathing system changed

System leak test

Bag-vent switch to vent position

Ventilation stopped when concentration of volatile anaesthetics dropped to < 5 ppm.

6

New sodium lime adapter without canister attached

Test lung ventilated with 10 lmin− 1 FGF

breathing circuit, breathing bag, sodium lime canister, bacterial filter and CO2 sampling line changed.

Bag-vent switch to vent position

7

30 s O2-flush

 

System leak test

breathing circuit, breathing bag, sodium lime canister, water trap, bacterial filter and CO2 sampling line changed.

8

New breathing circuits, breathing bag and CO2 sampling line attached

 

10 s O2-flush

System leak test

9

System leak test

 

Test lung ventilated with 15 lmin− 1 FGF

10 s O2-flush

10

APL-valve to 30cmH2O

  

Test lung ventilated with 15 lmin− 1 FGF

11

Test lung attached

   

12

Breathing bag removed

   

13

Sodium lime adapter removed

   

14

60 min washout: ventilation mode = volume controlled, O2 = 100%, FGF 15 lmin− 1, frequency = 40min− 1, Pmax = 70cmH2O, VT = 900 ml, PEEP = off, Tinsp = 0.7 s, Tplat = 20%

   

15

Standby mode

   

16

Piston diaphragm changed

   

17

Breathing circuits, breathing bag and CO2 sampling line removed, test lung

   

18

New sodium lime adapter and canister attached

   

19

30 s O2-flush

   

20

New breathing circuits, breathing bag and CO2 sampling line attached

   

21

System leak test

   

22

Breathing bag removed

ventilation mode = Man/Spon

APL-valve to 30cmH2O

FGF = 4 lmin−1

wait for “patient”

   

23

FGF = 10 lmin− 1

30 s O2-flush

breathing bag re-attached

attach bacterial filter

   

24

Test lung ventilated with a FGF of 12 lmin1

   
  1. Wash out procedures used in this study. ACF Activated charcoal filter, FGF Fresh gas flow, GE General Electric Healthcare, APL Adjustable pressure-limiting, VT Tidal volume, Pmax Maximum pressure, Tinsp Inspiratory time, Tplat Time of plateau pressure