From: ClearSight™ finger cuff versus invasive arterial pressure measurement in patients with body mass index above 45 kg/m2

Fig. 2

a Modified Bland-Altman plot for mean arterial pressure (MAP). The plot shows the agreement between measurements from invasive radial arterial monitoring (MAP-INV) and the finger cuff (MAP-FC). b Modified Bland-Altman plot for systolic blood pressure (SBP). The plot shows the agreement between measurements from invasive radial arterial monitoring (SBP-INV) and the finger cuff (SBP-FC). c Modified Bland-Altman plot for diastolic blood pressure (DBP). The plot shows the agreement between measurements from invasive radial arterial monitoring (DBP-INV) and the finger cuff (DBP-FC). Participants’ multiple measurements are presented individually as participant number. The middle horizontal line indicates the bias, the bottom and top lines are the 95% limits of agreement and shaded regions represent the 95% CIs for the lower and upper limits of agreement