From: Continuous EEG monitoring by a new simplified wireless headset in intensive care unit
Study group n = 20 | Control group n = 20 | |
---|---|---|
Time to assembly (min) | 6.2 ± 1.1 | 10.4 ± 2.3** |
Length of monitoring (hrs) | 57 ± 12 | 75 ± 15 ** |
Interventions (n) | 35 | 11 |
Interventions per patient (n) | 1.7 ± 1.2 | 0.5 ± 0.6** |
Electrodes replacement | 4 | 3 |
Gel or paste application | 31 | 8 |
Epileptic abnormalities | ||
Status epilepticus | 3 | 2 |
GPDs | 3 | 2 |
LPDs | 1 | 1 |
Non epileptic abnormalities | ||
Focal slow waves | 1 | 3 |
General slow waves | 4 | 3 |
EEG Asymmetries | 2 | 3 |