Fig. 3 | BMC Anesthesiology

Fig. 3

From: Perioperative non-invasive versus semi-invasive cardiac index monitoring in patients with bariatric surgery – a prospective observational study

Fig. 3

Four square plot of the concordance for cardiac index trending between Nexfin® and FloTrac™ for the total data sample. Hemodynamic trending interchangeability using a four-quadrant plot representation of the changes in cardiac index (CI) measurements from the total data sample. Data points in the left lower und right upper quadrant depict CI values with the same delta change (in %) – negative or positive. Number of values with changes in CI < 15% were excluded (exclusion zone, number of excluded values and remaining number of CI values). The concordance of the remaining values is also displayed in the diagram. An acceptable trending ability was assumed at a level of concordance > 92%. One data point lies outside the axis range