Fig. 2 | BMC Anesthesiology

Fig. 2

From: Perioperative non-invasive versus semi-invasive cardiac index monitoring in patients with bariatric surgery – a prospective observational study

Fig. 2

Correlation and Bland-Altman analyses of cardiac index measurements between Nexfin® and FloTrac™ for the total data sample. Pooled Spearman correlation analysis with correlation coefficient (rs) and P value shown in the left diagram. 2 data points lie outside of the axis range for better visualization of the diagram. The right diagram shows Bland-Altman plot of non-invasive (Nexfin®) and semi-invasive (FloTrac™) cardiac index (CI) measurements in l/min/m2. Bland-Altman analysis showing the mean difference and 95% limits of agreement for each comparison (bias ±1.96 standard deviation of the difference) as dots. 3 data points lie outside of the axis range for better visualization of the diagram. In addition, the value for the calculated percentage error (PE) is displayed in the diagram