From: Validation of noninvasive continuous arterial pressure measurement by ClearSight System™ during induction of anesthesia for cardiovascular surgery
r, Median (25-75%)
Within-subject Variability
Within-subject Precision
SAP (mmHg)
0.95 (0.89–0.96)
17.3 (4.7)
7.0 (2.6)
DAP (mmHg)
0.91 (0.86–0.95)
8.6 (2.3)
4.1 (1.4)
MAP (mmHg)
0.95 (0.91–0.96)
12.1 (2.7)
4.6 (1.4)