Author and year | Comparator | Protocol used for treatment | Device used | Position of the electrode |
---|---|---|---|---|
Low back pain | ||||
Holanda 2016 [14] | PRF treatment and lidocaine injection vs. laser irradiation | Pulse width: 20 ms with wash-out period of 480 ms; Frequency: 50 Hz; Amplitude: 45 V; Duration: 5 min (with wash out periods of 300 ms); Temperature: 42 °C | 150 mm RF probe with 5 mm active tip (company is not specified); COSMAN G4 pulse generator (Cosman Medical, Burlington, MA, USA). | L2 |
Lee 2018 [15] | Diagnostic block + PRF | Amplitude: 100 V; Duration: 240 s; Temperature: 40–42 °C | 20-gauge cannula and Cosman Four-Electrode Radiofrequency Generator (G4) (Cosman Medical, Burlington, MA, USA). | L2 - L5 and S1 |
Yang 2010 [16] | 2 implantation techniques | No stimulation parameters given | 22-gauge, SMK-C10 (Radionics Inc., Burlington, MA, USA). RF generator not specified. | L4 |
Hsu 2017 [17] | No comparator | Frequency: 2 Hz; Amplitude: 45 V; Duration: 120 s; Temperature: 42 °C | 10-cm 22-gauge sliced-tip cannula with 1 cm active tip (company is not specified); RF generator (Baylis Medical Company, Montreal, Canada). | L2 |
Tsou 2010 [18] | No comparator | Frequency: 2 Hz; Amplitude: 45 V; Duration: 120 s; Temperature: 42 °C | 10-cm, 22-gauge, curved-tip cannula with a 1 cm active tip electrode (company not specified); RF generator (Baylis Medical Co., Montreal, Canada). | L2-L5 and S1 |
Postsurgical pain | ||||
Albayrak 2017 [19] | TENS + exercise vs. TENS + exercise + PRF | Pulse width: 20 ms active and 480 ms silent periods; Frequency: 2 Hz; Amplitude: 45 V; Duration: 120 s; Temperature: 42 °C | 22-gauge cannula (OWL RF cannula 100 mm) with 5 mm active tip electrode (Diros Technology Inc., Canada); NeuroTherm 1100 RF generator (NeuroTherm, Wilmington, MA, USA). | L4 |
Cohen 2006 [20] | Intercostal nerve stimulation and medical management | Pulse width: 20 ms in 1 s cycle; Frequency: 2 Hz; Amplitude: 45 V; Duration: 120 s; Temperature: 42 °C. The procedure was repeated 4 times, for a total duration of 8 min. | 10 cm electrode with a 5 mm active tip (PMC22–100-5, Baylis Medical, Montreal, Quebec, Canada); PMG-115-TD, V2.0A RF generator (Baylis Medical Company, Montreal, Canada). | Exact DRGs are not specified |
Fam 2018 [21] | No comparator | Duration: 120 s; Temperature: 42 °C 2 cycles performed | 22-gauge, 10 cm, curved cannula with 10 mm active tip (Baylis Medical Company, Montreal, Canada). RF generator used not reported. | T2 and T3 |
Pain associated with herpes zoster | ||||
Kim 2017 [6] | Continuous epidural block (ropivacaine) | Pulse width: 20 ms; Frequency: 2 Hz; Amplitude: 45 V; Duration: 360 s; Temperature: 42 °C | 22-gauge 10 cm electrode with 10 mm active tip (Radionics Inc., Burlington, MA, USA); RF generator not specified. | Cervical, thoracic, lumbosacral (exact DRGs not specified) |
Cervicogenic headache | ||||
van Zundert 2003 [22] | No comparator | Pulse width: 20 ms; Amplitude: 45 V; Duration: 120 s (20 ms current and 480 ms without current); Temperature: 42 °C | 54 mm, 22-gauge SMK Pole needle with 4 mm active tip (Cotop International BV, Amsterdam, Netherlands); RFG 3C Plus RF generator (Radionics Inc. Burlington, MA, USA). | Cervical DRG |
Zhang 2011 [23] | No comparator | Duration: 360 s; Temperature: 42 °C | Information not given. | C2 |
Complex regional pain syndrome | ||||
Albayrak 2016 [24] | No comparator | Pulse width: 20 ms active and 480 ms silent periods; Frequency: 2 Hz; Amplitude: 40 V; Duration: 120 s; Temperature: 42 °C | 22-gauge cannula (OWL RF cannula 54 mm) with 4 mm active tip electrode (Diros Technology Inc., Canada); NeuroTherm 1100 RF generator (NeuroTherm, Wilmington, MA, USA). | C5 and C6 |
Apiliogullari 2015 [25] | No comparator | Pulse width: 20 ms active and 480 ms silent periods; Frequency: 2 Hz; Amplitude: 45 V; Duration: 120 s; Temperature: 42 °C | 22-gauge cannula (OWL RF cannula 54 mm) with 4 mm active tip electrode (Diros Technology Inc., Canada); NeuroTherm 1100 RF generator (NeuroTherm, Wilmington, MA, USA). | L4 and L5 |
Intractable vertebral metastatic pain | ||||
Arai 2015 [26] | No comparator | Pulse width: 20 ms active and 480 ms silent periods; Frequency: 2 Hz; Amplitude: 40 V; Duration: 120 s; Temperature: 42 °C | 5 mm active tip KT, guiding needle (Hakko Co. Ltd., Tokyo, Japan); RF generator JK-3 NeuroTherm (Morgan Automation Ltd., Liss, UK). | On each metastatic vertebral body, L1–5 and Th 7, 9–12 |
Scrotal and inguinal pain | ||||
Hofmeester 2013 [27] | No comparator | Pulse width: 8 ms; Frequency: 2 Hz; Amplitude: 45 V; Duration: 480 s; Temperature: 42 °C | Information not given. | T12, L1 and L2 |
Occipital radiating pain in rheumatoid arthritis | ||||
Lee 2015 [28] | No comparator | Duration: 120 s, three cycles performed; Temperature: 42 °C | 21-gauge 10 cm insulated needle (company is not specified); RF generator not specified. | C2 |
Chronic migraine | ||||
Li 2018 [29] | No comparator | Pulse width: 20 ms; Frequency: 2 Hz; Amplitude: 45 V; Duration: 900 s; Temperature: 42 °C | 22-gauge needle, RF generator G4 (Cosman Medical, Burlington, MA, USA). | C2 |