Skip to main content

Table 6 Data from intubation in Scenario F: Cervical collar stabilization and pharyngeal obstruction. Data are presented as median (IQR), or as number (percentage)

From: Comparison of the new flexible tip bougie catheter and standard bougie stylet for tracheal intubation by anesthesiologists in different difficult airway scenarios: a randomized crossover trial

parameter

standard bougie

flexible tip bougie catheter

p-value

Overall success rate, %

30 (94%)

32 (100%)

NS

Duration of 1st intubation attempt, s

53 (44–73)

44 (35–59)

0.002

Number of intubation attempts (%)

 1

21 (66%)

24 (72%)

 

 2

10 (31%)

6 (19%)

 

 3

1 (3%)

3 (9%)

 

Median (IQR)

1 (1–2)

1 (1–1.3)

NS

Number of optimization maneuvers (%)

 0

3 (9%)

10 (31%)

 

 1

8 (25%)

19 (59%)

 

 2

21 (66%)

3 (9%)

 

Median (IQR)

2 (1–2)

1 (0–1)

< 0.001

Severity of dental compression (%)

 0

0 (0%)

7 (22%)

 

 1

6 (19%)

10 (31%)

 

 2

26 (81%)

15 (47%)

 

Median (IQR)

2 (2–2)

1 (1–2)

0.008

Ease of use (1–100)

83 (72–90)

69 (54–77)

< 0.001

  1. NS Not statistically significant