Fig. 4 | BMC Anesthesiology

Fig. 4

From: A prospective observational study testing liquid crystal phase change type thermometer placed on skin against oesophageal/pharyngeal placed thermometers in participants undergoing general anesthesia

Fig. 4

Altman-Bland plots of skin temperature monitor on the middle of the sternum as compared to oesophageal temperature. (n = 92; 368 data points). a Altman-Bland Plots of skin temperature monitor on the middle of the sternum compared to oesophageal temperature WITH a forced air warmer (N = 75 [300 data points]). b Altman-Bland Plots of skin temperature monitor on the middle of the sternum compared to oesophageal temperature WITHOUT a forced air warmer (N = 17; 68 data points)