Table 3 Frequency of Aberrant Vital Signs Recorded by High Risk Bedside Monitors

From: Use of a novel electronic maternal surveillance system to generate automated alerts on the labor and delivery unit

Number of Recorded Valuesa

Maternal Early Warning Criteria

Authors’ “Flashing Red” Criteria

Authors’ Automated Paging Criteria

Criteria

Eventsb

Percent of Recorded Values

Events/24 Hours

Criteria

Events

Percent of Recorded Values

Events/24 Hours

Physician Paging Criteria

Pages to OB

Pages to OB/ 24 Hours

47,016

Systolic

BP < 90

1170

2.49%

12.86

Systolic BP < 85

581

1.24%

6.38

Systolic

BP < 85

87

0.956

47,016

Systolic

BP > 160

910

1.94%

10.00

Systolic

BP > = 160

1021

1.03%

11.22

Systolic

BP > = 160

260

2.85

97,558

Diastolic

BP > 100

1317

2.80%

14.47

Diastolic

BP > = 110

483

1.24%

5.31

Diastolic

BP > = 110

37

0.407

143,799

SpO2

<  95%

4996

5.12%

54.90

       

143,799

Heart Rate < 50

602

0.42%

6.62

       

47,016

Heart Rate > 120

4676

3.25%

51.38

Heart Rate > = 130

2168

4.61%

23.82

Heart Rate > = 130

190

2.09

 

Total Events Per 24 Hours

150.2

Total Events Per 24 Hours

46.7

Total Pages Per 24 Hours

6.30

  1. aThe data in this table are based on the vital signs recorded by the bedside monitors in 34 high risk patient rooms over a 91 day period from April 1, 2017 to July 1, 2017
  2. bThese vital signs frequencies in this table are taken directly from the bedside monitor without nurse verification, and therefore likely contain some artefactual readings