Skip to main content

Table 3 Table demonstrating haemodynamic measurements from the LiDCO and PulseCO monitors for up to eight hours. HR, MAP, CVP, LiDCO and PulseCO (median and IQR), bias and percentage error at measurements intervals.

From: A prospective study to evaluate the accuracy of pulse power analysis to monitor cardiac output in critically ill patients

 

Baseline

+1 hour

+ 2 hours

+4 hours

+8 hours

HR/min

104 (92–117)

110 (96–123)

111 (95–117)

107 (94–117)

107 (92–110)

MAP mmHg

81 (72–93)

84 (77–93)

86 (75–98)

81 (75–88)

81 (75–99)

CVP mmHg

16 (12–19)

16 (10–20)

13 (11–18)

16 (11–21)

15 (11–16)

LiDCO L/min

7.5 (5.1–9.0)

7.0 (5.0–10.6)

6.8 (5.6–10.1)

6.4 (5.3–9.8)

5.5 (4.9–6.7)

PulseCO L/min

#

7.5 (4.4–9.8)

8.6 (5.5–9.9)

7.2 (4.8–9.2)

6.1 (5.2–7.7)

Bias ± 2SD L/min

#

-0.3 ± 2.3

0.1 ± 1.9

-0.1 ± 2.2

0.2 ± 2.3

Percentage error

#

29%

22%

28%

36%